Literature DB >> 21611060

A survey of radiation dose to patients and operators during radiofrequency ablation using computed tomography.

A Saidatul1, Ca Azlan, Msa Megat Amin, Bjj Abdullah, Kh Ng.   

Abstract

Computed tomography (CT) fluoroscopy is able to give real time images to a physician undertaking minimally invasive procedures such as biopsies, percutaneous drainage, and radio frequency ablation (RFA). Both operators executing the procedure and patients too, are thus at risk of radiation exposure during a CT fluoroscopy.This study focuses on the radiation exposure present during a series of radio frequency ablation (RFA) procedures, and used Gafchromic film (Type XR-QA; International Specialty Products, USA) and thermoluminescent dosimeters (TLD-100H; Bicron, USA) to measure the radiation received by patients undergoing treatment, and also operators subject to scatter radiation.The voltage was held constant at 120 kVp and the current 70mA, with 5mm thickness. The duration of irradiation was between 150-638 seconds.Ultimately, from a sample of 30 liver that have undergone RFA, the study revealed that the operator received the highest dose at the hands, which was followed by the eyes and thyroid, while secondary staff dosage was moderately uniform across all parts of the body that were measured.

Entities:  

Keywords:  Gafchromic film; computed tomography fluoroscopy; radiofrequency ablation; thermoluminescent dosimeter

Year:  2010        PMID: 21611060      PMCID: PMC3097790          DOI: 10.2349/biij.6.1.e2

Source DB:  PubMed          Journal:  Biomed Imaging Interv J        ISSN: 1823-5530


INTRODUCTION

While the field of view (FOV) is concentrated on a more limited area in CT fluoroscopy, conventional CT uses 150-170 mA, and CT fluoroscopy tube current is invariably reduced to 50 mA. Other dosimetry practices are similar to the usual spiral CT scanning, along with tube voltage (kVp) being at similar levels. These real-time 2-D images are constantly updated while the x-ray tube rotates around the subject. Dynamic axial imaging is provided, with real-time observation of image reconstruction possible. Given this advantage, it is an invaluable procedure for monitoring the passage of a biopsy needle, or the tube during drainage. A notable divergence in the two techniques is that CT fluoroscopy is an interactive process, and as such requires the presence of an operator at the location of the gantry. This necessarily invokes exposure to scatter radiation, while conventional CT scanning allows the operator to control the process remotely. A non-surgical procedure, such as radio frequency ablation (RFA), is a localised treatment that offers the patient freedom from side effects without interfering with their general health, and a rapid return to normal activity. RFA uses the device of radiofrequency energy to which is inserted into target tissue (invariably a tumour) via the tip of a needle through the patient’s skin. On contact with the target tissue heat is produced, which kills the target tissue while sparing tissue that is healthy. In the event of the tumour causing bone matter to decay; this will not regenerate however, the dead tissue of a tumour will reduce in size.

MATERIALS AND METHODS

The University of Malaya Medical Centre provided the opportunity to gather the data for this study from thirty RFA liver treatments. The patients underwent sedation and the staff involved included a radiologist and anaesthetist, medical officers, a radiographer, along with nurses and support staff. The same radiologist and support staff participated in all 30 monitored treatments. Tube voltage was constant at 120 kVp, tube current at 70 mA and slice thickness was 5mm.

Patient Dose

The use of film in dosimetry provides precise information in respect of the location of the dose to the skin, and with calibration of the film, it also allows for quantitative measurement of the dose. To this end, Gafchromic® type XR-QA film (5" x 6") was placed at the rear of the patient (Fig. 1), and the radiation received during treatment was measured.
Figure 1

Diagram showing how Gafchromic film is positioned at the patient’s back.

Diagram showing how Gafchromic film is positioned at the patient’s back. Afterward, the film was converted to a digital format using a Mikrotek ScanMaker 4800 flatbed scanner (Carson, USA) with ScanWizard5 Version 7 software. Measuring the extent of the blackened area on the digital format allowed identification of the radiation dose received, and was made possible by Adobe Photoshop version 6.0 (Adobe Systems, California, USA).

Operator Dose

Thermo luminescent Dosimeters (TLD’s) were used to measure the scatter radiation received by the operator and other staff. Each individual had a sachet containing three TLD chips attached at the forehead, thyroid and middle finger (Fig. 2), and the radiologist alone wore a thyroid shield. A TLD reader (Harshaw 3500, Bicron, Ohio) then determined the average of these.
Figure 2

The TLD sachets were taped at the forehead, thyroid and middle finger of the radiologist and supporting staff (a) Radiologist – the TLD sachet was taped under the thyroid shield (b) Supporting staff – TLD sachet was taped at the neck. Supporting staff did not wear thyroid shield.

