Literature DB >> 26417581

Are anticoagulant independent mechanical valves within reach-fast prototype fabrication and in vitro testing of innovative bi-leaflet valve models.

Lawrence N Scotten1, Rolland Siegel1.   

Abstract

BACKGROUND: Exploration for causes of prosthetic valve thrombogenicity has frequently focused on forward or post-closure flow detail. In prior laboratory studies, we uncovered high amplitude flow velocities of short duration close to valve closure implying potential for substantial shear stress with subsequent initiation of blood coagulation pathways. This may be relevant to widely accepted clinical disparity between mechanical and tissue valves vis-à-vis thrombogenicity. With a series of prototype bi-leaflet mechanical valves, we attempt reduction of closure related velocities with the objective of identifying a prototype valve with thrombogenic potential similar to our tissue valve control. This iterative design approach may find application in preclinical assessment of valves for anticoagulation independence.
METHODS: Tested valves included: prototype mechanical bi-leaflet BVs (n=56), controls (n=2) and patented early prototype mechanicals (n=2) from other investigators. Pulsatile and quasi-steady flow systems were used for testing. Projected dynamic valve area (PDVA) was measured using previously described novel technology. Flow velocity over the open and closing periods was determined by volumetric flow rate/PDVA. For the closed valve interval, use was made of data obtained from quasi-steady back pressure/flow tests. Performance was ranked by a proposed thrombogenicity potential index (TPI) relative to tissue and mechanical control valves.
RESULTS: Optimization of the prototype valve designs lead to a 3-D printed model (BV3D). For the mitral/aortic site, BV3D has lower TPI (1.10/1.47) relative to the control mechanical valve (3.44/3.93) and similar to the control tissue valve (ideal TPI ≤1.0).
CONCLUSIONS: Using unique technology, rapid prototyping and thrombogenicity ranking, optimization of experimental valves for reduced thrombogenic potential was expedited and simplified. Innovative mechanical valve configurations were identified that merit consideration for further development which may bring the anti-coagulation independent mechanical valve within reach.

Entities:  

Keywords:  Prosthetic valve; flow velocities; rebound; thrombogenicity

Year:  2015        PMID: 26417581      PMCID: PMC4560704          DOI: 10.3978/j.issn.2305-5839.2015.08.18

Source DB:  PubMed          Journal:  Ann Transl Med        ISSN: 2305-5839


  30 in total

1.  New laboratory technique measures projected dynamic area of prosthetic heart valves.

Authors:  Lawrence N Scotten; David K Walker
Journal:  J Heart Valve Dis       Date:  2004-01

Review 2.  Research on biological and mechanical heart valves: experimental studies in chronic animal models.

Authors:  B Meuris
Journal:  Verh K Acad Geneeskd Belg       Date:  2002

3.  Reduction of procoagulant potential of b-datum leakage jet flow in bileaflet mechanical heart valves via application of vortex generator arrays.

Authors:  David W Murphy; Lakshmi P Dasi; Jelena Vukasinovic; Ari Glezer; Ajit P Yoganathan
Journal:  J Biomech Eng       Date:  2010-07       Impact factor: 2.097

Review 4.  Characterization of neurological injury in transcatheter aortic valve implantation: how clear is the picture?

Authors:  Jonathon P Fanning; Darren L Walters; David G Platts; Eamonn Eeles; Judith Bellapart; John F Fraser
Journal:  Circulation       Date:  2014-01-28       Impact factor: 29.690

5.  Simulating the Dynamics of an Aortic Valve Prosthesis in a Pulse Duplicator: Numerical Methods and Initial Experience.

Authors:  Boyce E Griffith; Vittoria Flamini; Abe DeAnda; Lawrence Scotten
Journal:  J Med Device       Date:  2013-12-05       Impact factor: 0.582

6.  Very long-term outcomes of the Carpentier-Edwards Perimount valve in aortic position.

Authors:  Thierry Bourguignon; Anne-Lorraine Bouquiaux-Stablo; Pascal Candolfi; Alain Mirza; Claudia Loardi; Marc-Antoine May; Rym El-Khoury; Michel Marchand; Michel Aupart
Journal:  Ann Thorac Surg       Date:  2015-01-09       Impact factor: 4.330

7.  Platelet glycoprotein Ibα ectodomain shedding and non-surgical bleeding in heart failure patients supported by continuous-flow left ventricular assist devices.

Authors:  Jingping Hu; Nandan K Mondal; Erik N Sorensen; Ling Cai; Hong-Bin Fang; Bartley P Griffith; Zhongjun J Wu
Journal:  J Heart Lung Transplant       Date:  2013-09-19       Impact factor: 10.247

8.  Quantification of Shear-Induced Platelet Activation: High Shear Stresses for Short Exposure Time.

Authors:  Jun Ding; Zengsheng Chen; Shuqiong Niu; Jiafeng Zhang; Nandan K Mondal; Bartley P Griffith; Zhongjun J Wu
Journal:  Artif Organs       Date:  2015-03-21       Impact factor: 3.094

Review 9.  Thromboembolic and bleeding complications in patients with mechanical heart valve prostheses.

Authors:  S C Cannegieter; F R Rosendaal; E Briët
Journal:  Circulation       Date:  1994-02       Impact factor: 29.690

10.  Challenges in cardiac device innovation: is neuroimaging an appropriate endpoint? Consensus from the 2013 Yale-UCL Cardiac Device Innovation Summit.

Authors:  Stephanie M Meller; Andreas Baumbach; Szilard Voros; Michael Mullen; Alexandra J Lansky
Journal:  BMC Med       Date:  2013-12-11       Impact factor: 8.775

View more
  2 in total

1.  Anticoagulant independent mechanical heart valves: viable now or still a distant holy grail.

Authors:  Aurelio Chaux; Richard J Gray; Jonathan C Stupka; Michael R Emken; Lawrence N Scotten; Rolland Siegel
Journal:  Ann Transl Med       Date:  2016-12

Review 2.  2006 to 2019 Story; percutaneously implantable aortic valve prototypes.

Authors:  Constantinos Zervides; Ornella Nohra; Gabriel Hunduma; Neil Wild Thomas; Ramy Samia
Journal:  J Cardiothorac Surg       Date:  2021-08-06       Impact factor: 1.637

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.