Cardiac output measurement by transpulmonary versus conventional thermodilution technique in intensive care patients after coronary artery bypass grafting.

R Gust1, A Gottschalk, H Bauer, B W Böttiger, H Böhrer, E Martin.   

Abstract

OBJECTIVE: The aim of the present study was to evaluate the correlation, accuracy, and precision of transpulmonary thermodilution cardiac output (CO) measurement. For this purpose, this technique was compared with the clinical gold standard, the CO measurement by pulmonary artery catheter in patients after coronary artery bypass grafting (CABG).
DESIGN: A prospective clinical study.
SETTING: A university medical center. PARTICIPANTS: Seventy-five patients in an intensive care unit (ICU) after CABG.
INTERVENTIONS: Standard (SCO) and transpulmonary thermodilution CO measurement (TPCO) measurements were simultaneously performed in triplicate by central venous injection of cooled saline solution. All variables were recorded at five different time points of measurement during weaning from mechanical ventilation.
MEASUREMENTS AND MAIN RESULTS: CO measurements yielded 375 data pairs. SCO ranged from 2.0 to 10.2 L/min, and TPCO from 1.3 to 10.6 L/min. During the entire observation period, TPCO measurements tended to yield relatively high values, whereas SCO measurements resulted in lower values. Correlation between TPCO and SCO measurements was significant (r = 0.73; p < 0.05), accompanied by an accuracy with a bias of 0.456 L/min (7.3%) and a precision of 1.156 L/min (18.5%).
CONCLUSION: In most patients, TPCO measurement will not replace the conventional technique by pulmonary artery catheter, but in some patients it offers an attractive, reliable, and safe method to determine CO.

Entities:  

Mesh:

Year:  1998        PMID: 9801970     DOI: 10.1016/s1053-0770(98)90093-3

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  6 in total

1.  Comparison between cardiac output measured by the pulmonary arterial thermodilution technique and that measured by the femoral arterial thermodilution technique in a pediatric animal model.

Authors:  M Rupérez; J López-Herce; C García; C Sánchez; E García; D Vigil
Journal:  Pediatr Cardiol       Date:  2003-12-23       Impact factor: 1.655

2.  Methods of monitoring shock.

Authors:  Ednan K Bajwa; Atul Malhotra; B Taylor Thompson
Journal:  Semin Respir Crit Care Med       Date:  2004-12       Impact factor: 3.119

3.  Correlation between cardiac output measured by the femoral arterial thermodilution technique pulmonary arterial and that measured by contour pulse analysis in a paediatric animal model.

Authors:  Jesús López-Herce; Marta Rupérez; César Sánchez; Cristina García; Elena García
Journal:  J Clin Monit Comput       Date:  2006-02-28       Impact factor: 2.502

4.  Accuracy of a novel approach to measuring arterial thermodilution cardiac output during intra-aortic counterpulsation.

Authors:  Werner Baulig; Philipp Schuett; Oliver Goedje; Edith R Schmid
Journal:  J Clin Monit Comput       Date:  2007-03-17       Impact factor: 1.977

5.  Transpulmonary thermodilution-derived cardiac function index identifies cardiac dysfunction in acute heart failure and septic patients: an observational study.

Authors:  Simon Ritter; Alain Rudiger; Marco Maggiorini
Journal:  Crit Care       Date:  2009-08-11       Impact factor: 9.097

6.  Noninvasive oscillometric cardiac output determination in the intensive care unit - comparison with invasive transpulmonary thermodilution.

Authors:  Alexander Reshetnik; Friederike Compton; Anna Schölzel; Markus Tölle; Walter Zidek; Markus van der Giet
Journal:  Sci Rep       Date:  2017-08-30       Impact factor: 4.379

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.