The TLD sachets were taped at the forehead, thyroid and middle finger of the radiologist and supporting staff (a) Radiologist – the TLD sachet was taped under the thyroid shield (b) Supporting staff – TLD sachet was taped at the neck. Supporting staff did not wear thyroid shield.

RESULT AND DISCUSSION

Radiation exposure during an RFA procedure that causes blackening on the Gafchromic film is shown in Fig. 3. Due to the fact that the x-ray tube has a constant orbit around the body of the patient, the darker stripes signifying higher doses of radiation are found in the same central location. Other alternating light and dark stripes occur below and above the dark band at the centre due to scout and helical scanning that is used to locate the target area prior to the insertion of the radio frequency probe into the liver.
Figure 3

The gafchromic film image after being exposed.

The gafchromic film image after being exposed. The radiation dose received at the entry point on the skin is provided as a function of time in Table 1, which shows that the dosage received increases with time. While the highest dosage received was 326±2.12 µGy for the duration 9patient.
Table 1

Entrance dose in relation to exposure time.

Exposure time (min)No. of cases, nEntrance Skin Dose (mGy)
2 < t < 3393 ± 7.3
3 < t < 43104 ± 3.8
4 < t < 54148 ± 5.0
5 < t < 64163 ± 2.2
6 < t < 74194 ± 4.6
7 < t < 86207 ± 2.1
8 < t < 93286 ± 3.1
9 < t < 103326 ± 2.1
Entrance dose in relation to exposure time. The World Health Organization (WHO) holds that skin erythema and temporary epilation of the skin will occur at dosage levels of 2 and 3 µGy, yet the basis for this study was conducted using a mere 32 µGy/min [2]. Ordinarily, CT fluoroscopy uses 120 kVp tube voltage and 70 mA tuber current for a typical RFA liver procedure, and with an exposure of approximately 90 minutes, will invariably cause skin erythema and temporary epilation of the skin to occur. Given that the dosage received will increase as time increases, the scatter radiation received by the radiologist and support staff conform to this presumption (Tables 2 and 3).
Table 2

Dose received by the radiologist in relation to duration of procedures.

Duration (min)No.of cases, nScattered Radiation Dose(Radiologist (µGy))
EyeThyroidHand
2 < t < 33104 ± 329 ± 3102 ± 2
3 < t < 43113 ± 431 ± 6108 ± 1
4 < t < 54154 ± 1136 ± 1194 ± 4
5 < t < 64160 ± 851 ± 4213 ± 5
6 < t < 74179 ± 1153 ± 19227 ± 4
7 < t < 86184 ± 554 ± 2273 ± 2
8 < t < 93245 ± 567 ± 1301 ± 28
9 < t < 103260 ± 473 ± 7354 ± 4
Table 3

Dose received by the supporting staff in relation to duration of procedures.

Duration (min)No.of cases, nScattered Radiation Dose(Supporting Staff (µGy))
EyeThyroidHand
2 < t < 3316 ± 215 ± 116 ± 1
3 < t < 4321 ± 116 ± 518 ± 1
4 < t < 5422 ± 417 ± 121 ± 1
5 < t < 6424 ± 218 ± 120 ± 2
6 < t < 7426 ± 319 ± 223 ± 1
7 < t < 8631 ± 627 ± 332 ± 5
8 < t < 9338 ± 232 ± 232 ± 1
9 < t < 10347 ± 340 ± 137 ± 2
Dose received by the radiologist in relation to duration of procedures. Dose received by the supporting staff in relation to duration of procedures. Dose rate received by operators (radiologist and supporting staff) at eye, thyroid and skin of hand. The distance from the x-ray tube proved to be fundamental to the extent of radiation received. Accordingly, the radiologist being the closest to the tube, received the greatest dose to the hand (354±4 µGy), and less radiation to the eye (260±4 µGy) and the thyroid (73±7 µGy). In order to insert the RF needle into the tumour, the radiologists hand is quite close to the gantry, and this provides sound basis of the results that emerged. The rate of exposure also proved illuminating, as the radiologist recorded higher levels to the hand (37±4 µGy/min) and eye (31± µGy/min), yet only a mere 9±1 µGy/min at the thyroid. Of course, this is basis for the contention that a thyroid shield will dramatically reduce exposure to radiation, with dose rates recorded at a reduction of 24-29% compared to the unprotected eye and skin. As one would expect, the supporting staff underwent far lower levels of scatter radiation. Here, the eye received the greatest dose (47±3 µGy ) with the thyroid (40±1 µGy), and lastly the skin (37±3 µGy). Due to the steep increase in distance away from the gantry, and the negligible difference in distance of the eye, thyroid and skin, the distribution of these results was limited. No thyroid shield was worn by supporting staff, and the rate of dosage received also supported its use in order to reduce exposure to scatter radiation. For the eye, thyroid and hand respectively, the data showed rates of 5±2 µGy/min, 4±1 µGy/min, and 4±1 µGy/min.

