Wall apposition assessment and performance comparison of distal protection filters.

Ender A Finol1, Gail M Siewiorek, Christine M Scotti, Michael H Wholey, Mark H Wholey.   

Abstract

PURPOSE: To assess the wall apposition of 3 distal protection filters used in carotid artery stenting (CAS) for cerebral protection and quantify the effect on the in vitro capture efficiency of the filters under simulated physiological flow conditions.
METHODS: The 3 distal protection filters (Angioguard XP, FilterWire EZ, and RX Accunet) were deployed in silicone flow models of 5.0-, 5.5-, and 6.0-mm inner diameter and were tested at a mean flow rate characteristic of the human internal carotid artery while injecting polydispersed microspheres simulating plaque emboli. The injected microspheres had a diameter larger than the pore size of the devices tested, so it was conjectured that any microspheres missed by the device traveled between the device basket and the vessel wall. Varying the diameter of the vessel phantom within the recommended vessel diameter treatment range for each device simulated the variability of vessel diameter in vivo, allowing the quantification of device wall apposition.
RESULTS: None of the devices tested completely prevented distal embolization in the flow model. The RX Accunet device has the best overall wall apposition, yielding gaps of 0.075% of the vessel cross-sectional area. The FilterWire EZ device had the best overall average filtration rate, failing to capture only 0.8% of plaque particles. There were no statistically significant differences in the wall apposition assessment or the capture efficiency of the RX Accunet and FilterWire EZ devices.
CONCLUSION: Several complications related to apposition of the filter basket on the vessel wall and device retrieval were detected in all the devices. It is inferred that the adaptability of the filter basket to conform to the vessel cross section at the site of deployment is the primary design variable responsible for distal embolization during CAS with cerebral protection.

Entities:  

Mesh:

Year:  2008        PMID: 18426268     DOI: 10.1583/07-2272.1

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  9 in total

1.  Blood flow changes caused by distal filter protection and catheter aspiration in the internal carotid artery during carotid stenting: evaluation using carotid Doppler sonography.

Authors:  T Sorimachi; K Nishino; K Morita; S Takeuchi; Y Ito; Y Fujii
Journal:  AJNR Am J Neuroradiol       Date:  2010-11-04       Impact factor: 3.825

2.  Computational modeling of distal protection filters.

Authors:  Gail M Siewiorek; Ender A Finol
Journal:  J Endovasc Ther       Date:  2010-12       Impact factor: 3.487

Review 3.  The Angioguard embolic protection device.

Authors:  Gail M Siewiorek; Mark K Eskandari; Ender A Finol
Journal:  Expert Rev Med Devices       Date:  2008-05       Impact factor: 3.166

4.  In vitro performance assessment of distal protection filters: pulsatile flow conditions.

Authors:  Gail M Siewiorek; Mark H Wholey; Ender A Finol
Journal:  J Endovasc Ther       Date:  2009-12       Impact factor: 3.487

5.  The association of clinical variables and filter design with carotid artery stenting thirty-day outcome.

Authors:  G M Siewiorek; R T Krafty; M H Wholey; E A Finol
Journal:  Eur J Vasc Endovasc Surg       Date:  2011-04-29       Impact factor: 7.069

6.  Magnetic resonance plaque imaging to predict the occurrence of the slow-flow phenomenon in carotid artery stenting procedures.

Authors:  Masahiko Sakamoto; Toshiaki Taoka; Hiroyuki Nakagawa; Katsutoshi Takayama; Takeshi Wada; Kaoru Myouchin; Toshiaki Akashi; Toshiteru Miyasaka; Akio Fukusumi; Satoru Iwasaki; Kimihiko Kichikawa
Journal:  Neuroradiology       Date:  2010-04       Impact factor: 2.804

7.  Vascular resistance in the carotid artery: an in vitro investigation of embolic protection filters.

Authors:  Gail M Siewiorek; Mark H Wholey; Ender A Finol
Journal:  J Vasc Interv Radiol       Date:  2008-08-29       Impact factor: 3.464

8.  Comparison of dual protection and distal filter protection as a distal embolic protection method during carotid artery stenting: a single-center carotid artery stenting experience.

Authors:  Yosuke Kajihara; Shigeyuki Sakamoto; Yoshihiro Kiura; Kazutoshi Mukada; Takahiro Chaki; Shiro Kajihara; Kaoru Kurisu
Journal:  Neurosurg Rev       Date:  2015-05-09       Impact factor: 3.042

9.  Mid-term and late results of endovascular treatment for symptomatic carotid artery stenosis under proximal protection.

Authors:  Paweł Latacz; Marian Simka; Marcin Krzanowski; Katarzyna Krzanowska; Paweł Brzegowy; Bartłomiej Łasocha; Tadeusz J Popiela
Journal:  Wideochir Inne Tech Maloinwazyjne       Date:  2020-04-20       Impact factor: 1.195