| Literature DB >> 28900369 |
Fatma Demet Arslan1, Inanc Karakoyun1, Banu Isbilen Basok1, Merve Zeytinli Aksit1, Anil Baysoy1, Yasemin Kilic Ozturk2, Yusuf Adnan Guclu2, Can Duman1.
Abstract
INTRODUCTION: Although serum-providing blood tubes with a barrier are still widely used due to their significant advantages, the use of blood tubes with a barrier to provide plasma is becoming widespread. We compared 22 analytes in a BD Vacutainer® Barricor LH Plasma tube for local clinical validation of this new lithium heparin tube with a barrier.Entities:
Keywords: blood analysis; blood specimen collection; evaluation; plasma; serum
Mesh:
Substances:
Year: 2017 PMID: 28900369 PMCID: PMC5575652 DOI: 10.11613/BM.2017.030706
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Analytical and clinical data on 22 analytes tested in 3 different types of blood collection tubes in comparison to the reference tube
| 36.8 | 43.7 | 1.43 | 40.2 | 40.2 | - 0.14 | 0.880 | 40.2 | - 0.14 | 0.447 | 40.3 | 0.11 | 0.466 | |
| 89 | 119 | 6.72 | 105 | 104 | - 0.46 | 0.061 | 101 | - 3.17 | < 0.001 | 100 | - 3.92 | < 0.001 | |
| 21 | 27 | 11.48 | 25 | 25 | 0.09 | 0.681 | 25 | 0.56 | 0.333 | 25 | 4.05 | 0.006 | |
| 21 | 30 | 6.54 | 25 | 26 | 1.93 | 0.022 | 25 | - 0.45 | 0.614 | 28 | 9.19 | < 0.001 | |
| 268 | 374 | 12.00 | 321 | 324 | 0.92 | 0.861 | 319 | - 0.78 | 0.334 | 321 | - 0.07 | 0.933 | |
| -0.76 | 4.47 | 14.20 | 1.86 | 1.79 | - 3.77 | 0.005 | 1.93 | 4.08 | 0.060 | 1.90 | 2.30 | 0.591 | |
| 7.39 | 12.08 | 8.95 | 9.74 | 10.05 | 3.19 | 0.083 | 10.07 | 3.39 | 0.019 | 10.16 | 4.39 | 0.001 | |
| 2.32 | 2.47 | 0.82 | 2.40 | 2.40 | - 0.05 | 0.880 | 2.42 | 0.54 | 0.018 | 2.40 | - 0.61 | 0.007 | |
| 97 | 109 | 0.50 | 103 | 103 | - 0.24 | 0.043 | 103 | - 0.06 | 0.606 | 103 | - 0.17 | 0.134 | |
| 73 | 110 | 3.96 | 91 | 91 | 0.04 | 0.763 | 91 | - 0.18 | 0.564 | 91 | 0.27 | 0.581 | |
| 6.7 | 8.4 | 19.20 | 7.6 | 7.6 | 0.84 | 0.911 | 7.5 | - 0.95 | 0.264 | 7.4 | - 2.30 | 0.004 | |
| 3.99 | 5.83 | 4.80 | 4.91 | 4.90 | - 0.16 | 0.824 | 4.89 | - 0.40 | 0.629 | 4.89 | - 0.63 | 0.395 | |
| 10.18 | 13.31 | 3.30 | 11.74 | 11.83 | 0.75 | 0.203 | 11.73 | - 0.12 | 0.893 | 11.87 | 1.07 | 0.091 | |
| 38 | 46 | 6.70 | 43 | 43 | 0.14 | 0.347 | 42 | - 2.20 | < 0.001 | 44 | 5.51 | 0.183 | |
| 6.0 | 6.6 | 2.34 | 6.3 | 6.3 | 0.03 | 0.683 | 6.3 | 0.62 | 0.006 | 6.1 | < 0.001 | ||
| 4.1 | 4.5 | 1.81 | 4.3 | 4.4 | 1.05 | 0.074 | - 7.05 | < 0.001 | 4.1 | < 0.001 | |||
| 170 | 207 | 4.30 | 188 | 205 | 8.85 | < 0.001 | 175 | - 6.99 | < 0.001 | 229* ( | < 0.001 | ||
| 133 | 150 | 0.23 | 142 | 140 | - 0.75 | < 0.001 | 141 | - 0.55 | < 0.001 | 141 | 0.012 | ||
| 67.1 | 74.7 | 1.36 | 71.0 | 71.2 | 0.34 | 0.013 | 74.6 | 5.18 | < 0.001 | 74.5 | < 0.001 | ||
