Literature DB >> 8841266

The harmonic scalpel: an intraoperative complication.

J T Awwad1, K Isaacson.   

Abstract

BACKGROUND: The harmonic scalpel is an ultrasonically activated surgical instrument for tissue dissection. Despite its expanding surgical applications, there are no reports about associated complications. CASE: A 35-year-old woman sustained injury to the sigmoid colon from the use of the harmonic scalpel during laparoscopic lysis of pelvic adhesions. The injury was identified and repaired laparoscopically in a primary fashion with no subsequent sequelae.
CONCLUSION: Acoustic energy coupling and overheating of the laparosonic blade extender sheath occur with bending of the instrument. This can happen with steering of the blade extender during laparoscopic surgery and may increase the exposure risk of adjacent tissues to injury.

Entities:  

Mesh:

Year:  1996        PMID: 8841266     DOI: 10.1016/0029-7844(96)00075-0

Source DB:  PubMed          Journal:  Obstet Gynecol        ISSN: 0029-7844            Impact factor:   7.661


  7 in total

1.  The author replies.

Authors:  E Malik; C Berg
Journal:  Surg Endosc       Date:  2001-01       Impact factor: 4.584

2.  Temperature safety profile of laparoscopic devices: Harmonic ACE (ACE), Ligasure V (LV), and plasma trisector (PT).

Authors:  F J Kim; M F Chammas; E Gewehr; M Morihisa; F Caldas; E Hayacibara; M Baptistussi; F Meyer; A C Martins
Journal:  Surg Endosc       Date:  2007-11-20       Impact factor: 4.584

3.  Evaluation of emissivity and temperature profile of laparoscopic ultrasonic devices (blades and passive jaws).

Authors:  Fernando J Kim; David Sehrt; Rodrigo Donalisio da Silva; Diedra Gustafson; Leticia Nogueira; Wilson R Molina
Journal:  Surg Endosc       Date:  2014-08-27       Impact factor: 4.584

4.  Comparative study of electrothermal bipolar vessel sealer and ultrasonic coagulating shears in laparoscopic colectomy.

Authors:  M Takada; T Ichihara; Y Kuroda
Journal:  Surg Endosc       Date:  2004-12-02       Impact factor: 4.584

5.  Electrothermal bipolar vessel sealing device vs. ultrasonic coagulating shears in laparoscopic colectomies: a comparative study.

Authors:  Roberto Campagnacci; Angelo de Sanctis; Maddalena Baldarelli; Massimiliano Rimini; Giovanni Lezoche; Mario Guerrieri
Journal:  Surg Endosc       Date:  2007-02-08       Impact factor: 4.584

Review 6.  Common uses and cited complications of energy in surgery.

Authors:  Ganesh Sankaranarayanan; Rajeswara R Resapu; Daniel B Jones; Steven Schwaitzberg; Suvranu De
Journal:  Surg Endosc       Date:  2013-04-23       Impact factor: 4.584

7.  Rupture of the spleen with the harmonic scalpel: case report of an unexplained complication.

Authors:  John E Morrison; Volker R Jacobs
Journal:  JSLS       Date:  2007 Apr-Jun       Impact factor: 2.172

  7 in total