Literature DB >> 14715065

Theoretical modeling and evaluation of the axial resolution of the adaptive optics scanning laser ophthalmoscope.

Krishnakumar Venkateswaran1, Austin Roorda, Fernando Romero-Borja.   

Abstract

We present axial resolution calculated using a mathematical model of the adaptive optics scanning laser ophthalmoscope (AOSLO). The peak intensity and the width of the axial intensity response are computed with the residual Zernike coefficients after the aberrations are corrected using adaptive optics for eight subjects and compared with the axial resolution of a diffraction-limited eye. The AOSLO currently uses a confocal pinhole that is 80 microm, or 3.48 times the width of the Airy disk radius of the collection optics, and projects to 7.41 microm on the retina. For this pinhole, the axial resolution of a diffraction-limited system is 114 microm and the computed axial resolution varies between 120 and 146 microm for the human subjects included in this study. The results of this analysis indicate that to improve axial resolution, it is best to reduce the pinhole size. The resulting reduction in detected light may demand, however, a more sophisticated adaptive optics system. The study also shows that imaging systems with large pinholes are relatively insensitive to misalignment in the lateral positioning of the confocal pinhole. However, when small pinholes are used to maximize resolution, alignment becomes critical. ( c) 2004 Society of Photo-Optical Instrumentation Engineers.

Entities:  

Mesh:

Year:  2004        PMID: 14715065     DOI: 10.1117/1.1627775

Source DB:  PubMed          Journal:  J Biomed Opt        ISSN: 1083-3668            Impact factor:   3.170


  17 in total

1.  Comparison of Control Algorithms for a MEMS-based Adaptive Optics Scanning Laser Ophthalmoscope.

Authors:  Kaccie Y Li; Sandipan Mishra; Pavan Tiruveedhula; Austin Roorda
Journal:  Proc Am Control Conf       Date:  2009-06-10

2.  Three-dimensional relationship between high-order root-mean-square wavefront error, pupil diameter, and aging.

Authors:  Raymond A Applegate; William J Donnelly; Jason D Marsack; Darren E Koenig; Konrad Pesudovs
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2007-03       Impact factor: 2.129

3.  Large-field-of-view, modular, stabilized, adaptive-optics-based scanning laser ophthalmoscope.

Authors:  Stephen A Burns; Remy Tumbar; Ann E Elsner; Daniel Ferguson; Daniel X Hammer
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2007-05       Impact factor: 2.129

4.  Adaptive optics scanning laser ophthalmoscopy for in vivo imaging of lamina cribrosa.

Authors:  Abhiram S Vilupuru; Nalini V Rangaswamy; Laura J Frishman; Earl L Smith; Ronald S Harwerth; Austin Roorda
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2007-05       Impact factor: 2.129

5.  Effect of aberrations and scatter on image resolution assessed by adaptive optics retinal section imaging.

Authors:  Justin M Wanek; Marek Mori; Mahnaz Shahidi
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2007-05       Impact factor: 2.129

6.  Imaging Glaucomatous Damage Across the Temporal Raphe.

Authors:  Gang Huang; Ting Luo; Thomas J Gast; Stephen A Burns; Victor E Malinovsky; William H Swanson
Journal:  Invest Ophthalmol Vis Sci       Date:  2015-06       Impact factor: 4.799

7.  Noninvasive measurements and analysis of blood velocity profiles in human retinal vessels.

Authors:  Zhangyi Zhong; Hongxin Song; Toco Yuen Ping Chui; Benno L Petrig; Stephen A Burns
Journal:  Invest Ophthalmol Vis Sci       Date:  2011-06-13       Impact factor: 4.799

8.  New wrinkles in retinal densitometry.

Authors:  Benjamin D Masella; Jennifer J Hunter; David R Williams
Journal:  Invest Ophthalmol Vis Sci       Date:  2014-10-14       Impact factor: 4.799

9.  Wide-vergence, multi-spectral adaptive optics scanning laser ophthalmoscope with diffraction-limited illumination and collection.

Authors:  Sanam Mozaffari; Francesco LaRocca; Volker Jaedicke; Pavan Tiruveedhula; Austin Roorda
Journal:  Biomed Opt Express       Date:  2020-02-26       Impact factor: 3.732

10.  Is oblique scanning laser ophthalmoscope applicable to human ocular optics? A feasibility study using an eye model for volumetric imaging.

Authors:  Wenjun Shao; Weiye Song; Ji Yi
Journal:  J Biophotonics       Date:  2020-03-03       Impact factor: 3.207

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.