Literature DB >> 15772598

A comparison of 18 different x-ray detectors currently used in dentistry.

Allan G Farman1, Taeko T Farman.   

Abstract

PURPOSE: There has been a proliferation of available dental x-ray detectors over the recent past. The purpose of this short technical report is to provide a basic comparison of spatial resolution, contrast perceptibility, and relative exposure latitudes of 18 current dental x-ray detectors, including solid-state systems (CCD and CMOS), photostimulable phosphors, and analog film.
METHODS: Spatial resolution was measured using a 0.025 mm Pb phantom test grid with a measurement range from 1.5 to 20 lp/mm. For contrast perceptibility, a 7-mm thick aluminum perceptibility test device with wells of 0.1-0.9 mm depth at 0.1 mm intervals and 1 defect of 1.5 mm was used. The relative exposure latitude was determined by expert consensus using clear discrimination of the enamel-dentin junction as the lower limit and pixel blooming or unacceptable levels of cervical burnout as the upper limit.
RESULTS: The highest spatial resolution was found with Kodak InSight film, RVG-ui (CCD), and RVG 6000 (CMOS) detectors, each of which achieved 20 lp/mm, followed by the Planmeca Dixi2 v3 at > or =16 lp/mm. Contrast resolution was at least to 0.2 mm through 7 mm aluminum for all 18 detectors, with the best results being found for the Visualix HDI, RVG-ui, RVG 5000, and RVG 6000 detectors and the Schick CDR wired and wireless systems. The greatest exposure ranges were found with photostimulable phosphors and the Kodak RVG 6000 and RVG 5000 detectors.
CONCLUSIONS: Most current x-ray detectors generally perform well in terms of spatial and contrast resolutions, and in terms of exposure latitude. These findings were independent of the modality applied.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15772598     DOI: 10.1016/j.tripleo.2004.04.002

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  42 in total

1.  Digital method for quantification of circumferential periodontal bone level using cone beam CT.

Authors:  Jonathan Fleiner; Christian Hannig; Dirk Schulze; Andres Stricker; Reinhilde Jacobs
Journal:  Clin Oral Investig       Date:  2012-03-21       Impact factor: 3.573

2.  Comparison of digital systems and conventional dental film for the detection of approximal enamel caries.

Authors:  A A Pontual; D P de Melo; S M de Almeida; F N Bóscolo; F Haiter Neto
Journal:  Dentomaxillofac Radiol       Date:  2010-10       Impact factor: 2.419

3.  A comprehensive in vitro study of image accuracy and quality for periodontal diagnosis. Part 2: the influence of intra-oral image receptor on periodontal measurements.

Authors:  Bart Vandenberghe; Hilde Bosmans; Jie Yang; Reinhilde Jacobs
Journal:  Clin Oral Investig       Date:  2010-05-12       Impact factor: 3.573

Review 4.  Modern dental imaging: a review of the current technology and clinical applications in dental practice.

Authors:  Bart Vandenberghe; Reinhilde Jacobs; Hilde Bosmans
Journal:  Eur Radiol       Date:  2010-06-11       Impact factor: 5.315

5.  Diagnostic accuracy of different imaging modalities in detection of proximal caries.

Authors:  B Senel; K Kamburoglu; O Uçok; S P Yüksel; T Ozen; H Avsever
Journal:  Dentomaxillofac Radiol       Date:  2010-12       Impact factor: 2.419

6.  The effect of delayed scanning of storage phosphor plates on occlusal caries detection.

Authors:  E Sogur; B G Baksi; A Mert
Journal:  Dentomaxillofac Radiol       Date:  2012-01-26       Impact factor: 2.419

7.  Comparison of changes in dental and bone radiographic densities in the presence of different soft-tissue simulators using pixel intensity and digital subtraction analyses.

Authors:  R S de Molon; R G Batitucci; R Spin-Neto; G M Paquier; C E Sakakura; G M Tosoni; G Scaf
Journal:  Dentomaxillofac Radiol       Date:  2013-09-04       Impact factor: 2.419

8.  Enterprise-wide implementation of digital radiography in oral and maxillofacial imaging: the University of Florida Dentistry System.

Authors:  Madhu K Nair; James C Pettigrew; Jeffrey S Loomis; Robert E Bates; Stephen Kostewicz; Boyd Robinson; Jean Sweitzer; Teresa A Dolan
Journal:  J Digit Imaging       Date:  2008-09-03       Impact factor: 4.056

9.  The modulation transfer function and signal-to-noise ratio of different digital filters: a technical approach.

Authors:  D D Brüllmann; B d'Hoedt
Journal:  Dentomaxillofac Radiol       Date:  2011-05       Impact factor: 2.419

10.  A comparison of the diagnostic accuracy of in vivo and in vitro photostimulable phosphor digital images in the detection of occlusal caries lesions.

Authors:  K Kamburoglu; B Senel; S P Yüksel; T Ozen
Journal:  Dentomaxillofac Radiol       Date:  2010-01       Impact factor: 2.419

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.