Literature DB >> 15365826

Detection of hyperbilirubinaemia in jaundiced full-term neonates by eye or by bilirubinometer?

Peter Szabo1, Martin Wolf, Hans Ulrich Bucher, Jean-Claude Fauchère, Daniel Haensse, Romaine Arlettaz.   

Abstract

UNLABELLED: The aim of this study was to compare predictions of hyperbilirubinaemia by eye, performed by trained physicians and nurses, with predictions obtained using two commercial bilirubinometers. Jaundice was assessed in 92 white and 48 non-white healthy full-term neonates using three non-invasive methods and by total serum bilirubin as the reference method. Clinical assessment of cephalocaudal progression of jaundice was carried out independently by a physician and by nurses. Simultaneously, the Minolta Airshields JM-102 was applied on the sternum, the BiliCheck on both the forehead and the sternum, and finally, serum bilirubin concentrations were determined. The Minolta JM-102 showed the best performance with r2 = 0.90, an intraclass correlation coefficient (ICC) of 0.93, and a 95% confidence interval (CI) of +/- 4 units (approx. 56 micromol/l). The BiliCheck performed slightly better on the forehead than over the sternum with r2=0.90, an ICC of 0.88, and a CI of +/- 62 microtmol/l. Assessment of jaundice by eye was least accurate with r2 = 0.74, an ICC of 0.67, and a CI of +/- 1.5 zones (corresponding to 75 Lmol/l). Skin pigmentation and ambient light both adversely affected noninvasive bilirubin estimation.
CONCLUSION: All three non-invasive methods are well suited for estimation of serum bilirubin but show large confidence intervals. In healthy term newborns, hyperbilirubinaemia (>250 Lmol/l) can be safely ruled out by eye if jaundice does not reach the abdomen or the extremities (Kramer zones 1 and 2), with < 22 units ( < 230 micromol/l) for the Minolta JM-102, or with a cut-off of 190 microlmol/l for the Bili-Check. If these respective thresholds are exceeded, serum bilirubin concentrations should be measured.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15365826     DOI: 10.1007/s00431-004-1533-7

Source DB:  PubMed          Journal:  Eur J Pediatr        ISSN: 0340-6199            Impact factor:   3.183


  21 in total

1.  Transcutaneous bilirubin measurement in newborn infants: evaluation of a new spectrophotometric method.

Authors:  M Beck; N Kau; H Schlebusch
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2003-07       Impact factor: 5.747

2.  Transcutaneous bilirubinometry decreases the need for serum bilirubin measurements and saves money.

Authors:  M J Maisels; E Kring
Journal:  Pediatrics       Date:  1997-04       Impact factor: 7.124

3.  Prediction of hyperbilirubinaemia in the healthy term newborn.

Authors:  X Carbonell; F Botet; J Figueras; A Riu-Godó
Journal:  Acta Paediatr       Date:  2001-02       Impact factor: 2.299

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  The meaning and use of the area under a receiver operating characteristic (ROC) curve.

Authors:  J A Hanley; B J McNeil
Journal:  Radiology       Date:  1982-04       Impact factor: 11.105

6.  Noninvasive measurement of total serum bilirubin in a multiracial predischarge newborn population to assess the risk of severe hyperbilirubinemia.

Authors:  V K Bhutani; G R Gourley; S Adler; B Kreamer; C Dalin; L H Johnson
Journal:  Pediatrics       Date:  2000-08       Impact factor: 7.124

7.  Can transcutaneous bilirubinometry reduce the need for blood tests in jaundiced full term babies?

Authors:  L Briscoe; S Clark; C W Yoxall
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2002-05       Impact factor: 5.747

8.  Assessment of a transcutaneous device in the evaluation of neonatal hyperbilirubinemia in a primarily Hispanic population.

Authors:  William D Engle; Gregory L Jackson; Dorothy Sendelbach; Denise Manning; William H Frawley
Journal:  Pediatrics       Date:  2002-07       Impact factor: 7.124

9.  Transcutaneous bilirubinometry: preliminary studies of noninvasive transcutaneous bilirubin meter in the Okayama National Hospital.

Authors:  I Yamanouchi; Y Yamauchi; I Igarashi
Journal:  Pediatrics       Date:  1980-02       Impact factor: 7.124

10.  Transcutaneous measurement of hyperbilirubinaemia: comparison of the Minolta jaundice meter and the Ingram icterometer.

Authors:  H Bilgen; Z Ince; E Ozek; N Bekiroglu; R Ors
Journal:  Ann Trop Paediatr       Date:  1998-12
View more
  7 in total

1.  Screening neonatal jaundice based on the sclera color of the eye using digital photography.

Authors:  Terence S Leung; Karan Kapur; Ashley Guilliam; Jade Okell; Bee Lim; Lindsay W MacDonald; Judith Meek
Journal:  Biomed Opt Express       Date:  2015-10-23       Impact factor: 3.732

2.  Use of a Smartphone App to Assess Neonatal Jaundice.

Authors:  James A Taylor; James W Stout; Lilian de Greef; Mayank Goel; Shwetak Patel; Esther K Chung; Aruna Koduri; Shawn McMahon; Jane Dickerson; Elizabeth A Simpson; Eric C Larson
Journal:  Pediatrics       Date:  2017-09       Impact factor: 7.124

3.  Impact of skin tone on photoacoustic oximetry and tools to minimize bias.

Authors:  Yash Mantri; Jesse V Jokerst
Journal:  Biomed Opt Express       Date:  2022-01-20       Impact factor: 3.732

4.  Agreement test of transcutaneous bilirubin and bilistick with serum bilirubin in preterm infants receiving phototherapy.

Authors:  Rinawati Rohsiswatmo; Hanifah Oswari; Radhian Amandito; Hikari Ambara Sjakti; Endang Windiastuti; Rosalina Dewi Roeslani; Indrayady Barchia
Journal:  BMC Pediatr       Date:  2018-09-29       Impact factor: 2.125

5.  Comparison of a new transcutaneous bilirubinometer (Bilimed) with serum bilirubin measurements in preterm and full-term infants.

Authors:  Tanja Karen; Hans Ulrich Bucher; Jean-Claude Fauchère
Journal:  BMC Pediatr       Date:  2009-11-12       Impact factor: 2.125

6.  Indirect neonatal hyperbilirubinemia in hospitalized neonates on the Thai-Myanmar border: a review of neonatal medical records from 2009 to 2014.

Authors:  L Thielemans; M Trip-Hoving; J Landier; C Turner; T J Prins; E M N Wouda; B Hanboonkunupakarn; C Po; C Beau; M Mu; T Hannay; F Nosten; B Van Overmeire; R McGready; V I Carrara
Journal:  BMC Pediatr       Date:  2018-06-12       Impact factor: 2.125

7.  Laboratory validation and field usability assessment of a point-of-care test for serum bilirubin levels in neonates in a tropical setting.

Authors:  Laurence Thielemans; Ahmar Hashmi; Dah Dah Priscilla; Moo Kho Paw; Tekel Pimolsorntong; Thatsanun Ngerseng; Bart Van Overmeire; Stephane Proux; François Nosten; Rose McGready; Verena I Carrara; Germana Bancone
Journal:  Wellcome Open Res       Date:  2018-11-23
  7 in total