Radiation effective doses to patients undergoing abdominal CT examinations.

D E Ware1, W Huda, P J Mergo, A L Litwiller.   

Abstract

PURPOSE: To determine the radiation effective dose to adult and pediatric patients undergoing abdominal computed tomographic (CT) examinations.
MATERIALS AND METHODS: Technique factors were obtained for three groups of randomly selected patients undergoing abdominal CT examinations: 31 children aged 10 years or younger; 32 young adults aged 11-18 years; and 36 adults older than 18 years. The radiographic techniques, together with the measured cross sections of patients, were used to estimate the total energy imparted to each patient. Each value of energy imparted was subsequently converted into the corresponding effective dose to the patient, taking into account the mass of the patient.
RESULTS: All abdominal CT examinations were performed at 120 kVp with a section thickness of approximately 7 mm for all sizes of patients. The mean number of CT sections increased from 22.0 for children to 31.5 for adults, and the mean quantity of x radiation in milliampere-seconds increased from 220 mAs for children to 290 mAs for adults. The mean values (+/- SD) of energy imparted were 72.1 mJ +/- 24.4 for children, 183.5 mJ +/- 44.8 for young adults, and 234.7 mJ +/- 89.4 for adults. The corresponding mean values of patient effective dose were 6.1 mSv +/- 1.4 for children, 4.4 mSv +/- 1.0 for young adults, and 3.9 mSv +/- 1.1 for adults.
CONCLUSION: Values of energy imparted to patients undergoing abdominal CT examinations were a factor of three higher in adults than in children, but the corresponding patient effective doses were 50% higher in children than in adults.

Entities:  

Mesh:

Year:  1999        PMID: 10207462     DOI: 10.1148/radiology.210.3.r99mr05645

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  25 in total

1.  Strategies of dose reduction.

Authors:  Donald P Frush
Journal:  Pediatr Radiol       Date:  2002-03-07

2.  Effective doses to adult and pediatric patients.

Authors:  Walter Huda
Journal:  Pediatr Radiol       Date:  2002-03-06

3.  In-patient to isocenter KERMA ratios in CT.

Authors:  Walter Huda; Kent M Ogden; Robert L Lavallee; Marsha L Roskopf; Ernest M Scalzetti
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

4.  Effect of Body Habitus on Radiation Dose During CT Fluoroscopy-Guided Spine Injections.

Authors:  Ronald J Viola; Giao B Nguyen; Terry T Yoshizumi; Sandra S Stinnett; Jenny K Hoang; Peter G Kranz
Journal:  Interv Neuroradiol       Date:  2014-10-17       Impact factor: 1.610

Review 5.  Responsible use of computed tomography in the evaluation of coronary artery disease and chest pain.

Authors:  David E Winchester; David C Wymer; Roger Y Shifrin; Steven M Kraft; James A Hill
Journal:  Mayo Clin Proc       Date:  2010-04       Impact factor: 7.616

Review 6.  CT dose reduction in children.

Authors:  Peter Vock
Journal:  Eur Radiol       Date:  2005-08-10       Impact factor: 5.315

7.  The accuracy of focused abdominal CT in patients presenting to the emergency department.

Authors:  Ali Latifi; Omid Torkzad; Fausto Labruto; Ulla Ullberg; Michael R Torkzad
Journal:  Emerg Radiol       Date:  2008-11-08

8.  Pediatric and adult trauma centers differ in evaluation, treatment, and outcomes for severely injured adolescents.

Authors:  Ashley E Walther; Richard A Falcone; Timothy A Pritts; Dennis J Hanseman; Bryce R H Robinson
Journal:  J Pediatr Surg       Date:  2016-04-12       Impact factor: 2.545

9.  Diagnostic ionizing radiation exposure in a population-based sample of children with inflammatory bowel diseases.

Authors:  Lena Palmer; Hans Herfarth; Carol Q Porter; Lynn A Fordham; Robert S Sandler; Michael D Kappelman
Journal:  Am J Gastroenterol       Date:  2009-08-18       Impact factor: 10.864

10.  Dose reduction in subsecond multislice spiral CT examination of children by online tube current modulation.

Authors:  H Greess; J Lutze; A Nömayr; H Wolf; T Hothorn; W A Kalender; W Bautz
Journal:  Eur Radiol       Date:  2004-03-30       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.