Literature DB >> 19995122

In vitro performance assessment of distal protection filters: pulsatile flow conditions.

Gail M Siewiorek1, Mark H Wholey, Ender A Finol.   

Abstract

PURPOSE: To evaluate in vitro the capture efficiency of a distal protection filter (DPF) used during carotid artery stenting and examine the longitudinal vascular impedance in the presence of a DPF.
METHODS: Four approved DPFs (Spider RX, FilterWire EZ, RX Accunet, and FiberNet) were evaluated in a physiologically realistic in vitro setup. A pulsatile programmable piston pump circulated a blood analog at a time-varying flow rate representative of the human common carotid artery. A silicone carotid bifurcation having average human dimensions was used for the carotid flow model. Microspheres ranging from 40 to 900 microm were injected to simulate embolization. The longitudinal vascular impedance was calculated as the ratio of the time-varying pressure gradient across the DPF to the time-varying flow rate in the internal carotid artery.
RESULTS: RX Accunet had the highest capture efficiency (99.4%) and Spider RX the lowest (78.1%). Spider RX increased the longitudinal vascular impedance the least after deployment (+23%), while FilterWire EZ increased the longitudinal vascular impedance the least after particles were injected (+29%). FiberNet increased longitudinal vascular impedance the most (+84%).
CONCLUSION: This investigation, unique for examining the effects of physiologically realistic pulsatile flow on DPF performance, can aid in the development of future generations of novel DPFs.

Entities:  

Mesh:

Year:  2009        PMID: 19995122      PMCID: PMC2793305          DOI: 10.1583/09-2874.1

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  25 in total

1.  Computer-controlled positive displacement pump for physiological flow simulation.

Authors:  D W Holdsworth; D W Rickey; M Drangova; D J Miller; A Fenster
Journal:  Med Biol Eng Comput       Date:  1991-11       Impact factor: 2.602

2.  Wall apposition assessment and performance comparison of distal protection filters.

Authors:  Ender A Finol; Gail M Siewiorek; Christine M Scotti; Michael H Wholey; Mark H Wholey
Journal:  J Endovasc Ther       Date:  2008-04       Impact factor: 3.487

3.  An in vitro analysis of a carotid artery stent with a protective porous membrane.

Authors:  Stefan Müller-Hülsbeck; Erhard J Hüsler; Silvio R Schaffner; Thomas Jahnke; Christoph Glass; Rüdiger Wenke; Martin Heller
Journal:  J Vasc Interv Radiol       Date:  2004-11       Impact factor: 3.464

4.  Carotid artery stenting is associated with increased complications in octogenarians: 30-day stroke and death rates in the CREST lead-in phase.

Authors:  Robert W Hobson; Virginia J Howard; Gary S Roubin; Thomas G Brott; Robert D Ferguson; Jeffrey J Popma; Darlene L Graham; George Howard
Journal:  J Vasc Surg       Date:  2004-12       Impact factor: 4.268

5.  Geometric characterization of stenosed human carotid arteries.

Authors:  R F Smith; B K Rutt; A J Fox; R N Rankin; D W Holdsworth
Journal:  Acad Radiol       Date:  1996-11       Impact factor: 3.173

6.  Updated review of the global carotid artery stent registry.

Authors:  Michael H Wholey; Nadim Al-Mubarek; Mark H Wholey
Journal:  Catheter Cardiovasc Interv       Date:  2003-10       Impact factor: 2.692

7.  Carotid angioplasty and stenting with and without cerebral protection: clinical alert from the Endarterectomy Versus Angioplasty in Patients With Symptomatic Severe Carotid Stenosis (EVA-3S) trial.

Authors:  J L Mas; G Chatellier; B Beyssen
Journal:  Stroke       Date:  2003-12-04       Impact factor: 7.914

8.  A randomized trial of carotid artery stenting with and without cerebral protection.

Authors:  Joel E Barbato; Ellen Dillavou; Michael B Horowitz; Tudor G Jovin; Emanuel Kanal; Suzanne David; Michel S Makaroun
Journal:  J Vasc Surg       Date:  2008-03-04       Impact factor: 4.268

9.  Beneficial effect of carotid endarterectomy in symptomatic patients with high-grade carotid stenosis.

Authors:  H J M Barnett; D W Taylor; R B Haynes; D L Sackett; S J Peerless; G G Ferguson; A J Fox; R N Rankin; V C Hachinski; D O Wiebers; M Eliasziw
Journal:  N Engl J Med       Date:  1991-08-15       Impact factor: 91.245

10.  Comparison of various cerebral protection devices used for carotid artery stent placement: an in vitro experiment.

Authors:  Stefan Müller-Hülsbeck; Thomas Jahnke; Carsten Liess; Christoph Glass; Jan Grimm; Martin Heller
Journal:  J Vasc Interv Radiol       Date:  2003-05       Impact factor: 3.464

View more
  6 in total

1.  Blood flow changes caused by distal filter protection and catheter aspiration in the internal carotid artery during carotid stenting: evaluation using carotid Doppler sonography.

Authors:  T Sorimachi; K Nishino; K Morita; S Takeuchi; Y Ito; Y Fujii
Journal:  AJNR Am J Neuroradiol       Date:  2010-11-04       Impact factor: 3.825

2.  Computational modeling of distal protection filters.

Authors:  Gail M Siewiorek; Ender A Finol
Journal:  J Endovasc Ther       Date:  2010-12       Impact factor: 3.487

3.  Treatment outcomes of carotid artery stenting with two types of distal protection filter device.

Authors:  Minoru Iko; Hiroshi Aikawa; Yoshinori Go; Kanji Nakai; Masanori Tsutsumi; Iwae Yu; Taichiro Mizokami; Kimiya Sakamoto; Ritsuro Inoue; Takafumi Mitsutake; Ayumu Eto; Hayatsura Hanada; Kiyoshi Kazekawa
Journal:  Springerplus       Date:  2014-03-08

4.  Comparison of Flow Impairment during Carotid Artery Stenting Using Two Types of Eccentric Filter Embolic Protection Devices.

Authors:  Kouhei Nii; Masanori Tsutsumi; Hitoshi Maeda; Hiroshi Aikawa; Ritsuro Inoue; Ayumu Eto; Kimiya Sakamoto; Takafumi Mitsutake; Hayatsura Hanada; Kiyoshi Kazekawa
Journal:  Neurol Med Chir (Tokyo)       Date:  2016-06-16       Impact factor: 1.742

Review 5.  Permanent Bilateral Carotid Filters for Stroke Prevention in Atrial Fibrillation.

Authors:  Tom De Potter; Ofer Yodfat; Guy Shinar; Avraham Neta; Vivek Y Reddy; Petr Neuzil; Roland Veltkamp; Stuart J Connolly
Journal:  Curr Cardiol Rep       Date:  2020-09-10       Impact factor: 2.931

6.  Combined Transradial and Transfemoral Approach With Ostial Vertebral Balloon Protection for the Treatment of Patients With Subclavian Steal Syndrome.

Authors:  Rami Fakih; Sudeepta Dandapat; Alan Mendez-Ruiz; Aldo A Mendez; Mudassir Farooqui; Cynthia Zevallos; Darko Quispe Orozco; David Hasan; James Rossen; Edgar A Samaniego; Colin Derdeyn; Santiago Ortega-Gutierrez
Journal:  Front Neurol       Date:  2020-10-22       Impact factor: 4.003

  6 in total