Effects of different centrifugation conditions on clinical chemistry and Immunology test results.

Elisabeth I Minder1, Adrian Schibli, Dagmar Mahrer, Predrag Nesic, Kathrin Plüer.   

Abstract

BACKGROUND: The effect of centrifugation time of heparinized blood samples on clinical chemistry and immunology results has rarely been studied. WHO guideline proposed a 15 min centrifugation time without citing any scientific publications. The centrifugation time has a considerable impact on the turn-around-time.
METHODS: We investigated 74 parameters in samples from 44 patients on a Roche Cobas 6000 system, to see whether there was a statistical significant difference in the test results among specimens centrifuged at 2180 g for 15 min, at 2180 g for 10 min or at 1870 g for 7 min, respectively. Two tubes with different plasma separators (both Greiner Bio-One) were used for each centrifugation condition. Statistical comparisons were made by Deming fit.
RESULTS: Tubes with different separators showed identical results in all parameters. Likewise, excellent correlations were found among tubes to which different centrifugation conditions were applied. Fifty percent of the slopes lay between 0.99 and 1.01. Only 3.6 percent of the statistical tests results fell outside the significance level of p < 0.05, which was less than the expected 5%. This suggests that the outliers are the result of random variation and the large number of statistical tests performed. Further, we found that our data are sufficient not to miss a biased test (beta error) with a probability of 0.10 to 0.05 in most parameters.
CONCLUSION: A centrifugation time of either 7 or 10 min provided identical test results compared to the time of 15 min as proposed by WHO under the conditions used in our study.

Entities:  

Year:  2011        PMID: 21569233      PMCID: PMC3123263          DOI: 10.1186/1472-6890-11-6

Source DB:  PubMed          Journal:  BMC Clin Pathol        ISSN: 1472-6890


  6 in total

1.  Evaluation of a centrifuge with rapid turnaround time for the preparation of plasma samples for measurement of common STAT markers on the ACS: 180 system.

Authors:  K Foster; P Datta; M Orswell; K Tasaico; A Alpert; B Bluestein
Journal:  Clin Lab       Date:  2000       Impact factor: 1.138

2.  Considerably reduced centrifugation time without increased hemolysis: evaluation of the new BD Vacutainer SSTTMII Advance.

Authors:  Birger Mensel; Ulrike Wenzel; Markus Roser; Jan Lüdemann; Matthias Nauck
Journal:  Clin Chem       Date:  2007-04       Impact factor: 8.327

3.  Evaluation of the performance and clinical impact of a rapid intraoperative parathyroid hormone assay in conjunction with preoperative imaging and concise parathyroidectomy.

Authors:  L R Johnson; G Doherty; T Lairmore; J F Moley; L M Brunt; J Koenig; M G Scott
Journal:  Clin Chem       Date:  2001-05       Impact factor: 8.327

4.  Performance of Deming regression analysis in case of misspecified analytical error ratio in method comparison studies.

Authors:  K Linnet
Journal:  Clin Chem       Date:  1998-05       Impact factor: 8.327

5.  Effect of blood collection tubes on total triiodothyronine and other laboratory assays.

Authors:  Raffick A R Bowen; Yung Chan; Joshua Cohen; Nadja N Rehak; Glen L Hortin; Gyorgy Csako; Alan T Remaley
Journal:  Clin Chem       Date:  2004-12-02       Impact factor: 8.327

6.  Rapid cortisol assay during adrenal vein sampling in patients with primary aldosteronism.

Authors:  Giulio Mengozzi; Denis Rossato; Chiara Bertello; Corrado Garrone; Alberto Milan; Roberto Pagni; Franco Veglio; Paolo Mulatero
Journal:  Clin Chem       Date:  2007-09-27       Impact factor: 8.327

  6 in total
  2 in total

1.  Pathologic Blood Samples Tolerate Exposure to Vibration and High Turbulence in Simulated Drone Flights, but Plasma Samples Should be Centrifuged After Flight.

Authors:  Karl Arne Johannessen; Nina Kristin Steen Wear; Karin Toska; Morten Hansbo; Jens Petter Berg; Erik Fosse
Journal:  IEEE J Transl Eng Health Med       Date:  2021-01-20       Impact factor: 3.316

2.  A 'smart' tube holder enables real-time sample monitoring in a standard lab centrifuge.

Authors:  Tony Hoang; Nicholas Moskwa; Ken Halvorsen
Journal:  PLoS One       Date:  2018-04-16       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.