Literature DB >> 1593435

Optical performance of decentered monofocal intraocular lenses.

R W Faaland1, L W Grossman.   

Abstract

A model eye water cell was used to evaluate the optical performance of biconvex, meniscus, and plano-convex (spheric and aspheric) monofocal poly(methyl methacrylate) intraocular lenses when the lenses were centered and when they were decentered 1 mm and 2 mm. Resolution, induced astigmatism, and modulation transfer function measurements were performed for all lenses with the more convex surface of the lens oriented toward the incident light. The same measurements were performed for the plano-convex and meniscus lenses in the reverse orientation. The lens shapes least affected by decentration were the biconvex and spheric plano-convex with the convex surface oriented toward the incident light. When centered, the aspheric plano-convex lenses had the best overall contrast performance based on the modulation transfer function measurements. However, once decentered the performance of the aspheric lenses approached that of the meniscus lenses in the reverse orientation, the lens shape which had the worst performance.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1593435     DOI: 10.1016/s0886-3350(13)80906-6

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  1 in total

1.  Star testing: a novel evaluation of intraocular lens optical quality.

Authors:  L Mitchell; A C B Molteno; T H Bevin; G Sanderson
Journal:  Br J Ophthalmol       Date:  2006-05       Impact factor: 4.638

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.