| Literature DB >> 29261759 |
Hee-Jung Chung1, Yoon Kyung Song1, Sung Kuk Hong1, Sang-Hyun Hwang2, Hee Seung Seo1, Dong Hee Whang3, Myung-Hyun Nam4, Do Hoon Lee1.
Abstract
Recently, because the quality of laboratory analyses has increased along with the need for quality improvement, several external quality control bodies have adapted performance specifications using the Desirable Biological Variation Database, termed "Ricos goals"; these criteria are more stringent than those presented in CLIA 88. In this study, we aimed to validate newly introduced serum separator tubes, Improvacutor, for routine clinical chemistry testing in accordance with Ricos goals and CLIA 88. Blood samples were collected from 100 volunteers into three types of serum vacuum tubes: Greiner Vacuette, Becton Dickinson (BD) Vacutainer, and Improve Improvacutor. The samples were subjected to 16 routine chemistry tests using a TBA-200fr NEO chemistry autoanalyzer. In the comparison analysis, all 16 test results were acceptable according to CLIA 88. However, in the comparison of Improve and BD tubes, creatinine showed 4.31% (+0.08 μmol/L) bias. This slightly exceeded the Desirable Specification for Inaccuracy Ricos limit of ±3.96%, but still satisfied the CLIS88 limit of ±26.52 μmol/L. The remaining 15 analytes performed acceptably according to the Desirable Specifications of Ricos. The correlation coefficient of 12 analytes was greater than 0.95 in Passing-Bablok regression analysis among the three tubes, but was lower for four analytes: calcium, sodium, potassium, and chloride. In the stability assay, only potassium tested in the Greiner tube revealed a larger positive bias (2.18%) than the Ricos Desirable Specification for Inaccuracy based on biologic variation (1.8%). The BD tube also showed a positive bias of 1.74%, whereas the new Improve tube showed the smallest positive bias of 1.17% in potassium level after 72 h storage. Thus, the results of this study demonstrate that recently introduced analytical performance specifications based on components of biological variation (Rico's goal) could be extended to criterion for performance evaluation and applied.Entities:
Mesh:
Substances:
Year: 2017 PMID: 29261759 PMCID: PMC5738124 DOI: 10.1371/journal.pone.0189882
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Statistical analysis of the results of blood samples from three types of collection tubes (mean ± SD).
| Test (unit) | BD | Greiner | Improve | ||
|---|---|---|---|---|---|
| Ca | (mmol/L) | 2.38 ± 0.10 | 2.38 ± 0.10 | 2.39 ± 0.10 | 0.753 |
| PHOS | (mmol/L) | 1.21 ± 0.16 | 1.21 ± 0.16 | 1.21 ± 0.16 | 0.964 |
| GLU | (mmol/L) | 5.8 ± 2.2 | 5.8 ± 2.2 | 5.9 ± 2.3 | 0.994 |
| BUN | (mmol/L) | 5.0 ± 1.6 | 5.0 ± 1.6 | 5.1 ± 1.7 | 0.948 |
| UA | (mmol/L) | 0.28 ± 0.08 | 0.28 ± 0.07 | 0.28 ± 0.08 | 0.994 |
| CHOL | (mmol/L) | 5.1 ± 1.0 | 5.1 ± 1.0 | 5.1 ± 1.0 | 0.916 |
| TP | (g/L) | 76.2 ± 3.8 | 75.8 ± 3.8 | 76.4 ± 4.0 | 0.553 |
| ALB | (g/L) | 44.1 ± 2.4 | 43.9 ± 2.3 | 44.2 ± 2.4 | 0.596 |
| TB | (μmol/L) | 11.6 ± 5.3 | 11.5 ± 5.3 | 11.6 ± 5.3 | 0.983 |
| ALP | (IU/L) | 67.0 ± 22.2 | 66.4 ± 21.8 | 67.4 ± 22.3 | 0.954 |
| AST | (IU/L) | 21.9 ± 18.7 | 21.9 ± 18.6 | 21.9 ± 18.8 | 1.000 |
| ALT | (IU/L) | 20.3 ± 17.3 | 20.4 ± 17.4 | 20.4 ± 17.5 | 0.999 |
| CREA | (μmol/L) | 61.9 ± 11.5 | 62.8 ± 12.4 | 64.5 ± 12.4 | 0.284 |
| Na | (mmol/L) | 141.4 ± 1.7 | 141.3 ± 1.7 | 141.6 ± 1.6 | 0.597 |
| K | (mmol/L) | 4.4 ± 0.4 | 4.4 ± 0.4 | 4.4 ± 0.4 | 0.807 |
| Cl | (mmol/L) | 104.0 ± 2.0 | 103.8 ± 2.3 | 103.9 ± 2.0 | 0.780 |
P values were calculated by ANOVA and post-hoc tests. The decimal point was marked as the reporting unit of clinical results. For example, the mean of calcium is shown as 2.38 mmol/L because the calcium level is clinically reported as 3.0 mmol/L.
Fig 1Bland-Altman plots for the representative 3 routine chemistry analytes: (A) Ca,(B) BUN, and (C) CREA for the three types of tubes. To compare the absolute differences vertically, three figures were generated for each analyte (BD Vacutainer versus Greiner Vacuette, Improvacutor versus Greiner Vacuette, and Improve Improvacutor versus BD Vacutainer) from the left. Solid lines denote the average difference and ± 2SD in SI units.
