BACKGROUND: Despite the importance of optical quality of an intraocular lens (IOL) on visual outcomes following cataract surgery, objective data on their optical quality are not readily available, and manufacturing standards are industry regulated. The star test is a classic test of optical quality based on examination of the Airy disc and expanded diffraction rings of a point source of light, used mainly for telescope and microscope objectives.
METHODS: A physical model eye cell allowed star testing of IOLs under conditions similar to the optical environment in which they operate. 18 IOLs were tested and results compared to actual images produced by these lenses in the model eye cell. Quantitative measures of star testing performance were developed.
RESULTS: The optical performance of the IOLs varied, some performing very poorly. Most lenses (13/17) performed better in reverse orientation, while aberrations induced by the haptics of foldable IOLs were also detected. There was excellent correlation between actual images formed and star testing parameters.
CONCLUSION: Star testing IOLs was a novel biomedical application of a centuries old, inexpensive method. A concerning variation of optical quality was found, suggesting IOL optical performance data should be more readily available. Independent, authority mandated IOL optical quality standards should be developed, and results readily available to ophthalmologists.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16622088      PMCID: PMC1857027          DOI: 10.1136/bjo.2005.082701

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  23 in total

1.  High-order aberrations with Hydroview H60M and AcrySof MA30BA intraocular lenses: comparative study.

Authors:  Futoshi Taketani; Toyoaki Matsuura; Eiichi Yukawa; Yoshiaki Hara
Journal:  J Cataract Refract Surg       Date:  2004-04       Impact factor: 3.351

2.  Optical performance of 3 intraocular lens designs in the presence of decentration.

Authors:  Griffith E Altmann; Louis D Nichamin; Stephen S Lane; Jay S Pepose
Journal:  J Cataract Refract Surg       Date:  2005-03       Impact factor: 3.351

3.  Optical performance of decentered monofocal intraocular lenses.

Authors:  R W Faaland; L W Grossman
Journal:  J Cataract Refract Surg       Date:  1992-05       Impact factor: 3.351

4.  Determining the imaging quality of intraocular lenses.

Authors:  N E Norrby; L W Grossman; E P Geraghty; C F Kreiner; M Mihori; A S Patel; V Portney; D M Silberman
Journal:  J Cataract Refract Surg       Date:  1998-05       Impact factor: 3.351

5.  Influence of intraocular lens optical design on high-order aberrations.

Authors:  Futoshi Taketani; Eiichi Yukawa; Toshiaki Yoshii; Yuuko Sugie; Yoshiaki Hara
Journal:  J Cataract Refract Surg       Date:  2005-05       Impact factor: 3.351

6.  Low contrast visual acuity in pseudophakic patients implanted with an anterior surface modified prolate intraocular lens.

Authors:  Federico Ricci; Gianluca Scuderi; Filippo Missiroli; Federico Regine; Angelica Cerulli
Journal:  Acta Ophthalmol Scand       Date:  2004-12

7.  Visual acuity and contrast sensitivity comparison between Tecnis and AcrySof SA60AT intraocular lenses: A multicenter randomized study.

Authors:  Roberto Bellucci; Antonio Scialdone; Lucio Buratto; Simonetta Morselli; Chiara Chierego; Alessandra Criscuoli; Giuseppe Moretti; Patricia Piers
Journal:  J Cataract Refract Surg       Date:  2005-04       Impact factor: 3.351

8.  Effect of cataract surgery incision location and intraocular lens type on ocular aberrations.

Authors:  Konrad Pesudovs; Holger Dietze; Owen G Stewart; Bruce A Noble; Michael J Cox
Journal:  J Cataract Refract Surg       Date:  2005-04       Impact factor: 3.351

9.  Improved functional vision with a modified prolate intraocular lens.

Authors:  Mark Packer; I Howard Fine; Richard S Hoffman; Patricia A Piers
Journal:  J Cataract Refract Surg       Date:  2004-05       Impact factor: 3.351

10.  Comparison of wavefront aberrations and optical quality of eyes implanted with five different intraocular lenses.

Authors:  Roberto Bellucci; Simonetta Morselli; Patricia Piers
Journal:  J Refract Surg       Date:  2004 Jul-Aug       Impact factor: 3.573

View more
  1 in total

1.  Intraocular lens dystrophic calcification after trans-scleral diode laser treatment for a cyclodialysis cleft.

Authors:  Anmar Abdul-Rahman; Philip House; Josephine Richards
Journal:  Am J Ophthalmol Case Rep       Date:  2018-06-20
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.