Dosimetric characterization and output verification for conical brachytherapy surface applicators. Part I. Electronic brachytherapy source.

Regina K Fulkerson1, John A Micka1, Larry A DeWerd1.   

Abstract

PURPOSE: Historically, treatment of malignant surface lesions has been achieved with linear accelerator based electron beams or superficial x-ray beams. Recent developments in the field of brachytherapy now allow for the treatment of surface lesions with specialized conical applicators placed directly on the lesion. Applicators are available for use with high dose rate (HDR)(192)Ir sources, as well as electronic brachytherapy sources. Part I of this paper will discuss the applicators used with electronic brachytherapy sources; Part II will discuss those used with HDR (192)Ir sources. Although the use of these applicators has gained in popularity, the dosimetric characteristics including depth dose and surface dose distributions have not been independently verified. Additionally, there is no recognized method of output verification for quality assurance procedures with applicators like these. Existing dosimetry protocols available from the AAPM bookend the cross-over characteristics of a traditional brachytherapy source (as described by Task Group 43) being implemented as a low-energy superficial x-ray beam (as described by Task Group 61) as observed with the surface applicators of interest.
METHODS: This work aims to create a cohesive method of output verification that can be used to determine the dose at the treatment surface as part of a quality assurance/commissioning process for surface applicators used with HDR electronic brachytherapy sources (Part I) and(192)Ir sources (Part II). Air-kerma rate measurements for the electronic brachytherapy sources were completed with an Attix Free-Air Chamber, as well as several models of small-volume ionization chambers to obtain an air-kerma rate at the treatment surface for each applicator. Correction factors were calculated using MCNP5 and EGSnrc Monte Carlo codes in order to determine an applicator-specific absorbed dose to water at the treatment surface from the measured air-kerma rate. Additionally, relative dose measurements of the surface dose distributions and characteristic depth dose curves were completed in-phantom.
RESULTS: Theoretical dose distributions and depth dose curves were generated for each applicator and agreed well with the measured values. A method of output verification was created that allows users to determine the applicator-specific dose to water at the treatment surface based on a measured air-kerma rate.
CONCLUSIONS: The novel output verification methods described in this work will reduce uncertainties in dose delivery for treatments with these kinds of surface applicators, ultimately improving patient care.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24506635      PMCID: PMC3987645          DOI: 10.1118/1.4862505

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  42 in total

1.  Monte Carlo simulation of a typical 60Co therapy source.

Authors:  G M Mora; A Maio; D W Rogers
Journal:  Med Phys       Date:  1999-11       Impact factor: 4.071

2.  Monte Carlo dosimetry of a new 192Ir high dose rate brachytherapy source.

Authors:  A Angelopoulos; P Baras; L Sakelliou; P Karaiskos; P Sandilos
Journal:  Med Phys       Date:  2000-11       Impact factor: 4.071

3.  AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology.

Authors:  C M Ma; C W Coffey; L A DeWerd; C Liu; R Nath; S M Seltzer; J P Seuntjens
Journal:  Med Phys       Date:  2001-06       Impact factor: 4.071

4.  Thermoluminescent dosimetry of the Symmetra 125I model I25.S06 interstitial brachytherapy seed.

Authors:  N S Patel; S T Chiu-Tsao; J F Williamson; P Fan; T Duckworth; D Shasha; L B Harrison
Journal:  Med Phys       Date:  2001-08       Impact factor: 4.071

5.  Thermoluminescent dosimetry of the selectseed 125I interstitial brachytherapy seed.

Authors:  G Anagnostopoulos; D Baltas; P Karaiskos; P Sandilos; P Papagiannis; L Sakelliou
Journal:  Med Phys       Date:  2002-05       Impact factor: 4.071

6.  Experimental determination of the anisotropy function and anisotropy factor for model 6711 I-125 seeds.

Authors:  R S Sloboda; G V Menon
Journal:  Med Phys       Date:  2000-08       Impact factor: 4.071

7.  Thermoluminescent dosimetry of the SourceTech Medical model STM1251 125I seed.

Authors:  Sou-Tung Chiu-Tsao; Tamara L Duckworth; Chih-Yun Hsiung; Zuofeng Li; Jeffrey Williamson; Neil S Patel; Louis B Harrison
Journal:  Med Phys       Date:  2003-07       Impact factor: 4.071

8.  The IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al-4 mm Cu HVL; 10-300 kV generating potential). Institution of Physics and Engineering in Medicine and Biology.

