Experimental investigation of a moving averaging algorithm for motion perpendicular to the leaf travel direction in dynamic MLC target tracking.

Jai-Woong Yoon1, Amit Sawant, Yelin Suh, Byung-Chul Cho, Tae-Suk Suh, Paul Keall.   

Abstract

PURPOSE: In dynamic multileaf collimator (MLC) motion tracking with complex intensity-modulated radiation therapy (IMRT) fields, target motion perpendicular to the MLC leaf travel direction can cause beam holds, which increase beam delivery time by up to a factor of 4. As a means to balance delivery efficiency and accuracy, a moving average algorithm was incorporated into a dynamic MLC motion tracking system (i.e., moving average tracking) to account for target motion perpendicular to the MLC leaf travel direction. The experimental investigation of the moving average algorithm compared with real-time tracking and no compensation beam delivery is described.
METHODS: The properties of the moving average algorithm were measured and compared with those of real-time tracking (dynamic MLC motion tracking accounting for both target motion parallel and perpendicular to the leaf travel direction) and no compensation beam delivery. The algorithm was investigated using a synthetic motion trace with a baseline drift and four patient-measured 3D tumor motion traces representing regular and irregular motions with varying baseline drifts. Each motion trace was reproduced by a moving platform. The delivery efficiency, geometric accuracy, and dosimetric accuracy were evaluated for conformal, step-and-shoot IMRT, and dynamic sliding window IMRT treatment plans using the synthetic and patient motion traces. The dosimetric accuracy was quantified via a tgamma-test with a 3%/3 mm criterion.
RESULTS: The delivery efficiency ranged from 89 to 100% for moving average tracking, 26%-100% for real-time tracking, and 100% (by definition) for no compensation. The root-mean-square geometric error ranged from 3.2 to 4.0 mm for moving average tracking, 0.7-1.1 mm for real-time tracking, and 3.7-7.2 mm for no compensation. The percentage of dosimetric points failing the gamma-test ranged from 4 to 30% for moving average tracking, 0%-23% for real-time tracking, and 10%-47% for no compensation.
CONCLUSIONS: The delivery efficiency of moving average tracking was up to four times higher than that of real-time tracking and approached the efficiency of no compensation for all cases. The geometric accuracy and dosimetric accuracy of the moving average algorithm was between real-time tracking and no compensation, approximately half the percentage of dosimetric points failing the gamma-test compared with no compensation.

Entities:  

Mesh:

Year:  2011        PMID: 21858989      PMCID: PMC3133804          DOI: 10.1118/1.3590384

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  29 in total

1.  Motion adaptive x-ray therapy: a feasibility study.

Authors:  P J Keall; V R Kini; S S Vedam; R Mohan
Journal:  Phys Med Biol       Date:  2001-01       Impact factor: 3.609

Review 2.  Errors and margins in radiotherapy.

Authors:  Marcel van Herk
Journal:  Semin Radiat Oncol       Date:  2004-01       Impact factor: 5.934

3.  Effects of motion on the total dose distribution.

Authors:  Thomas Bortfeld; Steve B Jiang; Eike Rietzel
Journal:  Semin Radiat Oncol       Date:  2004-01       Impact factor: 5.934

4.  Respiratory gating for liver tumors: use in dose escalation.

Authors:  Raquel Wagman; Ellen Yorke; Eric Ford; Philippe Giraud; Gikas Mageras; Bruce Minsky; Kenneth Rosenzweig
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-03-01       Impact factor: 7.038

5.  Real-time motion-adaptive-optimization (MAO) in TomoTherapy.

Authors:  Weiguo Lu; Mingli Chen; Kenneth J Ruchala; Quan Chen; Katja M Langen; Patrick A Kupelian; Gustavo H Olivera
Journal:  Phys Med Biol       Date:  2009-06-23       Impact factor: 3.609

6.  A technique for the quantitative evaluation of dose distributions.

Authors:  D A Low; W B Harms; S Mutic; J A Purdy
Journal:  Med Phys       Date:  1998-05       Impact factor: 4.071

7.  Evaluation of respiratory movement during gated radiotherapy using film and electronic portal imaging.

Authors:  E C Ford; G S Mageras; E Yorke; K E Rosenzweig; R Wagman; C C Ling
Journal:  Int J Radiat Oncol Biol Phys       Date:  2002-02-01       Impact factor: 7.038

8.  Synchronized moving aperture radiation therapy (SMART): average tumour trajectory for lung patients.

Authors:  Toni Neicu; Hiroki Shirato; Yvette Seppenwoolde; Steve B Jiang
Journal:  Phys Med Biol       Date:  2003-03-07       Impact factor: 3.609

9.  Four-dimensional IMRT treatment planning using a DMLC motion-tracking algorithm.

Authors:  Yelin Suh; Amit Sawant; Raghu Venkat; Paul J Keall
Journal:  Phys Med Biol       Date:  2009-05-28       Impact factor: 3.609

10.  Respiration tracking in radiosurgery.

Authors:  Achim Schweikard; Hiroya Shiomi; John Adler
Journal:  Med Phys       Date:  2004-10       Impact factor: 4.071

View more
  3 in total

1.  Use of dMLC for implementation of dynamic respiratory-gated radiation therapy.

Authors:  Eric W Pepin; Huanmei Wu; Hiroki Shirato
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

2.  Predictive modeling of respiratory tumor motion for real-time prediction of baseline shifts.

Authors:  A Balasubramanian; R Shamsuddin; B Prabhakaran; A Sawant
Journal:  Phys Med Biol       Date:  2017-01-11       Impact factor: 3.609

3.  Development and performance evaluation of a high-speed multileaf collimator.

Authors:  Xiang Zhang; Peiqing Ye; Hui Zhang
Journal:  J Appl Clin Med Phys       Date:  2016-12-29       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.