Frank G Shellock1, Matthew Gounis, Ajay Wakhloo.
Abstract
A detachable coil was assessed for magnetic field interactions, heating (body RF coil, SAR of 2.0-W/kg), and artifacts at 3-Tesla. The coil showed no magnetic field interactions and heating was negligible (+0.3 degrees C). Therefore, this coil is safe for a patient undergoing MR imaging at 3T or less. While artifacts may impair the ability to properly visualize anatomy in proximity to this implant, careful selection of imaging parameters can mitigate this problem.Entities:
Mesh:
Year: 2005 PMID: 15709137 PMCID: PMC7974076