Intrasession and intersession repeatability of the Pentacam system on posterior corneal assessment in the normal human eye.

Davie Chen1, Andrew K C Lam.   

Abstract

PURPOSE: To study the intrasession and intersession repeatability of the Pentacam system (Oculus Inc.) in measuring the posterior corneal shape.
SETTING: School of Optometry, Hong Kong Polytechnic University, Hong Kong, China.
METHODS: The posterior corneal shape in 1 eye (randomly selected) was measured with the Pentacam system. Three consecutive readings were taken for intrasession repeatability analysis. Measurements were repeated 1 to 2 weeks later for intersession repeatability analysis. The anterior best-fit sphere (BFS) and posterior BFS at the 5.0 mm and 8.0 mm zones, as well as the elevation of the posterior cornea at these 2 zones, were compared.
RESULTS: The anterior and posterior BFS demonstrated good intrasession repeatability (Cronbach alpha > or = 0.997; intraclass correlation coefficients > or = 0.990). No significant difference was found between the 2 visits. The 95% limits of agreement of posterior BFS were 52.3 to -41.7 microm at the 5.0 mm zone and 66.1 to -60.9 microm at the 8.0 mm zone. The elevation was similar between the 2 visits. The 95% limits of agreement were 4.27 to -5.29 microm at the 5.0 mm zone and 5.33 to -5.77 microm at the 8.0 mm zone.
CONCLUSIONS: Pentacam provided good performance in assessing the posterior cornea. To improve the intersession repeatability, it is suggested that 3 consecutive readings, rather than 1 image, be taken to generate an average BFS for analysis. When comparing the posterior corneal elevation between sessions, an average BFS generated from the first visit should be used for elevation calculation.

Entities:  

Mesh:

Year:  2007        PMID: 17321396     DOI: 10.1016/j.jcrs.2006.11.008

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  24 in total

1.  Validation of corneal topographic and aberrometric measurements obtained by color light-emitting diode reflection topography in healthy eyes.

Authors:  David P Piñero; Ainhoa Molina-Martín; Vicent J Camps; Dolores de Fez; María Teresa Caballero
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2019-09-03       Impact factor: 3.117

2.  Corneal elevation topography: best fit sphere, elevation distance, asphericity, toricity, and clinical implications.

Authors:  Damien Gatinel; Jacques Malet; Thanh Hoang-Xuan; Dimitri T Azar
Journal:  Cornea       Date:  2011-05       Impact factor: 2.651

3.  Effects of pterygium surgery on front and back corneal surfaces and anterior segment parameters.

Authors:  Ahmad Kheirkhah; Hamid Safi; Rahman Nazari; Reza Kaghazkanani; Hassan Hashemi; Mahmoud Jabbarvand Behrouz
Journal:  Int Ophthalmol       Date:  2012-04-07       Impact factor: 2.031

4.  Posterior corneal elevation changes and characteristic analysis 1 year after corneal collagen cross-linking for keratoconus.

Authors:  Jing Li; Shengsheng Wei; Yaohua Zhang; Dan Yang; Zhen Tian; Jing Du; Yong Li; Jianguo Liu; Yan Wang
Journal:  Int Ophthalmol       Date:  2022-01-06       Impact factor: 2.031

5.  The effect of posterior corneal flat meridian and astigmatism amount on the total corneal astigmatism estimated from anterior corneal measurements.

Authors:  Youngsub Eom; Su-Yeon Kang; Hyo Myung Kim; Jong Suk Song
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-07-20       Impact factor: 3.117

6.  Corneal topography from spectral optical coherence tomography (sOCT).

Authors:  Sergio Ortiz; Damian Siedlecki; Pablo Pérez-Merino; Noelia Chia; Alberto de Castro; Maciej Szkulmowski; Maciej Wojtkowski; Susana Marcos
Journal:  Biomed Opt Express       Date:  2011-11-04       Impact factor: 3.732

7.  Posterior corneal curvature assessment after Epi-LASIK for myopia: comparison of Orbscan II and Pentacam imaging.

Authors:  Yong-Soo Byun; So-Hyang Chung; Young-Geun Park; Choun-Ki Joo
Journal:  Korean J Ophthalmol       Date:  2012-01-14

8.  Evaluation of Central Corneal Thickness Using Corneal Dynamic Scheimpflug Analyzer Corvis ST and Comparison with Pentacam Rotating Scheimpflug System and Ultrasound Pachymetry in Normal Eyes.

Authors:  Ayong Yu; Weiqi Zhao; Giacomo Savini; Zixu Huang; Fangjun Bao; Weicong Lu; Qinmei Wang; Jinhai Huang
Journal:  J Ophthalmol       Date:  2015-12-01       Impact factor: 1.909

9.  Correlation between apical protrusion in the Scheimflug imaging and Corneal Hysteresis and Corneal Resistance factor by Ocular Response Analyzer, among refractive non-keratoconic Egyptian patients.

Authors:  Tamer Adel Refai
Journal:  Electron Physician       Date:  2015-10-19

10.  Anterior segment study with the Pentacam Scheimpflug camera in refractive surgery candidates.

Authors:  Masih Hashemi; Khalil Ghasemi Falavarjani; Gholam Hosseyn Aghai; Kaveh Abri Aghdam; Arzhang Gordiz
Journal:  Middle East Afr J Ophthalmol       Date:  2013 Jul-Sep
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.