| Literature DB >> 16894575 |
V Andrew Stenger1, Marius S Giurgi, Fernando E Boada, Douglas C Noll.
Abstract
An extension of the "UNaliasing by Fourier encoding the Overlaps using the temporaL Dimension" (UNFOLD) method to the excitation domain (XUNFOLD) is presented to improve the temporal resolution of multishot tailored RF (TRF) pulses. Multishot three-dimensional TRF pulses were designed to produce a time series of images with periodically aliased excitation profiles. The XUNFOLD method is shown to remove the excitation profile aliasing from the dynamic imaging data by filtering in the temporal frequency dimension. The technique is demonstrated to improve the temporal resolution of simulated functional MRI (fMRI) activation in a time series of brain images. Copyright (c) 2006 Wiley-Liss, Inc.Mesh:
Year: 2006 PMID: 16894575 PMCID: PMC3050053 DOI: 10.1002/mrm.21000
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668