Bias analysis and comparison of the results of blood sample analysis from the three types of collection tubes according to desirable specifications of Ricos and CLIA 88.
| Analyte | Laboratory Imprecision | Bias (%) | Desirable Specification for Inaccuracy | CLIA 88 | |||
|---|---|---|---|---|---|---|---|
| BD vs Greiner | Improve vs Greiner | Improve vs BD | |||||
| Ca | 0.98 | 0.28 | 0.44 | 0.16 | 0.82 | Target ± | 0.25 mmol/L |
| PHOS | 1.37 | 0.43 | 0.45 | 0.03 | 3.4 | Target ± | 10% |
| GLU | 0.72 | -0.16 | 0.43 | 0.59 | 2.3 | Target ± | 0.333 mmol/L |
| BUN | 1.25 | 0.61 | 1.51 | 0.90 | 5.6 | Target ± | 0.714 mmol/L |
| UA | 1.04 | -0.06 | 0.32 | 0.38 | 4.9 | Target ± | 17% |
| CHOL | 0.76 | 0.58 | 1.17 | 0.59 | 4.1 | Target ± | 10% |
| TP | 0.78 | 0.49 | 0.78 | 0.29 | 1.4 | Target ± | 10% |
| ALB | 0.74 | 0.41 | 0.77 | 0.36 | 1.4 | Target ± | 10% |
| TB | 1.47 | 0.74 | 1.19 | 0.44 | 8.95 | Target ± | 20% |
| ALP | 1.41 | 0.83 | 1.43 | 0.60 | 6.7 | Target ± | 30% |
| AST | 1.40 | 0.05 | -0.18 | -0.23 | 6.5 | Target ± | 20% |
| ALT | 2.44 | -0.59 | -0.20 | 0.39 | 11.48 | Target ± | 20% |
| CREA | 1.35 | -1.40 | 2.85 | 4.31 | 3.96 | Target ± | 26.52 μmol/L |
| Na | 0.47 | 0.08 | 0.17 | 0.09 | 0.23 | Target ± | 4.0 mmol/L |
| K | 0.81 | 0.80 | 0.73 | -0.07 | 1.8 | Target ± | 0.5 mmol/L |
| Cl | 0.50 | 0.18 | 0.16 | -0.02 | 0.5 | Target ± | 5% |
a Laboratory imprecision was verified in accordance with CLSI document EP15-A3 [23]
b Bias (%) = ([test tube mean − reference tube mean] / reference tube mean) × 100. The reference tube was the Greiner tube. In comparisons of the Improve versus BD tubes, the newly introduced Improve tubes were considered as the test tubes.
c Ricos goal, which is the desirable bias derived from biological variation [3, 4]. Decimal point was not unified and was marked as shown in the database. By rounding the decimal point and unifying it to one decimal place, it is possible to cause an error in judgment of acceptability.
d Allowable total error limits for linearity in the CAP survey were adapted because there is no criterion in CLIA 88.
Comparison of test results before and after 72-h storage of the same samples at 4°C.
| Analyte (unit) | BD | Greiner | Improve | Desirable specification for inaccuracy | ||||
|---|---|---|---|---|---|---|---|---|
| Bias (%) | Bias (%) | Bias (%) | ||||||
| Ca | (mmol/L) | −0.3 | 0.10 | 0.1 | 0.72 | −0.5 | 0.82 | |
| PHOS | (mmol/L) | 2.1 | < | 2.2 | < | 1.4 | < | 3.4 |
| GLU | (mmol/L) | 0.3 | 0.6 | < | 0.1 | 0.43 | 2.3 | |
| BUN | (mmol/L) | 0.8 | < | 1.2 | < | 0.4 | 0.09 | 5.6 |
| UA | (mmol/L) | 0.6 | 0.5 | 0.4 | 0.06 | 4.9 | ||
| CHOL | (mmol/L) | 0.4 | < | 0.7 | < | 0.2 | 0.08 | 4.1 |
| TP | (g/L) | −0.1 | 0.57 | 0.4 | < | 0.0 | 0.8 | 1.4 |
| ALB | (g/L) | 0.1 | 0.41 | 0.3 | 0.2 | 0.23 | 1.4 | |
| TB | (μmol/L) | −5.9 | < | −3.3 | < | −4.1 | < | 8.95 |
| ALP | (IU/L) | −0.5 | -0.6 | < | −1.5 | < | 6.7 | |
| AST | (IU/L) | 0.6 | 0.06 | 0.6 | 0.12 | −1.0 | < | 6.5 |
| ALT | (IU/L) | −2.2 | < | −2.4 | < | −2.9 | < | 11.48 |
| CREA | (μmol/L) | −0.2 | 0.64 | −0.3 | 0.59 | 1.6 | < | 3.96 |
| Na | (mmol/L) | 0.0 | 0.79 | 0.1 | 0.25 | −0.1 | 0.14 | 0.23 |
| K | (mmol/L) | 1.7 | < | 2.2 | < | 1.2 | < | 1.8 |
| Cl | (mmol/L) | −0.1 | 0.33 | 0.2 | 0.21 | −0.1 | 0.5 | |
P values were calculated by Student’s paired t -test.
Statistically significant values are shown in italic.
a Bias (%) = ([72 h tube mean − 0 h tube mean] / 0 h tube mean) × 100, in each tube.
b Ricos goal, which is the desirable bias derived from biological variation [3, 4]. Decimal point was not unified and was marked as shown in the database. By rounding the decimal point and unifying it to one decimal place, it is possible to cause an error in judgment of acceptability.