Literature DB >> 21493878

The modulation transfer function and signal-to-noise ratio of different digital filters: a technical approach.

D D Brüllmann1, B d'Hoedt.   

Abstract

OBJECTIVES: The aim of this study was to illustrate the influence of digital filters on the signal-to-noise ratio (SNR) and modulation transfer function (MTF) of digital images. The article will address image pre-processing that may be beneficial for the production of clinically useful digital radiographs with lower radiation dose.
METHODS: Three filters, an arithmetic mean filter, a median filter and a Gaussian filter (standard deviation (SD) = 0.4), with kernel sizes of 3 × 3 pixels and 5 × 5 pixels were tested. Synthetic images with exactly increasing amounts of Gaussian noise were created to gather linear regression of SNR before and after application of digital filters. Artificial stripe patterns with defined amounts of line pairs per millimetre were used to calculate MTF before and after the application of the digital filters.
RESULTS: The Gaussian filter with a 5 × 5 kernel size caused the highest noise suppression (SNR increased from 2.22, measured in the synthetic image, to 11.31 in the filtered image). The smallest noise reduction was found with the 3 × 3 median filter. The application of the median filters resulted in no changes in MTF at the different resolutions but did result in the deletion of smaller structures. The 5 × 5 Gaussian filter and the 5 × 5 arithmetic mean filter showed the strongest changes of MTF.
CONCLUSIONS: The application of digital filters can improve the SNR of a digital sensor; however, MTF can be adversely affected. As such, imaging systems should not be judged solely on their quoted spatial resolutions because pre-processing may influence image quality.

Mesh:

Year:  2011        PMID: 21493878      PMCID: PMC3520312          DOI: 10.1259/dmfr/33029984

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  16 in total

1.  Physical evaluation of a system for direct digital intra-oral radiography based on a charge-coupled device.

Authors:  K Yoshiura; H C Stamatakis; U Welander; W D McDavid; X Q Shi; S Ban; T Kawazu; M Tatsumi; S Kanda
Journal:  Dentomaxillofac Radiol       Date:  1999-09       Impact factor: 2.419

Review 2.  Radiographic image enhancement. Part I: spatial domain techniques.

Authors:  M Analoui
Journal:  Dentomaxillofac Radiol       Date:  2001-01       Impact factor: 2.419

Review 3.  Radiographic digital image enhancement. Part II: transform domain techniques.

Authors:  M Analoui
Journal:  Dentomaxillofac Radiol       Date:  2001-03       Impact factor: 2.419

4.  Fundamentals of image acquisition and processing in the digital era.

Authors:  A G Farman
Journal:  Orthod Craniofac Res       Date:  2003       Impact factor: 1.826

5.  Measurement of the presampling modulation transfer function of film digitizers using a curve fitting technique.

Authors:  F F Yin; M L Giger; K Doi
Journal:  Med Phys       Date:  1990 Nov-Dec       Impact factor: 4.071

Review 6.  Physical performance measures of radiographic imaging systems.

Authors:  A Workman; D S Brettle
Journal:  Dentomaxillofac Radiol       Date:  1997-05       Impact factor: 2.419

7.  Correction of background noise in direct digital dental radiography.

Authors:  T Yoshioka; C Kobayashi; H Suda; T Sasaki
Journal:  Dentomaxillofac Radiol       Date:  1996-11       Impact factor: 2.419

8.  Analytic approximation of the off-axis modulation transfer function of the eye.

Authors:  J A Jennings; W N Charman
Journal:  Vision Res       Date:  1997-03       Impact factor: 1.886

9.  Effect of beam energy and filtration on the signal-to-noise ratio of the Dexis intraoral X-ray detector.

Authors:  H Kitagawa; A G Farman
Journal:  Dentomaxillofac Radiol       Date:  2004-01       Impact factor: 2.419

10.  Task-specific enhancement filters in storage phosphor images from the Vistascan system for detection of proximal caries lesions of known size.

Authors:  Francisco Haiter-Neto; Marcia Spinelli Casanova; Morten Frydenberg; Ann Wenzel
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2009-01
View more
  7 in total

Review 1.  Spatial resolution in CBCT machines for dental/maxillofacial applications-what do we know today?

Authors:  D Brüllmann; R K W Schulze
Journal:  Dentomaxillofac Radiol       Date:  2015       Impact factor: 2.419

2.  Effect of beam quality and readout direction in the edge profile on the modulation transfer function of photostimulable phosphor systems via the edge method.

Authors:  Shinya Takarabe; Taku Kuramoto; Yusuke Shibayama; Hiroki Tsuru; Masato Tatsumi; Toyoyuki Kato; Kazutoshi Okamura; Kazunori Yoshiura
Journal:  J Med Imaging (Bellingham)       Date:  2021-07-20

3.  Image resolution in the digital era: notion and clinical implications.

Authors:  Vahid Rakhshan
Journal:  J Dent (Shiraz)       Date:  2014-12

4.  Effect of data binning and frame averaging for micro-CT image acquisition on the morphometric outcome of bone repair assessment.

Authors:  Milena Suemi Irie; Rubens Spin-Neto; Juliana Simeão Borges; Ann Wenzel; Priscilla Barbosa Ferreira Soares
Journal:  Sci Rep       Date:  2022-01-26       Impact factor: 4.379

5.  Terbium doped LiLuF4 nanocrystal scintillator-based flexible composite film for high resolution X-ray imaging.

Authors:  Hao Lu; Xieming Xu; Guiqing Feng; Baoping Sun; Shuaihua Wang; Shaofan Wu
Journal:  RSC Adv       Date:  2022-02-07       Impact factor: 3.361

6.  Spatiotemporal Rank Filtering Improves Image Quality Compared to Frame Averaging in 2-Photon Laser Scanning Microscopy.

Authors:  Henry Pinkard; Kaitlin Corbin; Matthew F Krummel
Journal:  PLoS One       Date:  2016-03-03       Impact factor: 3.240

7.  Effect of differences in pixel size on image characteristics of digital intraoral radiographic systems: a physical and visual evaluation.

Authors:  Taku Kuramoto; Shinya Takarabe; Kazutoshi Okamura; Kenshi Shiotsuki; Yusuke Shibayama; Hiroki Tsuru; Hiroshi Akamine; Masato Tatsumi; Toyoyuki Kato; Junji Morishita; Kazunori Yoshiura
Journal:  Dentomaxillofac Radiol       Date:  2020-05-07       Impact factor: 2.419

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.