Correlation between cardiac output measured by the femoral arterial thermodilution technique pulmonary arterial and that measured by contour pulse analysis in a paediatric animal model.

Jesús López-Herce1, Marta Rupérez, César Sánchez, Cristina García, Elena García.   

Abstract

OBJECTIVE: To evaluate the correlation between two methods for the determination of cardiac output: the femoral arterial thermodilution technique (FATD) and the arterial pulse contour analysis (PCCO) using the PiCCO catheter.
METHODS: We performed a prospective animal study using 51 immature Maryland pigs weighing 9 to 16 kg. A 4- or 5-Fr arterial PiCCO catheter was introduced into the femoral artery. In each animal, we made measurements of cardiac output at 30-minute intervals by femoral arterial thermodilution for a total of 209 measurements. We registered the previous PCCO and compared with the mean of two measures of FATD cardiac output.
RESULTS: Mean FATD was 1.73 +/- 0.60 L/min and mean PCCO was 1.78 +/- 0.70 L/min (no significant difference). The mean difference (bias) of differences (limits of agreement) was 0.04. The correlation founded between the two methods was 0.786 (95% confidence interval, 0.71-0.83).
CONCLUSIONS: Femoral arterial thermodilution cardiac output measurements correlates well with pulse contour analysis cardiac output in this paediatric animal model.

Entities:  

Mesh:

Year:  2006        PMID: 16508822     DOI: 10.1007/s10877-006-2229-5

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  18 in total

1.  Accuracy of beat-to-beat cardiac output monitoring by pulse contour analysis in hemodynamical unstable patients.

Authors:  O Gödje; R Friedl; A Hannekum
Journal:  Med Sci Monit       Date:  2001 Nov-Dec

2.  Comparison between cardiac output measured by the pulmonary arterial thermodilution technique and that measured by the femoral arterial thermodilution technique in a pediatric animal model.

Authors:  M Rupérez; J López-Herce; C García; C Sánchez; E García; D Vigil
Journal:  Pediatr Cardiol       Date:  2003-12-23       Impact factor: 1.655

3.  Assessment of cardiac output, intravascular volume status, and extravascular lung water by transpulmonary indicator dilution in critically ill neonates and infants.

Authors:  Holger Schiffmann; Bernhard Erdlenbruch; Dominique Singer; Sybille Singer; Egbert Herting; Andreas Hoeft; Wolfgang Buhre
Journal:  J Cardiothorac Vasc Anesth       Date:  2002-10       Impact factor: 2.628

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Intraoperative cardiac output monitoring: comparison of impedance cardiography and thermodilution.

Authors:  A C Perrino; A Lippman; C Ariyan; T Z O'Connor; M Luther
Journal:  J Cardiothorac Vasc Anesth       Date:  1994-02       Impact factor: 2.628

6.  Comparison of cardiac output assessed by pulse-contour analysis and thermodilution in patients undergoing minimally invasive direct coronary artery bypass grafting.

Authors:  W Buhre; A Weyland; S Kazmaier; G G Hanekop; M M Baryalei; M Sydow; H Sonntag
Journal:  J Cardiothorac Vasc Anesth       Date:  1999-08       Impact factor: 2.628

7.  Pulse contour analysis for cardiac output monitoring in cardiac surgery for congenital heart disease.

Authors:  Aman Mahajan; Afshin Shabanie; Judi Turner; Michael J Sopher; Jure Marijic
Journal:  Anesth Analg       Date:  2003-11       Impact factor: 5.108

8.  Cardiac output measurement by transpulmonary versus conventional thermodilution technique in intensive care patients after coronary artery bypass grafting.

Authors:  R Gust; A Gottschalk; H Bauer; B W Böttiger; H Böhrer; E Martin
Journal:  J Cardiothorac Vasc Anesth       Date:  1998-10       Impact factor: 2.628

9.  Comparison of continuous cardiac output measurements in patients after cardiac surgery.

Authors:  Frank Mielck; Wolfgang Buhre; Gunnar Hanekop; Theodor Tirilomis; Reinhard Hilgers; Hans Sonntag
Journal:  J Cardiothorac Vasc Anesth       Date:  2003-04       Impact factor: 2.628

Review 10.  Continuous measurement of cardiac output by the Fick principle in infants and children: comparison with the thermodilution method.

Authors:  C F Wippermann; R G Huth; F X Schmidt; J Thul; M Betancor; D Schranz
Journal:  Intensive Care Med       Date:  1996-05       Impact factor: 17.440

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.