CONCLUSIONS

While careful calibration is required, the essential element to this study was the use of Gafchromic film Type XR-QA. With it, precise quantitative measurements are possible when measuring the distribution of radiation from CT fluoroscopy. Due to the x-ray beam in CT fluoroscopy orbiting along the same channel, radiation is concentrated upon that area, the tissue surrounding which may be at risk. Conversely, the preliminary diagnostic scanning process did not record comparable results. The variables that influence the radiation received by patients and incidence of scatter radiation to operators are specifically, the duration of the procedure, the distance away from the gantry, and the dosimetric parameters. Clearly, radiologists and supporting staff need to be aware of the risks associated with exposure to scatter radiation, and adopt occupational health and safety practices in wearing lead aprons, gloves and a thyroid shield.
Table 4

Dose rate received by operators (radiologist and supporting staff) at eye, thyroid and skin of hand.

OrganRadiologist (μGy/min)Supporting staff (μGy/min)
Eye31 ± 25 ± 2
Thyroid9 ± 14 ± 1
Skin37 ± 44 ± 1
  6 in total

Review 1.  Real-time CT fluoroscopy: evolution of an interventional tool.

Authors:  B Daly; P A Templeton
Journal:  Radiology       Date:  1999-05       Impact factor: 11.105

2.  Dual-detector spiral CT arthrography of the knee: accuracy for detection of meniscal abnormalities and unstable meniscal tears.

Authors:  B C Vande Berg; F E Lecouvet; P Poilvache; J E Dubuc; B Bedat; B Maldague; J J Rombouts; J Malghem
Journal:  Radiology       Date:  2000-09       Impact factor: 11.105

3.  Radiation dose to patients and radiologists during transcatheter arterial embolization: comparison of a digital flat-panel system and conventional unit.

Authors:  Shigeru Suzuki; Shigeru Furui; Ikuo Kobayashi; Teiyu Yamauchi; Hiroshi Kohtake; Koji Takeshita; Koichi Takada; Masafumi Yamagishi
Journal:  AJR Am J Roentgenol       Date:  2005-10       Impact factor: 3.959

4.  The calibration of experimental self-developing Gafchromic HXR film for the measurement of radiation dose in computed tomography.

Authors:  K R Gorny; S L Leitzen; M R Bruesewitz; J M Kofler; N J Hangiandreou; C H McCollough
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

5.  Accurate skin dose measurements using radiochromic film in clinical applications.

Authors:  S Devic; J Seuntjens; W Abdel-Rahman; M Evans; M Olivares; E B Podgorsak; Té Vuong; Christopher G Soares
Journal:  Med Phys       Date:  2006-04       Impact factor: 4.071

6.  Tissue ablation with radiofrequency using multiprobe arrays.

Authors:  S N Goldberg; G S Gazelle; S L Dawson; W J Rittman; P R Mueller; D I Rosenthal
Journal:  Acad Radiol       Date:  1995-08       Impact factor: 3.173

  6 in total
  2 in total

1.  Evaluation of an X-Ray Dose Profile Derived from an Optically Stimulated Luminescent Dosimeter during Computed Tomographic Fluoroscopy.

Authors:  Hiroaki Hasegawa; Masanori Sato; Hiroshi Tanaka
Journal:  PLoS One       Date:  2015-07-07       Impact factor: 3.240

Review 2.  Radiation Exposure and Protection in Computed Tomography Fluoroscopy.

Authors:  Miyuki Nakatani; Shuji Kariya; Yasuyuki Ono; Takuji Maruyama; Yutaka Ueno; Atsushi Komemushi; Noboru Tanigawa
Journal:  Interv Radiol (Higashimatsuyama)       Date:  2022-06-03
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.