Effect of blood collection tubes on total triiodothyronine and other laboratory assays.

Raffick A R Bowen1, Yung Chan, Joshua Cohen, Nadja N Rehak, Glen L Hortin, Gyorgy Csako, Alan T Remaley.   

Abstract

BACKGROUND: Increased total triiodothyronine (TT(3)) assay results in apparently euthyroid patients triggered an investigation of the effect of blood collection tubes on serum TT(3) and other laboratory assays.
METHODS: We examined potential assay interference for three types of tubes: plastic Greiner Bio-One Vacuette; glass Becton Dickinson (BD) Vacutainer; and plastic BD Vacutainer SST tubes. Serum samples from apparently healthy volunteers (age range, 30-60 years; 15 males and 34 females) were collected in different tube types and analyzed in 17 immunoassays (n = 49), 30 clinical chemistry tests (n = 20), and 33 immunology assays (n = 15). Tube effects were also examined by adding pooled serum to different tube types.
RESULTS: TT(3) values, when measured by the IMMULITE 2000 but not the AxSYM analyzer, were significantly higher (P <0.0001) for SST (2.81 nmol/L) than either glass (2.15 nmol/L) or Vacuette (2.24 nmol/L) tubes. The effect was large enough to substantially shift the distribution of patient values, increasing the percentage of values above the reference interval from 11.3% to 35.8%. The degree of interference from SST tubes on TT(3) differed among various tube lots and could be attributed to a tube additive shared by other plastic tubes. Results from several other tests statistically differed among tube types, but differences were not considered to be clinically significant.
CONCLUSIONS: Assay interferences from blood collection tubes represent challenges to clinical laboratories because they are not detected by the usual quality-control or proficiency testing programs. Laboratories can, however, address this problem by monitoring distribution of patients' results.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15576427     DOI: 10.1373/clinchem.2004.043349

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  11 in total

1.  Effects of different centrifugation conditions on clinical chemistry and Immunology test results.

Authors:  Elisabeth I Minder; Adrian Schibli; Dagmar Mahrer; Predrag Nesic; Kathrin Plüer
Journal:  BMC Clin Pathol       Date:  2011-05-10

2.  Comparison of the effect of gel used in two different serum separator tubes for thyroid function tests.

Authors:  Müjgan Ercan; Esra Fırat Oğuz; Emiş Deniz Akbulut; Mustafa Yilmaz; Turan Turhan
Journal:  J Clin Lab Anal       Date:  2018-03-07       Impact factor: 2.352

3.  Evaluation of the thyroid in patients with hearing loss and enlarged vestibular aqueducts.

Authors:  Anne C Madeo; Ani Manichaikul; James C Reynolds; Nicholas J Sarlis; Shannon P Pryor; Thomas H Shawker; Andrew J Griffith
Journal:  Arch Otolaryngol Head Neck Surg       Date:  2009-07

4.  Different collection tubes in cardiac biomarkers detection.

Authors:  Massimo Daves; Diego Trevisan; Roberto Cemin
Journal:  J Clin Lab Anal       Date:  2008       Impact factor: 2.352

5.  Blood collection tube-related alterations in analyte concentrations in quality control material and serum specimens.

Authors:  Raffick A R Bowen; Annie Sattayapiwat; Verena Gounden; Alan T Remaley
Journal:  Clin Biochem       Date:  2013-11-14       Impact factor: 3.281

6.  Collection tubes with or without gel separator did not interfere with detection of rubella virus antibodies IgM and IgG.

Authors:  L C Oliveira; K H Kawasato; M S Otta; L P Lima; T S Okay
Journal:  J Clin Lab Anal       Date:  2007       Impact factor: 2.352

7.  Implementation of biological variation-based analytical performance specifications in the laboratory: Stringent evaluation of Improvacutor blood collection tubes.

Authors:  Hee-Jung Chung; Yoon Kyung Song; Sung Kuk Hong; Sang-Hyun Hwang; Hee Seung Seo; Dong Hee Whang; Myung-Hyun Nam; Do Hoon Lee
Journal:  PLoS One       Date:  2017-12-20       Impact factor: 3.240

Review 8.  Interferences from blood collection tube components on clinical chemistry assays.

Authors:  Raffick A R Bowen; Alan T Remaley
Journal:  Biochem Med (Zagreb)       Date:  2014-02-15       Impact factor: 2.313

Review 9.  Cytokines as biomarkers in rheumatoid arthritis.

Authors:  Agata Burska; Marjorie Boissinot; Frederique Ponchel
Journal:  Mediators Inflamm       Date:  2014-03-09       Impact factor: 4.711

10.  The local clinical validation of a new lithium heparin tube with a barrier: BD Vacutainer® Barricor LH Plasma tube.

Authors:  Fatma Demet Arslan; Inanc Karakoyun; Banu Isbilen Basok; Merve Zeytinli Aksit; Anil Baysoy; Yasemin Kilic Ozturk; Yusuf Adnan Guclu; Can Duman
Journal:  Biochem Med (Zagreb)       Date:  2017-08-28       Impact factor: 2.313

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.