| Literature DB >> 16214009 |
Keith R Berend1, Adolph V Lombardi, Thomas H Mallory, Joanne B Adams, Jackie H Russell, Kari L Groseth.
Abstract
Constrained acetabular components can treat or prevent instability after total hip arthroplasty (THA). We examine long-term results of 755 consecutive constrained THA in 720 patients (1986-1993; 62 primary, 59 conversion, 565 revision, 60 reimplantation, and 9 total femur). Eighty-three patients (88 THAs) were lost before 10-year follow-up, leaving 639 patients (667 THAs) available for study. Dislocation occurred in 117 hips (17.5%), in 37 (28.9%) of 128 constrained for recurrent dislocation, and 46 (28.2%) of 163 with dislocation history. Other reoperations were for aseptic loosening (51, 7.6% acetabular; 28, 4.2% stem; 16, 2.4% combined), infection (40, 6.0%), periprosthetic fracture (19, 2.8%), stem breakage (2, 0.3%), cup malposition (1, 0.1%), dissociated insert (1, 0.1%), dissociated femoral head (1, 0.1%), and impingement of 1 broken (0.1%) and 4 (0.6%) dissociated constraining rings. Although constrained acetabular components prevented recurrent dislocation in 71.1%, they should be used cautiously, with a 42.1% long-term failure rate observed in this series. Dislocation was common despite constraint with previous history as a significant risk.Entities:
Mesh:
Year: 2005 PMID: 16214009 DOI: 10.1016/j.arth.2005.06.001
Source DB: PubMed Journal: J Arthroplasty ISSN: 0883-5403 Impact factor: 4.757