Literature DB >> 18820984

Case reports: fractures of threaded cups: rare complications of a well-established implant.

Arnd T Hoburg1, Manav Mehta, Stephan Tohtz, Carsten Perka.   

Abstract

The use of cementless threaded cups in THA is a well-established treatment. Fractures of the cups are rare complications recorded in individual cases with material defects being discussed as the primary cause. We analyzed three cases of fractured cups. Although all three cups were well fixed to existing bone, we observed deficient osseous backing dorsocranially and abrasion particles. There were no signs of femoroacetabular impingement or infection. The cups showed corrosion debris. Scanning electron microscopic investigations showed characteristics of fretting and fretting-related corrosion. We concluded the fractures occurred because of fretting combined with inadequate bony support leading to fatigue of the material and subsequent fracture.

Entities:  

Mesh:

Year:  2008        PMID: 18820984      PMCID: PMC2635455          DOI: 10.1007/s11999-008-0526-7

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  15 in total

1.  Stress analysis of threaded cups.

Authors:  H Effenberger; U Witzel; F Lintner; W Rieger
Journal:  Int Orthop       Date:  2001       Impact factor: 3.075

2.  A 10-year minimum follow-up of hydroxyapatite-coated threaded cups: clinical, radiographic and survivorship analyses with comparison to the literature.

Authors:  Jean-Alain Epinette; Michael T Manley; James A D'Antonio; Avram A Edidin; William N Capello
Journal:  J Arthroplasty       Date:  2003-02       Impact factor: 4.757

3.  In vitro corrosion testing of modular hip tapers.

Authors:  Jay R Goldberg; Jeremy L Gilbert
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2003-02-15       Impact factor: 3.368

4.  Contact mechanics of metal-on-metal hip implants employing a metallic cup with a UHMWPE backing.

Authors:  F Liu; Z M Jin; P Grigoris; F Hirt; C Rieker
Journal:  Proc Inst Mech Eng H       Date:  2003       Impact factor: 1.617

5.  A multicenter retrieval study of the taper interfaces of modular hip prostheses.

Authors:  Jay R Goldberg; Jeremy L Gilbert; Joshua J Jacobs; Thomas W Bauer; Wayne Paprosky; Sue Leurgans
Journal:  Clin Orthop Relat Res       Date:  2002-08       Impact factor: 4.176

6.  Medium to long-term results with the cementless threaded Acetabular Munich II type cup.

Authors:  C Pellengahr; B Wegener; H Anetzberger; M Maier; P E Müller; H R Dürr; C Schulz; H Troullier; L Kühne; H J Refior
Journal:  Acta Chir Belg       Date:  2004-08       Impact factor: 1.090

7.  Acetabular reconstruction in revision hip surgery using femoral head block allograft.

Authors:  Philippe Piriou; Mark Norton; Jean-Luc Marmorat; Thierry Judet
Journal:  Orthopedics       Date:  2005-12       Impact factor: 1.390

8.  Fixation of acetabular cups without cement in total hip arthroplasty. A comparison of three different implant surfaces at a minimum duration of follow-up of five years.

Authors:  M T Manley; W N Capello; J A D'Antonio; A A Edidin; R G Geesink
Journal:  J Bone Joint Surg Am       Date:  1998-08       Impact factor: 5.284

9.  Extremely high fracture rate of a modular acetabular component with a sandwich polyethylene ceramic insertion for THA: a preliminary report.

Authors:  Jörn Kircher; Rainer Bader; Bettina Schroeder; Wolfram Mittelmeier
Journal:  Arch Orthop Trauma Surg       Date:  2008-06-21       Impact factor: 3.067

10.  Intergranular corrosion-fatigue failure of cobalt-alloy femoral stems. A failure analysis of two implants.

Authors:  J L Gilbert; C A Buckley; J J Jacobs; K C Bertin; M R Zernich
Journal:  J Bone Joint Surg Am       Date:  1994-01       Impact factor: 5.284

View more