Speech perception with mono- and quadrupolar electrode configurations: a crossover study.

Lucas H M Mens1, Carlo K Berenstein.   

Abstract

OBJECTIVE: To study the effect of two multipolar electrode configurations on speech perception, pitch perception, and the intracochlear electrical field. STUDY
DESIGN: Crossover design; within subject.
SETTING: Tertiary referral center. PATIENTS: Eight experienced adult cochlear implant users. INTERVENTION: Each subject used each of three experimental processors for 3 weeks. The following processors were compared that differed only in electrode configuration: 1) monopolar; 2) hybrid quadrupolar, in which half of the current returned to the extracochlear reference electrode and half to two electrodes immediately to the left and right of the active electrode; and 3) flat tripolar +2, which directed all the current to four reference electrodes (two on each side), separated from the active electrode by two inactive electrodes. All the processors used the standard Advanced Bionics HiRes speech-processing strategy, 12 channels, 1,220 pulses per second per channel, and with a pulse width of 33 (micros/phase).
RESULTS: The monopolar processors had the largest stimulation efficiency and the smallest dynamic range in linear current units. The reverse was true of flat tripolar +2 processor, whereas the hybrid quadrupolar processor fell in between. Insufficient loudness growth prevented the use of the flat tripolar +2 processor in three subjects. Word recognition did not differ between the clinically used 16-channel monopolar processor and the experimental monopolar processor, regardless of the differences in the number of channels, pulse rate, and duration of experience. Word recognition with the flat tripolar +2 processor was significantly poorer than with the monopolar and hybrid quadrupolar processors; monopolar and quadrupolar processors did not differ. There was no significant interaction between processor type and competing noise type (stationary or fluctuating), but performance at the higher level of fluctuating noise was best with the hybrid quadrupolar processor in almost all the subjects. Pitch scaling showed ceiling performance in five subjects and differed between processors in the two other subjects with imperfect tonotopy. Intracochlear current spread was considerable with the monopolar configuration; it was reduced with the hybrid quadrupolar configuration and virtually absent beyond the active electrodes with the tripolar configuration.
CONCLUSION: More confined configurations reduced the longitudinal width of the electrical field, which was expected to enhance channel separation, but no improvement in word recognition was found. More research is needed to test confined configurations that have enhanced efficiency and to evaluate the fundamental effects of configuration on channel discriminability.

Entities:  

Mesh:

Year:  2005        PMID: 16151343     DOI: 10.1097/01.mao.0000185060.74339.9d

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  40 in total

Review 1.  Probing the electrode-neuron interface with focused cochlear implant stimulation.

Authors:  Julie Arenberg Bierer
Journal:  Trends Amplif       Date:  2010-06

2.  Partial tripolar cochlear implant stimulation: Spread of excitation and forward masking in the inferior colliculus.

Authors:  Julie Arenberg Bierer; Steven M Bierer; John C Middlebrooks
Journal:  Hear Res       Date:  2010-08-18       Impact factor: 3.208

3.  Excitation Patterns of Standard and Steered Partial Tripolar Stimuli in Cochlear Implants.

Authors:  Ching-Chih Wu; Xin Luo
Journal:  J Assoc Res Otolaryngol       Date:  2015-12-21

4.  Assessment of electrode placement and audiological outcomes in bilateral cochlear implantation.

Authors:  George B Wanna; Jack H Noble; Theodore R McRackan; Benoit M Dawant; Mary S Dietrich; Linsey D Watkins; Alejandro Rivas; Theodore A Schuman; Robert F Labadie
Journal:  Otol Neurotol       Date:  2011-04       Impact factor: 2.311

5.  Psychophysical versus physiological spatial forward masking and the relation to speech perception in cochlear implants.

Authors:  Michelle L Hughes; Lisa J Stille
Journal:  Ear Hear       Date:  2008-06       Impact factor: 3.570

6.  Current focusing and steering: modeling, physiology, and psychophysics.

Authors:  Ben H Bonham; Leonid M Litvak
Journal:  Hear Res       Date:  2008-04-06       Impact factor: 3.208

7.  Impact of electrode design and surgical approach on scalar location and cochlear implant outcomes.

Authors:  George B Wanna; Jack H Noble; Matthew L Carlson; René H Gifford; Mary S Dietrich; David S Haynes; Benoit M Dawant; Robert F Labadie
Journal:  Laryngoscope       Date:  2014-05-30       Impact factor: 3.325

8.  Current steering with partial tripolar stimulation mode in cochlear implants.

Authors:  Ching-Chih Wu; Xin Luo
Journal:  J Assoc Res Otolaryngol       Date:  2012-12-19

9.  Role of electrode placement as a contributor to variability in cochlear implant outcomes.

Authors:  Charles C Finley; Timothy A Holden; Laura K Holden; Bruce R Whiting; Richard A Chole; Gail J Neely; Timothy E Hullar; Margaret W Skinner
Journal:  Otol Neurotol       Date:  2008-10       Impact factor: 2.311

10.  An attempt to improve bilateral cochlear implants by increasing the distance between electrodes and providing complementary information to the two ears.

Authors:  Richard S Tyler; Shelley A Witt; Camille C Dunn; Ann Perreau; Aaron J Parkinson; Blake S Wilson
Journal:  J Am Acad Audiol       Date:  2010-01       Impact factor: 1.664

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.