Literature DB >> 23402672

Performance characteristics of a positron projection imager for mouse whole-body imaging.

Jurgen Seidel1, Wenze Xi, John W Kakareka, Thomas J Pohida, Elaine M Jagoda, Michael V Green, Peter L Choyke.   

Abstract

INTRODUCTION: We describe a prototype positron projection imager (PPI) for visualizing the whole-body biodistribution of positron-emitting compounds in mouse-size animals. The final version of the PPI will be integrated into the MONICA portable dual-gamma camera system to allow the user to interchangeably image either single photon or positron-emitting compounds in a shared software and hardware environment.
METHODS: A mouse is placed in the mid-plane between two identical, opposed, pixelated LYSO arrays separated by 21.8-cm and in time coincidence. An image of the distribution of positron decays in the animal is formed on this mid-plane by coincidence events that fall within a small cone angle perpendicular to the two detectors and within a user-specified energy window. We measured the imaging performance of this device with phantoms and in tests performed in mice injected with various compounds labeled with positron-emitting isotopes.
RESULTS: Representative performance measurements yielded the following results (energy window 250-650keV, cone angle 3.5°): resolution in the image mid-plane, 1.66-mm (FWHM), resolution ±1.5-cm above and below the image plane, 2.2-mm (FWHM), sensitivity: 0.237-cps/kBq (8.76-cps/μCi) (18)F (0.024% absolute). Energy resolution was 15.9% with a linear-count-rate operating range of 0-14.8MBq (0-400μCi) and a corrected sensitivity variation across the field-of-view of <3%. Whole-body distributions of [(18)F] FDG and [(18)F] fluoride were well visualized in mice of typical size.
CONCLUSION: Performance measurements and field studies indicate that the PPI is well suited to whole-body positron projection imaging of mice. When integrated into the MONICA gamma camera system, the PPI may be particularly useful early in the drug development cycle where, like MONICA, basic whole-body biodistribution data can direct future development of the agent under study and where logistical factors (e.g., available imaging space, non-portability, and cost) may be limitations. Published by Elsevier Inc.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23402672      PMCID: PMC3596450          DOI: 10.1016/j.nucmedbio.2012.12.003

Source DB:  PubMed          Journal:  Nucl Med Biol        ISSN: 0969-8051            Impact factor:   2.408


  6 in total

1.  List-mode maximum-likelihood reconstruction applied to positron emission mammography (PEM) with irregular sampling.

Authors:  R H Huesman; G J Klein; W W Moses; J Qi; B W Reutter; P R Virador
Journal:  IEEE Trans Med Imaging       Date:  2000-05       Impact factor: 10.048

2.  The positron emission mammography/tomography breast imaging and biopsy system (PEM/PET): design, construction and phantom-based measurements.

Authors:  Raymond R Raylman; Stan Majewski; Mark F Smith; James Proffitt; William Hammond; Amarnath Srinivasan; John McKisson; Vladimir Popov; Andrew Weisenberger; Clifford O Judy; Brian Kross; Srikanth Ramasubramanian; Larry E Banta; Paul E Kinahan; Kyle Champley
Journal:  Phys Med Biol       Date:  2008-01-10       Impact factor: 3.609

3.  MONICA: a compact, portable dual gamma camera system for mouse whole-body imaging.

Authors:  Wenze Xi; Jurgen Seidel; John W Kakareka; Thomas J Pohida; Diane E Milenic; James Proffitt; Stan Majewski; Andrew G Weisenberger; Michael V Green; Peter L Choyke
Journal:  Nucl Med Biol       Date:  2010-04       Impact factor: 2.408

4.  Performance Characteristics of BGO Detectors for a Low Cost Preclinical PET Scanner.

Authors:  H Zhang; N T Vu; Q Bao; R W Silverman; B N Berry-Pusey; A Douraghy; D A Williams; F R Rannou; D B Stout; A F Chatziioannou
Journal:  IEEE Trans Nucl Sci       Date:  2010-06-01       Impact factor: 1.679

5.  Imaging of weak-source distributions in LSO-based small-animal PET scanners.

Authors:  Andrew L Goertzen; Joon Young Suk; Christopher J Thompson
Journal:  J Nucl Med       Date:  2007-09-14       Impact factor: 10.057

6.  Dual-modality molecular imaging using antibodies labeled with activatable fluorescence and a radionuclide for specific and quantitative targeted cancer detection.

Authors:  Mikako Ogawa; Celeste A S Regino; Jurgen Seidel; Michael V Green; Wenze Xi; Mark Williams; Nobuyuki Kosaka; Peter L Choyke; Hisataka Kobayashi
Journal:  Bioconjug Chem       Date:  2009-11       Impact factor: 4.774

  6 in total
  2 in total

1.  Comparison of planar, PET and well-counter measurements of total tumor radioactivity in a mouse xenograft model.

Authors:  Michael V Green; Jurgen Seidel; Mark R Williams; Karen J Wong; Anita Ton; Falguni Basuli; Peter L Choyke; Elaine M Jagoda
Journal:  Nucl Med Biol       Date:  2017-06-30       Impact factor: 2.408

Review 2.  Advances in Preclinical PET.

Authors:  Stephen S Adler; Jurgen Seidel; Peter L Choyke
Journal:  Semin Nucl Med       Date:  2022-03-18       Impact factor: 4.802

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.