Authors: 
Journal:  Phys Med Biol       Date:  1996-12       Impact factor: 3.609

9.  Dosimetry of interstitial brachytherapy sources: recommendations of the AAPM Radiation Therapy Committee Task Group No. 43. American Association of Physicists in Medicine.

Authors:  R Nath; L L Anderson; G Luxton; K A Weaver; J F Williamson; A S Meigooni
Journal:  Med Phys       Date:  1995-02       Impact factor: 4.071

10.  Surface applicators for high dose rate brachytherapy in AIDS-related Kaposi's sarcoma.

Authors:  M D Evans; M Yassa; E B Podgorsak; T N Roman; L J Schreiner; L Souhami
Journal:  Int J Radiat Oncol Biol Phys       Date:  1997-10-01       Impact factor: 7.038

View more
  8 in total

1.  Recommendations of the Spanish brachytherapy group (GEB) of Spanish Society of Radiation Oncology (SEOR) and the Spanish Society of Medical Physics (SEFM) for high-dose rate (HDR) non melanoma skin cancer brachytherapy.

Authors:  S Rodríguez; M Arenas; C Gutierrez; J Richart; J Perez-Calatayud; F Celada; M Santos; A Rovirosa
Journal:  Clin Transl Oncol       Date:  2017-08-14       Impact factor: 3.405

Review 2.  Commissioning and periodic tests of the Esteya(®) electronic brachytherapy system.

Authors:  Cristian Candela-Juan; Yury Niatsetski; Zoubir Ouhib; Facundo Ballester; Javier Vijande; Jose Perez-Calatayud
Journal:  J Contemp Brachytherapy       Date:  2015-05-13

3.  Efficacy and safety of electronic brachytherapy for superficial and nodular basal cell carcinoma.

Authors:  Rosa Ballester-Sánchez; Olga Pons-Llanas; Cristian Candela-Juan; Francisco Javier Celada-Alvarez; Blanca de Unamuno-Bustos; Margarita Llavador-Ros; Antonio Ballesta-Cuñat; Christopher A Barker; Alejandro Tormo-Mico; Rafael Botella-Estrada; Jose Perez-Calatayud
Journal:  J Contemp Brachytherapy       Date:  2015-06-09

4.  Dosimetric characteristics of a new unit for electronic skin brachytherapy.

Authors:  Teresa Garcia-Martinez; Jan-Pieter Chan; Jose Perez-Calatayud; Facundo Ballester
Journal:  J Contemp Brachytherapy       Date:  2014-02-19

5.  Biological characterization of a novel in vitro cell irradiator.

Authors:  Tyler L Fowler; Michael M Fisher; Alison M Bailey; Bryan P Bednarz; Randall J Kimple
Journal:  PLoS One       Date:  2017-12-12       Impact factor: 3.240

6.  Optimum radiation source for radiation therapy of skin cancer.

Authors:  Habib Safigholi; William Y Song; Ali S Meigooni
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

7.  Varian HDR surface applicators - commissioning and clinical implementation.

Authors:  Ileana Iftimia; Andrea B McKee; Per H Halvorsen
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102

8.  Investigation of a source model for a new electronic brachytherapy tandem by film measurement.

Authors:  Elijah Martin; Keith Sowards; Brian Wang
Journal:  J Appl Clin Med Phys       Date:  2018-08-13       Impact factor: 2.102

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.