Two-dimensional dynamic simulation of platelet activation during mechanical heart valve closure.

S Krishnan1, H S Udaykumar, J S Marshall, K B Chandran.   

Abstract

A major drawback in the operation of mechanical heart valve prostheses is thrombus formation in the near valve region. Detailed flow analysis in this region during the valve closure phase is of interest in understanding the relationship between shear stress and platelet activation. A fixed-grid Cartesian mesh flow solver is used to simulate the blood flow through a bi-leaflet mechanical valve employing a two-dimensional geometry of the leaflet with a pivot point representing the hinge region. A local mesh refinement algorithm allows efficient and fast flow computations with mesh adaptation based on the gradients of the flow field in the leaflet-housing gap at the instant of valve closure. Leaflet motion is calculated dynamically based on the fluid forces acting on it employing a fluid-structure interaction algorithm. Platelets are modeled and tracked as point particles by a Lagrangian particle tracking method which incorporates the hemodynamic forces on the particles. A platelet activation model is included to predict regions which are prone to platelet activation. Closure time of the leaflet is validated against experimental studies. Results show that the orientation of the jet flow through the gap between the housing and the leaflet causes the boundary layer from the valve housing to be drawn in by the shear layer separating from the leaflet. The interaction between the separating shear layers is seen to cause a region of intensely rotating flow with high shear stress and high residence time of particles leading to high likelihood of platelet activation in that region.

Mesh:

Year:  2006        PMID: 17013660     DOI: 10.1007/s10439-006-9194-5

Source DB:  PubMed          Journal:  Ann Biomed Eng        ISSN: 0090-6964            Impact factor:   3.934


  20 in total

1.  Influence of three mechanical bileaflet prosthetic valve designs on the three-dimensional flow field inside a simulated aorta.

Authors:  Toshinosuke Akutsu; Akira Matsumoto
Journal:  J Artif Organs       Date:  2010-10-07       Impact factor: 1.731

2.  Are anticoagulant independent mechanical valves within reach-fast prototype fabrication and in vitro testing of innovative bi-leaflet valve models.

Authors:  Lawrence N Scotten; Rolland Siegel
Journal:  Ann Transl Med       Date:  2015-08

3.  The effect of implantation orientation of a bileaflet mechanical heart valve on kinematics and hemodynamics in an anatomic aorta.

Authors:  Iman Borazjani; Fotis Sotiropoulos
Journal:  J Biomech Eng       Date:  2010-11       Impact factor: 2.097

4.  Flow interactions with cells and tissues: cardiovascular flows and fluid-structure interactions. Sixth International Bio-Fluid Mechanics Symposium and Workshop, March 28-30, 2008, Pasadena, California.

Authors:  Morton H Friedman; Rob Krams; Krishnan B Chandran
Journal:  Ann Biomed Eng       Date:  2010-03       Impact factor: 3.934

5.  Dynamic particle image velocimetry study of the aortic flow field of contemporary mechanical bileaflet prostheses.

Authors:  Toshinosuke Akutsu; Jun Saito; Ryota Imai; Tomoyuki Suzuki; Xiu Dong Cao
Journal:  J Artif Organs       Date:  2008-07-06       Impact factor: 1.731

Review 6.  Review of numerical methods for simulation of mechanical heart valves and the potential for blood clotting.

Authors:  Mohamad Shukri Zakaria; Farzad Ismail; Masaaki Tamagawa; Ahmad Fazli Abdul Aziz; Surjatin Wiriadidjaja; Adi Azrif Basri; Kamarul Arifin Ahmad
Journal:  Med Biol Eng Comput       Date:  2017-07-26       Impact factor: 2.602

7.  Tortuosity triggers platelet activation and thrombus formation in microvessels.

Authors:  Jennifer K W Chesnutt; Hai-Chao Han
Journal:  J Biomech Eng       Date:  2011-12       Impact factor: 2.097

8.  Impact of design parameters on bileaflet mechanical heart valve flow dynamics.

Authors:  Vijay Govindarajan; Holavanahalli S Udaykumar; Luke H Herbertson; Steven Deutsch; Keefe B Manning; Krishnan B Chandran
Journal:  J Heart Valve Dis       Date:  2009-09

9.  FLOW DYNAMIC COMPARISON BETWEEN RECESSED HINGE AND OPEN PIVOT BI-LEAFLET HEART VALVE DESIGNS.

Authors:  V Govindarajan; H S Udaykumar; K B Chandran
Journal:  J Mech Med Biol       Date:  2009-06-01       Impact factor: 0.897

Review 10.  Lagrangian postprocessing of computational hemodynamics.

Authors:  Shawn C Shadden; Amirhossein Arzani
Journal:  Ann Biomed Eng       Date:  2014-07-25       Impact factor: 3.934