Signal-to-noise ratio and dose to the lens of the eye for computed tomography examination of the brain using an automatic tube current modulation system.

Supawitoo Sookpeng1, Chitsanupong Butdee2.   

Abstract

PURPOSE: The study aimed to evaluate the image quality in terms of signal-to-noise ratio (SNR) and dose to the lens of the eye and the other nearby organs from the CT brain scan using an automatic tube current modulation (ATCM) system with or without CT gantry tilt is needed.
METHODS: An anthropomorphic phantom was scanned with different settings including use of different ATCM, fixed tube current time product (mAs) settings and degree angles of gantry tilt. Gafchromic film XR-QA2 was used to measure absorbed dose of the organs. Relative doses and SNR for the various scan settings were compared with the reference setting of the fixed 330 mAs.
RESULTS: Average absorbed dose for the lens of the eyes varied from 8.7 to 21.7 mGy. The use of the ATCM system with the gantry tilt resulted in up to 60% decrease in the dose to the lens of the eye. SNR significantly decreased while tilting the gantry using the fixed mAs techniques, compared to that of the reference setting. However, there were no statistical significant differences for SNRs between the reference setting and all ATCM settings.
CONCLUSIONS: Compared to the reference setting of the fixed effective mAs, using the ATCM system and appropriate tilting, the gantry resulted in a substantial decrease in the dose to the lens of the eye while preserving signal-to-noise ratio. CT brain examination should be carefully controlled to optimize dose for lens of the eye and image quality of the examination.

Keywords:  Absorbed dose; CT brain; Eye lens; Tube current modulation

Mesh:

Year:  2016        PMID: 27909929     DOI: 10.1007/s10140-016-1470-6

Source DB:  PubMed          Journal:  Emerg Radiol        ISSN: 1070-3004


  22 in total

1.  Tomographic reconstruction for tilted helical multislice CT.

Authors:  J Hsieh
Journal:  IEEE Trans Med Imaging       Date:  2000-09       Impact factor: 10.048

2.  Radiation dose for routine clinical adult brain CT: Variability on different scanners at one institution.

Authors:  Tracy A Jaffe; Jenny K Hoang; Terry T Yoshizumi; Greta Toncheva; Carolyn Lowry; Carl Ravin
Journal:  AJR Am J Roentgenol       Date:  2010-08       Impact factor: 3.959

3.  The calibration of experimental self-developing Gafchromic HXR film for the measurement of radiation dose in computed tomography.

Authors:  K R Gorny; S L Leitzen; M R Bruesewitz; J M Kofler; N J Hangiandreou; C H McCollough
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

4.  Dose and energy dependence of response of Gafchromic XR-QA film for kilovoltage x-ray beams.

Authors:  O Rampado; E Garelli; S Deagostini; R Ropolo
Journal:  Phys Med Biol       Date:  2006-05-24       Impact factor: 3.609

Review 5.  Computed tomography--an increasing source of radiation exposure.

Authors:  David J Brenner; Eric J Hall
Journal:  N Engl J Med       Date:  2007-11-29       Impact factor: 91.245

6.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103.

Authors: 
Journal:  Ann ICRP       Date:  2007

Review 7.  Raise the bar and lower the dose: current and future strategies for radiation dose reduction in head and neck imaging.

Authors:  M Ibrahim; H Parmar; E Christodoulou; S Mukherji
Journal:  AJNR Am J Neuroradiol       Date:  2013-02-28       Impact factor: 3.825

8.  Peak skin and eye lens radiation dose from brain perfusion CT based on Monte Carlo simulation.

Authors:  Di Zhang; Chris H Cagnon; J Pablo Villablanca; Cynthia H McCollough; Dianna D Cody; Donna M Stevens; Maria Zankl; John J Demarco; Adam C Turner; Maryam Khatonabadi; Michael F McNitt-Gray
Journal:  AJR Am J Roentgenol       Date:  2012-02       Impact factor: 3.959

9.  Estimation of Eye Lens Dose During Brain Scans Using Gafchromic Xr-QA2 Film in Various Multidetector CT Scanners.

Authors:  Philomina Akhilesh; Arti R Kulkarni; Shramika H Jamhale; S D Sharma; Rajesh Kumar; D Datta
Journal:  Radiat Prot Dosimetry       Date:  2017-04-25       Impact factor: 0.972

10.  Rising use of diagnostic medical imaging in a large integrated health system.

Authors:  Rebecca Smith-Bindman; Diana L Miglioretti; Eric B Larson
Journal:  Health Aff (Millwood)       Date:  2008 Nov-Dec       Impact factor: 6.301

View more
  1 in total

1.  Dosimetric changes with computed tomography automatic tube-current modulation techniques.

Authors:  Sofia Spampinato; Anna Maria Gueli; Pietro Milone; Luigi Angelo Raffaele
Journal:  Radiol Phys Technol       Date:  2018-04-06
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.