Blood collection tube-related alterations in analyte concentrations in quality control material and serum specimens.

Raffick A R Bowen1, Annie Sattayapiwat2, Verena Gounden3, Alan T Remaley3.   

Abstract

OBJECTIVES: Several previous studies have described the effects of interfering substances on clinical assay results; however, the effects of exogenous substances, particularly additives from blood collection tubes on quality control (QC) specimens and serum specimens have not been well examined. This study examines the effects of blood-collection tube additives on total triiodothyronine (TT3), and thyroxine (TT4), cortisol, and routine clinical chemistry tests in QC and serum specimens from apparently healthy volunteers.
METHODS: QC and serum specimens were poured or collected into different blood collection tubes. TT3 and TT4, cortisol, and routine chemistry tests were analyzed from the different blood-collection tube types.
RESULTS: The findings of this study demonstrate statistically and/or clinically significant blood collection tube-related alterations in the TT3, TT4, and cortisol concentrations of QC specimens and TT4 concentrations from serum specimens.
CONCLUSIONS: These findings have important implications for clinical laboratories, demonstrating that QC specimens should ideally, like patients' specimens, be poured into blood collection tubes. This strategy would reveal any adverse effects caused by blood collection tubes, which otherwise would not likely be detected by most routine QC practices. The results of this study also show the importance of producing blood collection tubes that contain additives that are truly inert and do not adversely affect clinical laboratory testing.
Copyright © 2013 The Canadian Society of Clinical Chemists. All rights reserved.

Entities:  

Keywords:  Blood collection tube; Quality control material; Thyroxine; Triiodothyronine

Mesh:

Substances:

Year:  2013        PMID: 24240064      PMCID: PMC3944914          DOI: 10.1016/j.clinbiochem.2013.11.003

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  17 in total

1.  Comparison of BD Vacutainer SST Plus Tubes with BD SST II Plus Tubes for common analytes.

Authors:  V J Bush; M R Janu; F Bathur; A Wells; A Dasgupta
Journal:  Clin Chim Acta       Date:  2001-04       Impact factor: 3.786

Review 2.  Specimen materials, target values and commutability for external quality assessment (proficiency testing) schemes.

Authors:  W Greg Miller
Journal:  Clin Chim Acta       Date:  2003-01       Impact factor: 3.786

3.  Stability studies of twenty-four analytes in human plasma and serum.

Authors:  Bobby L Boyanton; Kenneth E Blick
Journal:  Clin Chem       Date:  2002-12       Impact factor: 8.327

Review 4.  Laboratory quality control: using patient data to assess analytical performance.

Authors:  Steven C Kazmierczak
Journal:  Clin Chem Lab Med       Date:  2003-05       Impact factor: 3.694

Review 5.  Thoughts on quality-control systems: a laboratorian's perspective.

Authors:  G S Cembrowski
Journal:  Clin Chem       Date:  1997-05       Impact factor: 8.327

Review 6.  Preanalytical quality improvement: in quality we trust.

Authors:  Giuseppe Lippi; Kathleen Becan-McBride; Darina Behúlová; Raffick A Bowen; Stephen Church; Joris Delanghe; Kjell Grankvist; Steve Kitchen; Mads Nybo; Matthias Nauck; Nora Nikolac; Vladimir Palicka; Mario Plebani; Sverre Sandberg; Ana-Maria Simundic
Journal:  Clin Chem Lab Med       Date:  2013-01       Impact factor: 3.694

7.  Evaluation of evacuated blood-collection tubes: effects of three types of polymeric separators on therapeutic drug-monitoring specimens.

Authors:  M Landt; C H Smith; G L Hortin
Journal:  Clin Chem       Date:  1993-08       Impact factor: 8.327

8.  Introduction to sample size determination and power analysis for clinical trials.

Authors:  J M Lachin
Journal:  Control Clin Trials       Date:  1981-06

9.  Suitability of collection tubes with separator gels for collecting and storing blood samples for therapeutic drug monitoring (TDM).

Authors:  J Karppi; K K Akerman; M Parviainen
Journal:  Clin Chem Lab Med       Date:  2000-04       Impact factor: 3.694

10.  Effect of blood collection tubes on total triiodothyronine and other laboratory assays.

Authors:  Raffick A R Bowen; Yung Chan; Joshua Cohen; Nadja N Rehak; Glen L Hortin; Gyorgy Csako; Alan T Remaley
Journal:  Clin Chem       Date:  2004-12-02       Impact factor: 8.327

View more
  4 in total

1.  Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D.

Authors:  Songlin Yu; Weiyan Zhou; Xinqi Cheng; Huiling Fang; Ruiping Zhang; Qian Cheng; Jianhua Han; Wei Su; Liangyu Xia; Ling Qiu
Journal:  PLoS One       Date:  2016-11-10       Impact factor: 3.240

2.  Verification of the Canadian Laboratory Initiative on Paediatric Reference Intervals (CALIPER) reference values in Croatian children and adolescents.

Authors:  Renata Zrinski Topić; Jasna Leniček Krleža
Journal:  Biochem Med (Zagreb)       Date:  2020-06-15       Impact factor: 2.313

3.  A comparative study on the effect of blood collection tubes on stress oxidative markers.

Authors:  Alireza Bastin; Saba Fooladi; Amir Hossein Doustimotlagh; Sina Vakili; Amir Hashem Aminizadeh; Sanaz Faramarz; Hamidreza Shiri; Mohammad Hadi Nematollahi
Journal:  PLoS One       Date:  2022-04-06       Impact factor: 3.240

Review 4.  Interferences from blood collection tube components on clinical chemistry assays.

Authors:  Raffick A R Bowen; Alan T Remaley
Journal:  Biochem Med (Zagreb)       Date:  2014-02-15       Impact factor: 2.313

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.