Early polyethylene wear and excessive acetabular granuloma in an uncemented HA-coated total hip arthroplasty--midterm results of a prospective study.

Jörg Axel Karl Ohnsorge1, Jack Davis, Uwe Maus, Muharrem Saklak, Markus Weisskopf, Dieter Christian Wirtz.   

Abstract

This is a prospective review of 135 HA-coated ABG I total hip arthroplasty (THA) systems with a mean clinical and radiographic follow-up of 8.5 years. The 5-year survival rate was 85%, but 22% of the patients were dissatisfied. Revision THA was already indicated in 28% of the patients, with 26% indicated for cup loosening. PE wear was detected by x-ray in 42%. Disproportionate substantial wear with an average linear loss of 2.6 mm at the inner rim of the insert was observed in 23% of the cases. The mean annual wear rate was calculated 0.1-0.25 mm/year. Laboratory examination of the retrieved PE revealed polishing, cracks, and subsurface delamination. Radiographic evidence of acetabular cysts were found to be excessive granuloma during surgery. Polarization microscopy revealed debris particles phagocytized by reticuloendothelial cells. Results confirm the general opinion that aseptic osteolysis is a cell-mediated process driven by the presence of particles generated from wear debris. The findings also suggest that the main reasons for the failure of the first-generation ABG hip system were an insufficient locking mechanism and poor PE congruency, and not solely poor PE quality.

Entities:  

Year:  2006        PMID: 18751822      PMCID: PMC2488178          DOI: 10.1007/s11420-006-9001-8

Source DB:  PubMed          Journal:  HSS J        ISSN: 1556-3316


  30 in total

1.  The effects of cementless femoral stem shape and proximal surface texture on 'fit-and-fill' characteristics and on bone remodeling.

Authors:  H J Laine; T J Puolakka; T Moilanen; K J Pajamäki; J Wirta; M U Lehto
Journal:  Int Orthop       Date:  2000       Impact factor: 3.075

2.  Medium-term results of the ABG total hip arthroplasty in young patients.

Authors:  Konstantinos A Giannikas; Robert Din; Sahzad Sadiq; Timothy H Dunningham
Journal:  J Arthroplasty       Date:  2002-02       Impact factor: 4.757

3.  Premature wear and osteolysis in an HA-coated, uncemented total hip arthroplasty.

Authors:  P Duffy; J L Sher; P F Partington
Journal:  J Bone Joint Surg Br       Date:  2004-01

4.  The ABG hydroxyapatite-coated hip prosthesis: one hundred consecutive operations with average 6-year follow-up.

Authors:  A Rogers; R Kulkarni; E M Downes
Journal:  J Arthroplasty       Date:  2003-08       Impact factor: 4.757

5.  Clinical and radiographic evaluation of total hip replacement. A standard system of terminology for reporting results.

Authors:  R C Johnston; R H Fitzgerald; W H Harris; R Poss; M E Müller; C B Sledge
Journal:  J Bone Joint Surg Am       Date:  1990-02       Impact factor: 5.284

6.  Severe osteolysis after third-body wear due to hydroxyapatite particles from acetabular cup coating.

Authors:  E W Morscher; A Hefti; U Aebi
Journal:  J Bone Joint Surg Br       Date:  1998-03

7.  Bonding of bone to apatite-coated implants.

Authors:  R G Geesink; K de Groot; C P Klein
Journal:  J Bone Joint Surg Br       Date:  1988-01

8.  [Stretched polyethylene (UHMWPE)--a new substance modification for minimizing wear in artificial knee joint replacement].

Authors:  D C Wirtz; E Schopphoff; D Weichert; F U Niethard
Journal:  Biomed Tech (Berl)       Date:  2001-12       Impact factor: 1.411

9.  Radiographic analysis of a low-modulus titanium-alloy femoral total hip component. Two to six-year follow-up.

Authors:  A Sarmiento; T A Gruen
Journal:  J Bone Joint Surg Am       Date:  1985-01       Impact factor: 5.284

10.  Dislocations after total hip-replacement arthroplasties.

Authors:  G E Lewinnek; J L Lewis; R Tarr; C L Compere; J R Zimmerman
Journal:  J Bone Joint Surg Am       Date:  1978-03       Impact factor: 5.284

View more
  4 in total

1.  Mechanical failure of metal-polyethylene sandwich liner in metal-on-metal total hip replacement.

Authors:  Yasushi Oshima; Joseph F Fetto
Journal:  J Mater Sci Mater Med       Date:  2015-01-20       Impact factor: 3.896

2.  Inferior survival of hydroxyapatite versus titanium-coated cups at 15 years.

Authors:  Maiken Stilling; Ole Rahbek; Kjeld Søballe
Journal:  Clin Orthop Relat Res       Date:  2009-03-28       Impact factor: 4.176

Review 3.  Early subsidence of shape-closed hip arthroplasty stems is associated with late revision. A systematic review and meta-analysis of 24 RSA studies and 56 survival studies.

Authors:  Paul van der Voort; Bart G Pijls; Marc J Nieuwenhuijse; Jorrit Jasper; Marta Fiocco; Josepha W M Plevier; Saskia Middeldorp; Edward R Valstar; Rob G H H Nelissen
Journal:  Acta Orthop       Date:  2015       Impact factor: 3.717

4.  Results of a hydroxyapatite-coated femoral stem (Corail) in Chinese: a minimum 10-year follow-up.

Authors:  Jianda Xu; Zikang Xie; Jianning Zhao; Yi Gao; Hong Zhao; Libo Peng; Yuxing Qu
Journal:  Springerplus       Date:  2016-11-15
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.