An analysis of an implantable dosimeter system for external beam therapy.

Robert D Black1, Charles W Scarantino, Gregory G Mann, Mitchell S Anscher, Robert D Ornitz, Benjamin E Nelms.   

Abstract

BACKGROUND AND
PURPOSE: To review the data from an implantable radiation dosimetry system used in a clinical setting and to examine correlations between dosimeter readings and potential causative error sources.
MATERIALS AND METHODS: MOSFET (metal oxide semiconductor field effect transistor) based encapsulated dosimeters were evaluated in a phantom (in vitro) and in a study with 18 patients. The dosimeters were placed in the gross tumor volume or in collateral normal tissue. Predicted dose values were established by imaging the dosimeters in the planning CTs.
RESULTS: The in vitro study confirmed that bounding cumulative errors due to setup, planning, and machine output within a +/-5% level is achievable. In patients, it was found that deviations from the targeted dose often exceeded the 5% level.
CONCLUSIONS: The use of an implantable dosimeter system could provide an effective empiric check on the dose delivered at depth. Such a tool may have value for institutional quality assurance, as well as for therapy delivered to individual patients.

Entities:  

Mesh:

Year:  2005        PMID: 16111599      PMCID: PMC2543134          DOI: 10.1016/j.ijrobp.2005.05.025

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  27 in total

1.  Evaluation of ultrasound-based prostate localization for image-guided radiotherapy.

Authors:  K M Langen; J Pouliot; C Anezinos; M Aubin; A R Gottschalk; I-C Hsu; D Lowther; Y-M Liu; K Shinohara; L J Verhey; V Weinberg; M Roach
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-11-01       Impact factor: 7.038

2.  It is necessary to validate each individual IMRT treatment plan before delivery. For the proposition.

Authors:  Chester Ramsey
Journal:  Med Phys       Date:  2003-09       Impact factor: 4.071

3.  IMRT may be used to excess because of its higher reimbursement from Medicare. For the proposition.

Authors:  Bhudatt R Paliwal
Journal:  Med Phys       Date:  2004-01       Impact factor: 4.071

4.  Incorporating an improved dose-calculation algorithm in conformal radiotherapy of lung cancer: re-evaluation of dose in normal lung tissue.

Authors:  Katrien De Jaeger; Mischa S Hoogeman; Martijn Engelsman; Yvette Seppenwoolde; Eugène M F Damen; Ben J Mijnheer; Liesbeth J Boersma; Joos V Lebesque
Journal:  Radiother Oncol       Date:  2003-10       Impact factor: 6.280

5.  The current NRC definitions of therapy misadministration are vague, do not reflect the norms of clinical practice, and should be rewritten. For the proposition.

Authors:  Howard Amols
Journal:  Med Phys       Date:  2004-04       Impact factor: 4.071

6.  Synergistic effects of hyperoxic gas breathing and reduced oxygen consumption on tumor oxygenation: a theoretical model.

Authors:  Timothy W Secomb; Richard Hsu; Mark W Dewhirst
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-06-01       Impact factor: 7.038

7.  Implementing IMRT in clinical practice: a joint document of the American Society for Therapeutic Radiology and Oncology and the American Association of Physicists in Medicine.

Authors:  James M Galvin; Gary Ezzell; Avraham Eisbrauch; Cedric Yu; Brian Butler; Ying Xiao; Isaac Rosen; Julian Rosenman; Michael Sharpe; Lei Xing; Ping Xia; Tony Lomax; Daniel A Low; Jatinder Palta
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-04-01       Impact factor: 7.038

8.  Guidance document on delivery, treatment planning, and clinical implementation of IMRT: report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee.

Authors:  Gary A Ezzell; James M Galvin; Daniel Low; Jatinder R Palta; Isaac Rosen; Michael B Sharpe; Ping Xia; Ying Xiao; Lei Xing; Cedric X Yu
Journal:  Med Phys       Date:  2003-08       Impact factor: 4.071

9.  Independent verification of ultrasound based image-guided radiation treatment, using electronic portal imaging and implanted gold markers.

Authors:  Frank Van den Heuvel; Tanya Powell; Edward Seppi; Peter Littrupp; Mubashra Khan; Yue Wang; Jeffrey D Forman
Journal:  Med Phys       Date:  2003-11       Impact factor: 4.071

10.  Treatment planning for lung cancer: traditional homogeneous point-dose prescription compared with heterogeneity-corrected dose-volume prescription.

Authors:  Steven J Frank; Kenneth M Forster; Craig W Stevens; James D Cox; Ritsuko Komaki; Zhongxing Liao; Susan Tucker; Xuemei Wang; Roy E Steadham; Chris Brooks; George Starkschall
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-08-01       Impact factor: 7.038

View more
  8 in total

1.  Investigation of an implantable dosimeter for single-point water equivalent path length verification in proton therapy.

Authors:  Hsiao-Ming Lu; Greg Mann; Ethan Cascio
Journal:  Med Phys       Date:  2010-11       Impact factor: 4.071

2.  The observed variance between predicted and measured radiation dose in breast and prostate patients utilizing an in vivo dosimeter.

Authors:  Charles W Scarantino; Bradley R Prestidge; Mitchel S Anscher; Carolyn R Ferree; William T Kearns; Robert D Black; Natasha G Bolick; Gloria P Beyer
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-10-01       Impact factor: 7.038

3.  Radioluminescence characterization of in situ x-ray nanodosimeters: Potential real-time monitors and modulators of external beam radiation therapy.

Authors:  Jeffrey S Souris; Shih-Hsun Cheng; Charles Pelizzari; Nai-Tzu Chen; Patrick La Riviere; Chin-Tu Chen; Leu-Wei Lo
Journal:  Appl Phys Lett       Date:  2014-11-20       Impact factor: 3.791

4.  Uncertainties encountered in implementation of adaptive planning with in vivo dosimeters.

Authors:  M T Studenski; S J Gardner; R B Den
Journal:  Radiol Phys Technol       Date:  2014-09-19

5.  In vivo real-time rectal wall dosimetry for prostate radiotherapy.

Authors:  Nicholas Hardcastle; Dean L Cutajar; Peter E Metcalfe; Michael L F Lerch; Vladimir L Perevertaylo; Wolfgang A Tomé; Anatoly B Rosenfeld
Journal:  Phys Med Biol       Date:  2010-07-07       Impact factor: 3.609

6.  A study on rectal dose measurement in phantom and in vivo using Gafchromic EBT3 film in IMRT and CyberKnife treatments of carcinoma of prostate.

Authors:  K Ganapathy; P G G Kurup; V Murali; M Muthukumaran; S Balaji Subramanian; J Velmurugan
Journal:  J Med Phys       Date:  2013-07

7.  Investigation into the use of a MOSFET dosimeter as an implantable fiducial marker.

Authors:  Stephen F Kry; Michael Price; Zhonglu Wang; Firas Mourtada; Mohammad Salehpour
Journal:  J Appl Clin Med Phys       Date:  2009-01-27       Impact factor: 2.102

8.  Commissioning and implementation of an implantable dosimeter for radiation therapy.

Authors:  Ivan Buzurovic; Timothy N Showalter; Matthew T Studenski; Robert B Den; Adam P Dicker; Junsheng Cao; Ying Xiao; Yan Yu; Amy Harrison
Journal:  J Appl Clin Med Phys       Date:  2013-03-04       Impact factor: 2.102

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.