On the material and the tribology of alumina-alumina couplings for hip joint prostheses.

A Walter1.   

Abstract

The yearly sphericity deviation ranged from 0.03 mu to 3.7 mm in 48 retrieved implants with alumina--alumina bearing balls and cups. Excessive amounts of wear are predominantly design dependent or caused by malalignment. But the abrasion phenomena occurring in the cup centers and in the corresponding ball areas are related to unfavorable function zones. The influence of the lubricating gap geometry as studied in laboratory tests shows the form of sphericity deviations to be of decisive importance for wear and friction, if extended over small angles. Then, the contact stresses attributable to the actual effective curvatures of the bearing exceed the resistance to abrasion of the material under boundary lubrication conditions. There are remarkable differences between the early and actual material quality. Based on the state of knowledge of material aspects of wear criteria for the optimization of material and design, promising tolerable wear rates may be achieved.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1516324

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  6 in total

1.  A squeaky reputation: the problem may be design-dependent.

Authors:  Javad Parvizi; Bahar Adeli; Justin C Wong; Camilo Restrepo; Richard H Rothman
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

2.  Oxidized zirconium head on crosslinked polyethylene liner in total hip arthroplasty: a 7- to 12-year in vivo comparative wear study.

Authors:  George K Karidakis; Theofilos Karachalios
Journal:  Clin Orthop Relat Res       Date:  2015-12       Impact factor: 4.176

3.  Factors affecting modular acetabular ceramic liner insertion: a biomechanical analysis.

Authors:  James P McAuley; Douglas A Dennis; Justin Grostefon; William G Hamilton
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

Review 4.  Thirty years of experience with alumina-on-alumina bearings in total hip arthroplasty.

Authors:  Didier Hannouche; Amine Zaoui; Frédéric Zadegan; Laurent Sedel; Rémy Nizard
Journal:  Int Orthop       Date:  2010-12-30       Impact factor: 3.075

5.  Ceramic-ceramic bearing decreases osteolysis: a 20-year study versus ceramic-polyethylene on the contralateral hip.

Authors:  Philippe Hernigou; Sebastien Zilber; Paolo Filippini; Alexandre Poignard
Journal:  Clin Orthop Relat Res       Date:  2009-03-13       Impact factor: 4.176

Review 6.  Emergence of the alumina matrix composite in total hip arthroplasty.

Authors:  B Masson
Journal:  Int Orthop       Date:  2007-11-27       Impact factor: 3.075

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.