Comparison and reproducibility of transthoracic bioimpedance and dual beam Doppler ultrasound measurement of cardiac function in healthy volunteers.

H W Ng1, T Walley, Y Tsao, A M Breckenridge.   

Abstract

1. We compared the ease of use and reproducibility of two noninvasive methods, transthoracic electrical bioimpedance (TEB) (BoMed NCCOM3-R7) and non-imaging dual beam Doppler ultrasound (Quantascope--Vital Science), in cardiac output (CO) and stroke volume (SV) measurement in healthy volunteers at rest and during physiological stress, both short term and from day to day. 2. The TEB method was easier to use and not dependent on the operator. The TEB method was more reproducible both in the short term and from day to day. Both devices were able to detect CO and SV changes under physiological stress, but were less reproducible day to day during exercise. 3. At supine rest, the within subject coefficient of variation between time point for TEB was 4.6% for CO and 6.1% for SV, and 7.9% for CO and 7.4% for SV with Doppler. 4. The results from each device showed a linear correlation coefficient (r) of 0.69 for CO (P less than 0.0005) and 0.64 for SV (P less than 0.0005). The correlation coefficient improved to 0.76 (P less than 0.0005) when only changes in CO and SV were considered. There was no systematic difference in the changes detected by the two methods, but the individual variation was wide.

Entities:  

Mesh:

Year:  1991        PMID: 1777364      PMCID: PMC1368518          DOI: 10.1111/j.1365-2125.1991.tb03899.x

Source DB:  PubMed          Journal:  Br J Clin Pharmacol        ISSN: 0306-5251            Impact factor:   4.335


  12 in total

1.  Blood flow measurement using the attenuation-compensated volume flowmeter.

Authors:  C F Hottinger; J D Meindl
Journal:  Ultrason Imaging       Date:  1979-01       Impact factor: 1.578

2.  Non-invasive determination of cardiac output by Doppler echocardiography and electrical bioimpedance.

Authors:  D B Northridge; I N Findlay; J Wilson; E Henderson; H J Dargie
Journal:  Br Heart J       Date:  1990-02

3.  Multicomponent noninvasive physiologic monitoring of circulatory function.

Authors:  W C Shoemaker; P L Appel; H B Kram; R C Nathan; J L Thompson
Journal:  Crit Care Med       Date:  1988-05       Impact factor: 7.598

4.  Pharmacodynamic monitoring during acute intervention in ischaemic heart disease using a new echo-Doppler device.

Authors:  B Silke; S P Verma; S H Taylor
Journal:  Br J Clin Pharmacol       Date:  1990-06       Impact factor: 4.335

5.  A clinical comparison between cardiac output measured by thermodilution versus noninvasive thoracic electrical bioimpedance.

Authors:  J A Mattar; A C Baruzzi; D Diament; R T Szynkier; J de Felippe; P L da Luz; J O Auler; S Lage; F Pileggi; A Jatene
Journal:  Acute Care       Date:  1986

6.  Comparison of measurements of cardiac output by bioimpedance and thermodilution in severely ill surgical patients.

Authors:  P L Appel; H B Kram; J Mackabee; A W Fleming; W C Shoemaker
Journal:  Crit Care Med       Date:  1986-11       Impact factor: 7.598

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

Review 8.  Systolic time intervals: a review of the method in the non-invasive investigation of cardiac function in health, disease and clinical pharmacology.

Authors:  S Hassan; P Turner
Journal:  Postgrad Med J       Date:  1983-07       Impact factor: 2.401

Review 9.  Is the measurement of cardiac output useful in clinical practice?

Authors:  S H Taylor; B Silke
Journal:  Br J Anaesth       Date:  1988       Impact factor: 9.166

10.  Continuous noninvasive real-time monitoring of stroke volume and cardiac output by thoracic electrical bioimpedance.

Authors:  D P Bernstein
Journal:  Crit Care Med       Date:  1986-10       Impact factor: 7.598

View more
  9 in total

1.  Agreement and reproducibility of the estimates of cardiovascular function by impedance cardiography and M-mode echocardiography in healthy subjects.

Authors:  C de Mey; J Matthews; R Butzer; V Schroeter; G G Belz
Journal:  Br J Clin Pharmacol       Date:  1992-07       Impact factor: 4.335

2.  Comparative effects of ivabradine, a selective heart rate-lowering agent, and propranolol on systemic and cardiac haemodynamics at rest and during exercise.

Authors:  Robinson Joannides; Nicholas Moore; Michaela Iacob; Patricia Compagnon; Guy Lerebours; Jean-François Menard; Christian Thuillez
Journal:  Br J Clin Pharmacol       Date:  2006-02       Impact factor: 4.335

3.  Impedance cardiography.

Authors:  H W Ng; T Walley
Journal:  Br J Clin Pharmacol       Date:  1993-07       Impact factor: 4.335

4.  Dose-dependent heart rate reducing effect of nizatidine, a histamine H2-receptor antagonist.

Authors:  H Hinrichsen; A Halabi; G Fuhrmann; W Kirch
Journal:  Br J Clin Pharmacol       Date:  1993-05       Impact factor: 4.335

5.  Disagreement between standard transthoracic impedance cardiography and the automated transthoracic electrical bioimpedance method in estimating the cardiovascular responses to phenylephrine and isoprenaline in healthy man.

Authors:  C De Mey; D Enterling
Journal:  Br J Clin Pharmacol       Date:  1993-04       Impact factor: 4.335

6.  Short-term postural reflexes in diabetic patients with autonomic dysfunction.

Authors:  Y Zhang; L A Critchley; Y H Tam; B Tomlinson
Journal:  Diabetologia       Date:  2003-12-03       Impact factor: 10.122

7.  Lack of pharmacodynamic interactions between acute dose flosequinan and xamoterol. A pilot study in healthy subjects.

Authors:  H W Ng; T J Walley; Y Tsao; A M Breckenridge
Journal:  Eur J Clin Pharmacol       Date:  1994       Impact factor: 2.953

8.  Haemodynamic comparison of amlodipine and atenolol in essential hypertension using the quantascope.

Authors:  T C Tham; N Herity; S Guy; B Silke
Journal:  Br J Clin Pharmacol       Date:  1993-12       Impact factor: 4.335

9.  Central Hemodynamics Measured During 5 Repetition Maximum Free Weight Resistance Exercise.

Authors:  Jonathan S Howard; Cherilyn N McLester; Thomas W Evans; John R McLester; Jimmy P Calloway
Journal:  Int J Exerc Sci       Date:  2018-01-01
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.