The accuracy of the FreeStyle Navigator continuous glucose monitoring system in children with type 1 diabetes.

Darrell M Wilson1, Roy W Beck, William V Tamborlane, Mariya J Dontchev, Craig Kollman, Peter Chase, Larry A Fox, Katrina J Ruedy, Eva Tsalikian, Stuart A Weinzimer.   

Abstract

OBJECTIVE: To evaluate the accuracy and precision of the FreeStyle Navigator continuous glucose monitoring system in children with type 1 diabetes. RESEARCH DESIGN AND METHODS: In 30 children with type 1 diabetes (mean age 11.2 +/- 4.1 years), the Navigator glucose values were compared with reference serum glucose values of blood samples obtained in an inpatient clinical research center and measured in a central laboratory using a hexokinase enzymatic method and in an outpatient setting with a FreeStyle meter. Median absolute difference (AD) and median relative absolute difference (RAD) were computed for sensor-reference and sensor-sensor pairs.
RESULTS: The median AD and RAD were 17 mg/dl and 12%, respectively, for 1,811 inpatient sensor-reference pairs and 20 mg/dl and 14%, respectively, for 8,639 outpatient pairs. The median RAD between two simultaneous Navigator measurements (n = 1,971) was 13%. Ninety-one percent of sensors in the inpatient setting and 81% of sensors in the outpatient setting had a median RAD < or = 20%.
CONCLUSIONS: The Navigator's accuracy does not yet approach the accuracy of current-generation home glucose meters, but it is sufficient to believe that the device has the potential to be an important adjunct to treatment of youth with type 1 diabetes.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17192334      PMCID: PMC1764630          DOI: 10.2337/dc06-1407

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


  10 in total

1.  Intensive insulin therapy in critically ill patients.

Authors:  G van den Berghe; P Wouters; F Weekers; C Verwaest; F Bruyninckx; M Schetz; D Vlasselaers; P Ferdinande; P Lauwers; R Bouillon
Journal:  N Engl J Med       Date:  2001-11-08       Impact factor: 91.245

2.  Evaluation of factors affecting CGMS calibration.

Authors:  Bruce A Buckingham; Craig Kollman; Roy Beck; Andrea Kalajian; Rosanna Fiallo-Scharer; Michael J Tansey; Larry A Fox; Darrell M Wilson; Stuart A Weinzimer; Katrina J Ruedy; William V Tamborlane
Journal:  Diabetes Technol Ther       Date:  2006-06       Impact factor: 6.118

3.  Intensive insulin therapy in the medical ICU.

Authors:  Greet Van den Berghe; Alexander Wilmer; Greet Hermans; Wouter Meersseman; Pieter J Wouters; Ilse Milants; Eric Van Wijngaerden; Herman Bobbaers; Roger Bouillon
Journal:  N Engl J Med       Date:  2006-02-02       Impact factor: 91.245

4.  Improvement in glycemic excursions with a transcutaneous, real-time continuous glucose sensor: a randomized controlled trial.

Authors:  Satish Garg; Howard Zisser; Sherwyn Schwartz; Timothy Bailey; Roy Kaplan; Samuel Ellis; Lois Jovanovic
Journal:  Diabetes Care       Date:  2006-01       Impact factor: 19.112

5.  Evaluation and comparison of 10 glucose methods and the reference method recommended in the proposed product class standard (1974).

Authors:  R B Passey; R L Gillum; J B Fuller; F M Urry; M L Giles
Journal:  Clin Chem       Date:  1977-01       Impact factor: 8.327

6.  Accuracy of newer-generation home blood glucose meters in a Diabetes Research in Children Network (DirecNet) inpatient exercise study.

Authors:  Stuart A Weinzimer; Roy W Beck; H Peter Chase; Larry A Fox; Bruce A Buckingham; William V Tamborlane; Craig Kollman; Julie Coffey; Dongyuan Xing; Katrina J Ruedy
Journal:  Diabetes Technol Ther       Date:  2005-10       Impact factor: 6.118

7.  Alarms based on real-time sensor glucose values alert patients to hypo- and hyperglycemia: the guardian continuous monitoring system.

Authors:  Bruce Bode; Kenneth Gross; Nancy Rikalo; Sherwyn Schwartz; Timothy Wahl; Casey Page; Todd Gross; John Mastrototaro
Journal:  Diabetes Technol Ther       Date:  2004-04       Impact factor: 6.118

8.  The accuracy of the CGMS in children with type 1 diabetes: results of the diabetes research in children network (DirecNet) accuracy study.

Authors: 
Journal:  Diabetes Technol Ther       Date:  2003       Impact factor: 6.118

9.  The accuracy of the GlucoWatch G2 biographer in children with type 1 diabetes: results of the diabetes research in children network (DirecNet) accuracy study.

Authors: 
Journal:  Diabetes Technol Ther       Date:  2003       Impact factor: 6.118

10.  Spurious reporting of nocturnal hypoglycemia by CGMS in patients with tightly controlled type 1 diabetes.

Authors:  Kilty McGowan; William Thomas; Antoinette Moran
Journal:  Diabetes Care       Date:  2002-09       Impact factor: 19.112

  10 in total
  44 in total

1.  Continuous subcutaneous glucose monitoring in children with type 1 diabetes.

Authors:  Amnon Zung; Zvi Zadik
Journal:  Pediatrics       Date:  2002-02       Impact factor: 7.124

2.  Effect of short-term use of a continuous glucose monitoring system with a real-time glucose display and a low glucose alarm on incidence and duration of hypoglycemia in a home setting in type 1 diabetes mellitus.

Authors:  Raymond J Davey; Timothy W Jones; Paul A Fournier
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

Review 3.  Toward closing the loop: an update on insulin pumps and continuous glucose monitoring systems.

Authors:  Tandy Aye; Jen Block; Bruce Buckingham
Journal:  Endocrinol Metab Clin North Am       Date:  2010-09       Impact factor: 4.741

4.  Use of continuous glucose monitoring in young children with type 1 diabetes: implications for behavioral research.

Authors:  Susana R Patton; Laura B Williams; Sally J Eder; Megan J Crawford; Lawrence Dolan; Scott W Powers
Journal:  Pediatr Diabetes       Date:  2011-02       Impact factor: 4.866

5.  A human pilot study of the fluorescence affinity sensor for continuous glucose monitoring in diabetes.

Authors:  Ralph Dutt-Ballerstadt; Colton Evans; Arun P Pillai; Eric Orzeck; Rafal Drabek; Ashok Gowda; Roger McNichols
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

Review 6.  A tale of two compartments: interstitial versus blood glucose monitoring.

Authors:  Eda Cengiz; William V Tamborlane
Journal:  Diabetes Technol Ther       Date:  2009-06       Impact factor: 6.118

7.  Performance evaluation of three continuous glucose monitoring systems: comparison of six sensors per subject in parallel.

Authors:  Guido Freckmann; Stefan Pleus; Manuela Link; Eva Zschornack; Hans-Martin Klötzer; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

8.  Intense exercise in type 1 diabetes: exploring the role of continuous glucose monitoring.

Authors:  Ludovic Jean Chassin; Malgorzata E Wilinska; Roman Hovorka
Journal:  J Diabetes Sci Technol       Date:  2007-07

9.  Clinical overview of continuous glucose monitoring.

Authors:  Bruce Buckingham
Journal:  J Diabetes Sci Technol       Date:  2008-03

10.  The accuracy of the Guardian RT continuous glucose monitor in children with type 1 diabetes.

Authors: 
Journal:  Diabetes Technol Ther       Date:  2008-08       Impact factor: 6.118

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.