Literature DB >> 15811748

Optical performance of 3 intraocular lens designs in the presence of decentration.

Griffith E Altmann1, Louis D Nichamin, Stephen S Lane, Jay S Pepose.   

Abstract

PURPOSE: To study the theoretical optical performance of 3 intraocular lens (IOL) designs in the presence of IOL decentration.
SETTING: Optics Center, Bausch & Lomb, Rochester, New York, USA.
METHODS: A ray-tracing program was used to evaluate the effect of IOL decentration on the optical performance of 3 silicone IOLs (LI61U, Bausch & Lomb; Tecnis Z9000, Advanced Medical Optics; and a new aberration-free IOL [SofPort AO, Bausch & Lomb]) in an experimental model eye. The study was done using pupil diameters of 3.0 mm, 4.0 mm, and 5.0 mm and IOL decentrations of 0 mm, 0.25 mm, 0.50 mm, 0.75 mm, and 1.00 mm. The modulation transfer functions were computed and plotted. A Monte Carlo simulation analysis with 1000 trials with IOL decentration randomly varying for each pupil size was performed.
RESULTS: Decentration of LI61U and Tecnis Z9000 IOLs led to asymmetrical higher-order aberrations that adversely affected the optical performance of the model eye; performance was not affected with the aberration-free IOL because it lacks inherent spherical aberration. Optical performance with the aberration-free IOL was better than with the LI61U IOL as the former has less spherical aberration and did not introduce other aberrations when decentered. Performance with the aberration-free IOL was better than with the Tecnis Z9000 IOL for 3.0 mm, 4.0 mm, and 5.0 mm pupils when decentration exceeded 0.15 mm, 0.30 mm, and 0.38 mm, respectively. Performance with the LI61U IOL was better than with the Tecnis Z9000 IOL for 3.0 mm, 4.0 mm, and 5.0 mm pupils when decentration exceeded 0.3 mm, 0.5 mm, and 0.5 mm, respectively. Monte Carlo simulations showed the expected postoperative results of the LI61U IOL and aberration-free IOL would be repeatable and predictable, whereas the outcomes with the Tecnis Z9000 IOL would vary widely.
CONCLUSIONS: The optical performance of the model eye was not affected by decentration of an aspheric IOL designed to have no inherent spherical aberration. With decentration, the performance with the new IOL was better than with a conventional spherical IOL and an aspheric IOL designed to offset the spherical aberration of an average cornea.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15811748     DOI: 10.1016/j.jcrs.2004.09.024

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  27 in total

1.  [Investigation of the theoretical image quality of aspheric intraocular lenses by decentration. Hoya AF-1 iMics1 und Zeiss ASPHINA(TM) (Invent ZO)].

Authors:  M Gillner; A Langenbucher; T Eppig
Journal:  Ophthalmologe       Date:  2012-03       Impact factor: 1.059

2.  Optical performance of toric intraocular lenses in the presence of decentration.

Authors:  Bin Zhang; Jin-Xue Ma; Dan-Yan Liu; Ying-Hua Du; Cong-Rong Guo; Yue-Xian Cui
Journal:  Int J Ophthalmol       Date:  2015-08-18       Impact factor: 1.779

3.  Star testing: a novel evaluation of intraocular lens optical quality.

Authors:  L Mitchell; A C B Molteno; T H Bevin; G Sanderson
Journal:  Br J Ophthalmol       Date:  2006-05       Impact factor: 4.638

4.  Light-adjustable lens: customizing correction for multifocality and higher-order aberrations.

Authors:  Christian A Sandstedt; Shiao H Chang; Robert H Grubbs; Daniel M Schwartz
Journal:  Trans Am Ophthalmol Soc       Date:  2006

5.  Comparison of contrast sensitivity, depth of field and ocular wavefront aberrations in eyes with an IOL with zero versus positive spherical aberration.

Authors:  Jay S Pepose; Mujtaba A Qazi; Keith H Edwards; Jeff P Sanderson; Edwin J Sarver
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-03-11       Impact factor: 3.117

6.  Custom optimization of intraocular lens asphericity.

Authors:  Douglas D Koch; Li Wang
Journal:  Trans Am Ophthalmol Soc       Date:  2007

7.  Optical quality of aspheric toric intraocular lenses at different degrees of decentering.

Authors:  Cari Pérez-Vives; Teresa Ferrer-Blasco; David Madrid-Costa; Santiago García-Lázaro; Robert Montés-Micó
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-04-22       Impact factor: 3.117

8.  Comparative study of visual function and ocular aberrations of two different one-piece designed hydrophilic acrylic intraocular lens.

Authors:  Yuxin Zhao; Zhaoxia Wang; Xia Tian; Xuehong Wang; Xining Gao
Journal:  Int Ophthalmol       Date:  2017-05-30       Impact factor: 2.031

9.  Pseudolentogenic astigmatic effect of multifocal intraocular lenses: non-corneal ocular residual astigmatism (N-CORA) as a new parameter in astigmatic change analysis.

Authors:  Andreas Frings; Johannes Steinberg; Vasyl Druchkiv; Stephan J Linke; Toam Katz
Journal:  Int Ophthalmol       Date:  2016-09-24       Impact factor: 2.031

10.  Ocular residual astigmatism (ORA) in pre-cataract eyes prior to and after refractive lens exchange.

Authors:  Toam Katz; Johannes Steinberg; Vasyl Druchkiv; Stephan J Linke; Andreas Frings
Journal:  Int Ophthalmol       Date:  2016-09-13       Impact factor: 2.031

View more