Yang Kyeung Cho1, Hwa Seok Chang, Tae Yoon La, Donghyun Ji, Hyunkyung Kim, Jin A Choi, Man Soo Kim.
Abstract
PURPOSE: We evaluated various preoperative anterior segment parameters measured with a Pentacam rotating Scheimpflug camera and compared them with those of conventional methods. We also evaluated the effect of different parameters on corneal endothelial cells after cataract surgery.Entities:
Keywords: Anterior segment parameters; Corneal endothelium; Pentacam
Mesh:
Year: 2010 PMID: 21052508 PMCID: PMC2955271 DOI: 10.3341/kjo.2010.24.5.284
Source DB: PubMed Journal: Korean J Ophthalmol ISSN: 1011-8942
Baseline patient data
LOCS III = Lens Opacities Classification System.
Fig. 1Measurement of pupil size a measured before capsulorrhexis (PupilCCC). PupilCCC was measured before capsulorrhexis using a blunt-ended, viscoelastics injection needle of known length.
Change of endothelial cell density, coefficient of variation and hexagonality over time
Preop=preoperative measurement; Pod 3D = measurement at postoperative 3 day; Pod 2M = measurement at postoperative 2 month.
*Paired t-test, compared with preoperative value at each follow-up.
Nuclear opacity measured with Pentacam distribution according to LOCS III classification
LOCS III = Lens Opacities Classification System. τb(Kendall's tau b) = 0.414, p = 0.000.
Univariate association of our seven variables with the significance values predicting percentage change of endothelial cell density, coefficient of variation and, hexagonality at postoperative day 3 (3D) and month 2 (2M)
Pupilpreop = pupil size as measured preoperatively; PupilCCC = pupil size a measured before capsulorrhexis; ACDpentacam = anterior chamber depth measured by Pentacam; ACV = anterior chamber volume; LTsono=lens thickness measured by ultrasound; Viscoelasticamount = amount of viscoelastic; Densitometrypentacam = lens opacity measured by Pentacam densitometry.
*p < 0.05, statistically significant.
Final stepwise, multiple regression model for predicting percentage change of endothelial cell density at postoperative day 3 (3D) and month 2 (2M)
Densitometrypentacam = lens opacity measured by Pentacam densitometry; Viscoelasticsamount = amount of viscoelastics; PupilCCC = pupil size as measured before capsulorrhexis; ACVΠ = anterior chamber volume.
*p < 0.05, statistically significant.
Fig. 2Correlation of anterior chamber depth (mm) by Pentacam (ACDpentacam) and ultrasound (ACDsono) (r = 0.822, p < 0.05).
Fig. 3Correlation between anterior chamber depth (ACD) measured by Pentacam (ACDpentacam) and anterior chamber volume (ACV) (r = 0.650, p = 0.000).
Fig. 4Correlation between anterior chamber volume (ACV) and injected amount of viscoelastics before capsulorrhexis (r = -0.021, p = 0.850).
Fig. 5Correlation between anterior chamber depth by Pentacam (ACDpentacam) and injected amount of viscoelastics before capsulorrhexis (r = 0.123, p = 0.253).