Functional MR imaging using gradient-echo echo-planar imaging in the presence of large static field inhomogeneities.

R T Constable1.   

Abstract

At 1.5 T, the field strength of most clinical MR imagers, gradient-echo imaging is the primary imaging method for measuring brain activation, as such sequences are highly sensitive to changes in blood oxygenation or T2* effects. Unfortunately, gradient-echo sequences are also extremely sensitive to magnetic field inhomogeneities, and this sensitivity has precluded examination of regions of cortex near field inhomogeneities with functional MR imaging. This article presents a gradient-echo echo-planar imaging method that uses variable amplitude scaling on the slice-select refocusing lobe to generate images compensated for static field inhomogeneities. A technique for constructing composite images to be used in statistical tests for activation is also presented. The method is shown to produce clean activation maps in the presence of large static field inhomogeneities. The technique retains the sensitivity of gradient-echo imaging to changes in blood oxygenation while removing the sensitivity to large static field inhomogeneities.

Mesh:

Substances:

Year:  1995        PMID: 8748497     DOI: 10.1002/jmri.1880050622

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  18 in total

1.  Three-dimensional tailored RF pulses for the reduction of susceptibility artifacts in T(*)(2)-weighted functional MRI.

Authors:  V A Stenger; F E Boada; D C Noll
Journal:  Magn Reson Med       Date:  2000-10       Impact factor: 4.668

2.  Multishot 3D slice-select tailored RF pulses for MRI.

Authors:  V Andrew Stenger; Fernando E Boada; Douglas C Noll
Journal:  Magn Reson Med       Date:  2002-07       Impact factor: 4.668

3.  Variable-density spiral 3D tailored RF pulses.

Authors:  V Andrew Stenger; Fernando E Boada; Douglas C Noll
Journal:  Magn Reson Med       Date:  2003-11       Impact factor: 4.668

4.  Excitation UNFOLD (XUNFOLD) to improve the temporal resolution of multishot tailored RF pulses.

Authors:  V Andrew Stenger; Marius S Giurgi; Fernando E Boada; Douglas C Noll
Journal:  Magn Reson Med       Date:  2006-09       Impact factor: 4.668

5.  Imaging mesial temporal lobe activation during scene encoding: comparison of fMRI using BOLD and arterial spin labeling.

Authors:  María A Fernández-Seara; Jiongjiong Wang; Ze Wang; Marc Korczykowski; Matthias Guenther; David A Feinberg; John A Detre
Journal:  Hum Brain Mapp       Date:  2007-12       Impact factor: 5.038

6.  Simultaneous z-shim method for reducing susceptibility artifacts with multiple transmitters.

Authors:  Weiran Deng; Cungeng Yang; Vijayanand Alagappan; Lawrence L Wald; Fernando E Boada; V Andrew Stenger
Journal:  Magn Reson Med       Date:  2009-02       Impact factor: 4.668

7.  Variable slice thickness (VAST) EPI for the reduction of susceptibility artifacts in whole-brain GE-EPI at 7 Tesla.

Authors:  Sascha Brunheim; Sören Johst; Viktor Pfaffenrot; Stefan Maderwald; Harald H Quick; Benedikt A Poser
Journal:  MAGMA       Date:  2017-07-10       Impact factor: 2.310

Review 8.  Multi-echo acquisition.

Authors:  Stefan Posse
Journal:  Neuroimage       Date:  2011-10-25       Impact factor: 6.556

9.  Distortion and signal loss in medial temporal lobe.

Authors:  Cheryl A Olman; Lila Davachi; Souheil Inati
Journal:  PLoS One       Date:  2009-12-03       Impact factor: 3.240

10.  Dual-echo Z-shimmed proton resonance frequency-shift magnetic resonance thermometry near metallic ablation probes: Technique and temperature precision.

Authors:  Yuxin Zhang; Megan E Poorman; William A Grissom
Journal:  Magn Reson Med       Date:  2017-02-10       Impact factor: 4.668