The calibration of experimental self-developing Gafchromic HXR film for the measurement of radiation dose in computed tomography.

K R Gorny1, S L Leitzen, M R Bruesewitz, J M Kofler, N J Hangiandreou, C H McCollough.   

Abstract

A prototype, self-developing Gafchromic HXR film has sensitivity an order of magnitude larger than that of the commercially available Gafchromic XR film used in interventional radiological applications. The higher sensitivity of the HXR film allows the possibility of acquisition of high-resolution calibrated dose profiles within the diagnostic range of exposure levels, below 10 R (87.7 mGy). We employed a commercially available, optical flatbed scanner for digitization of the film and image analysis software to determine the response of the HXR films to ionizing radiation. Spatial uniformity and temporal repeatability of the flatbed scanner were determined and used in optimization of the digitization protocol. The HXR film postexposure density growth and sensitivity to ambient light were determined using multiple scans of two simultaneously exposed sheets, one stored in light-tight conditions and the other continuously illuminated with white light. A calibrated step wedge of the HXR film was obtained by simultaneous irradiation of a portion of a film strip and a calibrated ionization chamber using a radiographic x-ray tube with beam characteristics matched to a typical CT scanner (8 mm Al HVL, 120 kVp). Repeated digitization of the calibration film was used to determine the precision of the film response measurements. The precision, as measured by the standard deviation of multiple measurements, was better than 1% over the full dynamic range of film response. This precision was measured using exposures ranging from 0.5 to 12 R (4.4 to 105.3 mGy). This exposure range is highly relevant to x-ray computed tomography. Preliminary radiation dose profiles demonstrate the utility of this technique.

Entities:  

Mesh:

Year:  2005        PMID: 15895584     DOI: 10.1118/1.1862802

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  Experimental estimates of peak skin dose and its relationship to the CT dose index using the CTDI head phantom.

Authors:  Hugo de las Heras; Ronaldo Minniti; Sean Wilson; Chad Mitchell; Marlene Skopec; Claudia C Brunner; Kish Chakrabarti
Journal:  Radiat Prot Dosimetry       Date:  2013-07-16       Impact factor: 0.972

2.  Signal-to-noise ratio and dose to the lens of the eye for computed tomography examination of the brain using an automatic tube current modulation system.

Authors:  Supawitoo Sookpeng; Chitsanupong Butdee
Journal:  Emerg Radiol       Date:  2016-12-02

3.  In vitro dose measurements in a human cadaver with abdomen/pelvis CT scans.

Authors:  Da Zhang; Atul Padole; Xinhua Li; Sarabjeet Singh; Ranish Deedar Ali Khawaja; Diego Lira; Tianyu Liu; Jim Q Shi; Alexi Otrakji; Mannudeep K Kalra; X George Xu; Bob Liu
Journal:  Med Phys       Date:  2014-09       Impact factor: 4.071

4.  A feasibility study using radiochromic films for fast neutron 2D passive dosimetry.

Authors:  Samuel L Brady; Rathnayaka Gunasingha; Terry T Yoshizumi; Calvin R Howell; Alexander S Crowell; Brent Fallin; Anton P Tonchev; Mark W Dewhirst
Journal:  Phys Med Biol       Date:  2010-08-06       Impact factor: 3.609

5.  Computed tomography dose assessment for a 160 mm wide, 320 detector row, cone beam CT scanner.

Authors:  J Geleijns; M Salvadó Artells; P W de Bruin; R Matter; Y Muramatsu; M F McNitt-Gray
Journal:  Phys Med Biol       Date:  2009-05-06       Impact factor: 3.609

6.  A survey of radiation dose to patients and operators during radiofrequency ablation using computed tomography.

Authors:  A Saidatul; Ca Azlan; Msa Megat Amin; Bjj Abdullah; Kh Ng
Journal:  Biomed Imaging Interv J       Date:  2010-01-01
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.