P Duffy1, J L Sher, P F Partington.
Abstract
The ABG I cementless hip prosthesis has demonstrated unacceptably high rates of wear and osteolysis in our patients. We performed a retrospective study of 97 hips implanted between 1992 and 1998. Radiographic analysis revealed high rates of wear of the polyethylene liner with marked peri-acetabular osteolysis. Clinical examination indicated that many of these patients were initially asymptomatic. Wear-related problems have required ten hips to be revised and a furher 13 are awaiting revision. This gives a failure rate of 24% at a mean follow-up of 69 months. Contributing factors are likely to include poor wear characteristics of the polyethylene liners which were gamma irradiated in air, and increased wear debris caused by a poor fit of the polyethylene liner within the shell. We believe that all ABG I implants should be immediately reviewed and remain under careful, long-term follow-up.Entities:
Mesh:
Substances:
Year: 2004 PMID: 14765862
Source DB: PubMed Journal: J Bone Joint Surg Br ISSN: 0301-620X