Literature DB >> 18183396

A long-term follow-up study of the cementless THA with anatomic stem/HGPII cup with 22-mm head.

Yutaka Nakoshi1, Masahiro Hasegawa, Akihiro Sudo, Atsumasa Uchida.   

Abstract

The anatomic femoral component and Harris-Galante porous II (HGPII) cup were developed to provide more reliable bone ingrowth. We performed 20 cementless total hip arthroplasties (THAs) with anatomic stem/HGPII cup with 22-mm head in 14 consecutive patients, and evaluated the clinical and radiological results for a mean follow-up of 12.8 years. The all-anatomically designed stem provided excellent clinical and radiographic results. Four acetabular components underwent revision: three for fracture of the locking mechanism and wear of the polyethylene liner and one for the locking mechanism failure with dislocation of the HGPII cup. The abduction angles of the four revised acetabular components were apparently higher. The survivorship 13 years after surgery was 78%. Our findings show good long-term results using the anatomic femoral component, while the HGPII cup combined with 22-mm head seems to have poor durability due to locking mechanism failure.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18183396      PMCID: PMC2899065          DOI: 10.1007/s00264-007-0505-8

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  25 in total

1.  Radiological demarcation of cemented sockets in total hip replacement.

Authors:  J G DeLee; J Charnley
Journal:  Clin Orthop Relat Res       Date:  1976 Nov-Dec       Impact factor: 4.176

2.  Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results.

Authors:  C A Engh; J D Bobyn; A H Glassman
Journal:  J Bone Joint Surg Br       Date:  1987-01

3.  Polyethylene liner exchange of the Harris-Galante porous I and II acetabular components without cement: results and complications.

Authors:  Paul F Lachiewicz; Elizabeth S Soileau
Journal:  J Arthroplasty       Date:  2006-10       Impact factor: 4.757

4.  Polyethylene wear of the porous-coated anatomic total hip arthroplasty with an average 11-year follow-up.

Authors:  H Kawamura; R B Bourne; M J Dunbar; C H Rorabeck
Journal:  J Arthroplasty       Date:  2001-12       Impact factor: 4.757

5.  "Modes of failure" of cemented stem-type femoral components: a radiographic analysis of loosening.

Authors:  T A Gruen; G M McNeice; H C Amstutz
Journal:  Clin Orthop Relat Res       Date:  1979-06       Impact factor: 4.176

6.  Second-generation cementless total hip arthroplasty. Eight to eleven-year results.

Authors:  M J Archibeck; R A Berger; J J Jacobs; L R Quigley; S Gitelis; A G Rosenberg; J O Galante
Journal:  J Bone Joint Surg Am       Date:  2001-11       Impact factor: 5.284

7.  Press-fit metal-backed alumina sockets: a minimum 5-year followup study.

Authors:  P Bizot; M Larrouy; J Witvoet; L Sedel; R Nizard
Journal:  Clin Orthop Relat Res       Date:  2000-10       Impact factor: 4.176

8.  Charnley total hip arthroplasty with cement. Minimum twenty-five-year follow-up.

Authors:  J J Callaghan; J C Albright; D D Goetz; J P Olejniczak; R C Johnston
Journal:  J Bone Joint Surg Am       Date:  2000-04       Impact factor: 5.284

9.  Cementless acetabular fixation at fifteen years. A comparison with the same surgeon's results following acetabular fixation with cement.

Authors:  John L Gaffey; John J Callaghan; Douglas R Pedersen; Devon D Goetz; Patrick M Sullivan; Richard C Johnston
Journal:  J Bone Joint Surg Am       Date:  2004-02       Impact factor: 5.284

Review 10.  Locking mechanism failure in the Harris-Galante porous acetabular component associated with recurrent hip dislocation.

Authors:  Christopher L Peters; Charles L Sullivan
Journal:  J Arthroplasty       Date:  2002-06       Impact factor: 4.757

View more
  3 in total

1.  Long-term results of the standard Wagner cup.

Authors:  Uwe Demmelmeyer; Annemarie Schraml; Wolfgang Hönle; Alexander Schuh
Journal:  Int Orthop       Date:  2009-04-08       Impact factor: 3.075

2.  Total hip replacement arthroplasty with Mallory-Head system--minimum ten-year follow-up results.

Authors:  Joo-Hyoun Song; Dong-Hyun Kim; Juyoung Kim
Journal:  Int Orthop       Date:  2012-08-04       Impact factor: 3.075

3.  Fixation of an anatomically designed cementless stem in total hip arthroplasty.

Authors:  Shigeru Nakamura; Noriyuki Arai; Takateru Kobayashi; Takashi Matsushita
Journal:  Adv Orthop       Date:  2012-04-10
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.