| Literature DB >> 30268107 |
Rinawati Rohsiswatmo1,2, Hanifah Oswari1, Radhian Amandito2, Hikari Ambara Sjakti1, Endang Windiastuti1, Rosalina Dewi Roeslani1, Indrayady Barchia3.
Abstract
BACKGROUND: This study compares the minimally invasive Bilistick and a noninvasive method with standard Total Serum Bilirubin (TSB) measurement in preterm newborns receiving phototherapy. We assess the agreement of Transcutaneous Bilirubinometer (TcB) and Bilistick bilirubin measurements with standard TSB measurement in preterm infants receiving phototherapy.Entities:
Keywords: Bilistick; Indonesia; Phototherapy; Preterm infants; Transcutaneous bilirubin
Mesh:
Substances:
Year: 2018 PMID: 30268107 PMCID: PMC6162930 DOI: 10.1186/s12887-018-1290-9
Source DB: PubMed Journal: BMC Pediatr ISSN: 1471-2431 Impact factor: 2.125
Characteristics of subjects (n = 94)
| Characteristics | |
|---|---|
| Gender | |
| Male | 53 (56.4%) |
| Female | 41 (43.6%) |
| Age (days) Mean ± SD (range) | 2.94 ± 1.66 (1 to 10) |
| Gestational age (weeks) Mean ± SD (range) | 31.27 ± 2.32 (26 to 34) |
| Weight (g) Mean ± SD (range) | 1466.73 ± 442.24 (700 to 2450) |
Profile of bilirubin levels
| Measurement | Bilirubin level |
|---|---|
| Before phototherapy | |
| Total serum bilirubin (mg/dL) Mean ± SD (range) | 10.18 ± 2.91 (4.43 to 19.70) |
| Transcutaneous (mg/dL) Mean ± SD (range) | 10.99 ± 3.07 (4.70 to 19.60) |
| Bilistick (mg/dL) Mean ± SD (range) | 8.68 ± 2.78 (4.20 to 14.80) |
| 24 h of phototherapy | |
| Total serum bilirubin (mg/dL) Mean ± SD (range) | 9.75 ± 3.04 (3.25 to 18.20) |
| Transcutaneous (mg/dL) Mean ± SD (range) | 10.18 ± 3.44 (1.90 to 19.70) |
| Bilistick (mg/dL) Mean ± SD (range) | 8.33 ± 2.86 (3.20 to 17.90) |
| 48 h of phototherapy | |
| Total serum bilirubin (mg/dL) Median ± SD (range) | 8 (0.58 to 21.26) |
| Transcutaneous (mg/dL) Median ± SD (range) | 7.9 (1.70 to 22.30) |
| Bilistick (mg/dL) Median ± SD (range) | 6.85 (1.20 to 17.20) |
Fig. 1Scatter and Bland Altman plot between TcB and TSB (a and b), Bilistick and TSB (c and d) before phototherapy, Y equation = linear equation obtained from linear regression analysis. Y = estimated TSB, X = Bilirubin level of the device being tested () = average difference () = agreement limit () = tendency of average difference () = 0 point, standard to observe distance with red line
Fig. 2Scatter and Bland Altman plots between TcB and TSB (a and b), Bilistick and TSB (c and d) at 24 h of phototherapy. Y equation = linear equation obtained from linear regression analysis. Y = estimated TSB, X = Bilirubin level of the device being tested () = average difference () = agreement limit () = tendency of average difference () = 0 point, standard to observe distance with red line
Fig. 3Scatter and Bland Altman plot between TcB and TSB (a and b), Bilistick and TSB (c and d) at 48 h of phototherapy. Y equation = linear equation obtained from linear regression analysis. Y = estimated TSB, X = Bilirubin level of the device being tested () = average difference () = agreement limit () = tendency of average difference () = 0 point, standard to observe distance with red line