| 1.44 | 1.55 | 9.70 | 1.51 | 1.50 | 0.71 | 0.879 | 1.54 | 1.96 | 0.003 | 1.52 | 0.83 | 0.103 | |
| 266 | 374 | 4.87 | 315 | 315 | - 0.54 | 0.232 | 322 | 1.55 | < 0.001 | 313 | 1.03 | 0.281 | |
| 12.8 | 15.0 | 5.57 | 13.9 | 13.9 | - 0.20 | 0.278 | 13.6 | - 0.50 | 0.096 | 13.9 | - 0.26 | 0.290 | |
| (±) SCL – (upper/lower) significant change limit. Biasd - desirable quality specifications for bias. SEM - standard error of the mean. %Bias - difference between the compared tube results and the reference tube results. Z-tube - glass tube without additive | |||||||||||||
The effects of two different centrifugation times on AST, K, and LD in Barricor plasma
| 23 | 19 - 30 | 23 | 19 - 30 | 0.687 | |
| 4.3 | 4.1 - 4.9 | 4.2 | 4.1 - 4.8 | 0.084 | |
| 200 | 188 - 237 | 210 | 187 - 239 | 0.131 | |
| Barricor - lithium heparin tube with barrier. IQR - interquartile range. AST, aspartate aminotransferase; K, potassium; LD, lactate dehydrogenase. P < 0.05 was considered statistically significant. *Centrifuged at 2360xg. | |||||
The effects of two different blood collection techniques on AST, K, LD and free Hb results in Barricor plasma samples
| 24 | 26 (20 – 28) | 5.6 | 0.031 | 24 | 26 (23 - 27) | 7.0 | 0.004 | 6.54 | 0.944 | ||
| 23 | 25 (21 – 26) | 23 | 24 (21 - 25) | 0.129 | |||||||
| 4.0 | 4.1 (3.9 - 4.4) | 0.7 | 0.046 | 3.9 | 3.9 (3.7 - 4.2) | - 1.7 | 0.414 | 1.81 | 0.221 | ||
| 4.0 | 4.0 (3.9 - 4.4) | 4.0 | 4.0 (3.7 - 4.3) | 0.799 | |||||||
| 189 | 181 (175 – 214) | 21.2 | 0.007 | 178 | 178 (168 - 186) | 12.4 | 0.005 | 4.30 | 0.185 | ||
| 156 | 155 (142 – 172) | 158 | 158 (147 - 168) | 0.767 | |||||||
| 0.87 | 0.37 (0.11 - 1.46) | NC | 0.953 | 0.17 | 0.13 (0.06 - 0.20) | NC | 0.153 | ND | 0.139 | ||
| 0.76 | 0.44 (0.10 - 1.65) | 0.25 | 0.22 (0.18 - 0.33) | 0.169 | |||||||
| IQR - interquartile range. %Bias - difference between Barricor tube results and LiH tube results. Biasd - desirable quality specifications for bias. NC - not calculated. ND - not determined. AST, aspartate aminotransferase; K, potassium; LD, lactate dehydrogenase; free Hb, free haemoglobin. P < 0.05 was considered statistically significant. | |||||||||||
The effects of individual differences on AST, K, and LD in Barricor plasma
| 30 | 25 (21 - 28) | 10.6 | 26 | 23 (20 - 32) | 7.4 | 0.715 | |
| 4.0 | 4.1 (3.7 - 4.2) | - 5.8 | 4.2 | 4.2 (3.9 - 4.4) | - 3.8 | 0.053 | |
| 240 | 223 (185 - 265) | 25.3 | 218 | 209 (175 - 254) | 16.3 | 0.429 | |
| Barricor - lithium heparin tube with barrier. IQR - interquartile range. %Bias - difference between Barricor and Z tube’ test results in patients and healthy individuals. AST, aspartate aminotransferase; K, potassium; LD, lactate dehydrogenase. P < 0.05 was considered statistically significant. | |||||||