Evaluation of the impedance tonometers TGDc-01 and iCare according to the international ocular tonometer standards ISO 8612.

Peter Christian Ruokonen1, Thomas Schwenteck, Jörg Draeger.   

Abstract

BACKGROUND: Two new tonometers have been introduced that are based on the impedance principle. Both the TGDc-01 (transpalpebral measurement) and the iCare (corneal measurement) do not require corneal anaesthesia. The present work presents an evaluation of both devices.
METHODS: Comparative measurements using one of the new tonometers and applanation tonometry were performed by one investigator according to the international standard for ocular tonometer (ISO 8612). Measurements were performed on 445 eyes without corneal pathology from 243 patients. Six measurements were performed for iCare and 3 for TGDc, immediately followed by 3 applanation tonometry measurements.
RESULTS: The correlation coefficient with respect to applanation tonometry was 0.81 for TGDc and 0.95 for iCare. TGDc-01 measurements showed an average deviation of 3.1 +/- 2.6 mmHg to those of Perkins applanation tonometry. The maximum difference was 28.7 mmHg below and 9.8 mmHg above the results of applanation tonometry. iCare showed an average deviation of 2.5 +/- 1.1 mmHg to Goldmann tonometry. The maximum difference was 14.5 mmHg below and 9.8 mmHg above.
CONCLUSIONS: The results of both new tonometers showed a good correlation with the reference applanation tonometric methods, but the strict requirements of ISO 8612 are not fulfilled by either tonometer at present. Additionally, transpalpebral measurements with the TGDc-01 showed unacceptably high variability.

Entities:  

Mesh:

Year:  2007        PMID: 17318572     DOI: 10.1007/s00417-006-0483-3

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  14 in total

1.  A new induction-based impact method for measuring intraocular pressure.

Authors:  A I Kontiola
Journal:  Acta Ophthalmol Scand       Date:  2000-04

2.  [Prospective comparison of the new indentation tonometer TGdC-01, the non-contact tonometer PT100 and the conventional Goldmann applanation tonometer].

Authors:  A Müller; L Godenschweger; G E Lang; J Kampmeier
Journal:  Klin Monbl Augenheilkd       Date:  2004-09       Impact factor: 0.700

3.  Comparison of intraocular pressure measurements with the digital tonometer TGDc-01 'PRA' and the Goldmann applanation tonometer.

Authors:  Michael W Meyer; Roland Gockeln; Ludwig Hoy; Andrea Meyer; Carl Erb
Journal:  Ophthalmic Res       Date:  2004 Sep-Oct       Impact factor: 2.892

4.  Measurement of the intraocular pressure with the "transpalpebral tonometer" TGDc-01 in comparison with applanation tonometry.

Authors:  Dirk Sandner; Andreas Böhm; Sandra Kostov; Lutz Pillunat
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2004-12-22       Impact factor: 3.117

5.  Transpalpebral tonometry: reliability and comparison with Goldmann applanation tonometry and palpation in healthy volunteers.

Authors:  A Troost; S H Yun; K Specht; F Krummenauer; O Schwenn
Journal:  Br J Ophthalmol       Date:  2005-03       Impact factor: 4.638

Review 6.  Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach.

Authors:  M J Doughty; M L Zaman
Journal:  Surv Ophthalmol       Date:  2000 Mar-Apr       Impact factor: 6.048

7.  Central corneal thickness in low-tension glaucoma.

Authors:  N Ehlers; F K Hansen
Journal:  Acta Ophthalmol (Copenh)       Date:  1974

8.  Falsely elevated intraocular pressure due to increased central corneal thickness.

Authors:  A M Bron; C Creuzot-Garcher; S Goudeau-Boutillon; P d'Athis
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  1999-03       Impact factor: 3.117

9.  Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry*.

Authors:  Liane H van der Jagt; Nomdo M Jansonius
Journal:  Ophthalmic Physiol Opt       Date:  2005-09       Impact factor: 3.117

10.  Measuring intraocular pressure with the Pulsair 3000 and Rebound tonometers in elderly patients without an anesthetic.

Authors:  Antti Kontiola; Päivi Puska
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2003-11-21       Impact factor: 3.117

View more
  7 in total

1.  Evaluation of the Icare-ONE rebound tonometer as a self-measuring intraocular pressure device in normal subjects.

Authors:  Ioannis Halkiadakis; Aimilianos Stratos; George Stergiopoulos; Eleni Patsea; Sotiris Skouriotis; Panagiotis Mitropoulos; Dimitrios Papaconstantinou; Gerasimos Georgopoulos
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-08       Impact factor: 3.117

2.  [Influence of corneal thickness on intraocular pressure measurements following Descemet's stripping automated endothelial keratoplasty (DSAEK)].

Authors:  M K J Klamann; A-K B Maier; J Gonnermann; N Torun; P C Ruokonen
Journal:  Ophthalmologe       Date:  2012-11       Impact factor: 1.059

3.  The influence of refractive errors on IOP measurement by rebound tonometry (ICare) and Goldmann applanation tonometry.

Authors:  Teresio Avitabile; Antonio Longo; Daniela Rocca; Roberta Amato; Caterina Gagliano; Marine Castaing
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-08-29       Impact factor: 3.117

4.  Comparison of ICare and IOPen vs Goldmann applanation tonometry according to international standards 8612 in glaucoma patients.

Authors:  Milena Pahlitzsch; Jeanette Brünner; Johannes Gonnermann; Anna-Karina B Maier; Necip Torun; Eckart Bertelmann; Matthias Kj Klamann
Journal:  Int J Ophthalmol       Date:  2016-11-18       Impact factor: 1.779

5.  [Glaucoma attack following Hughes tarsoconjunctival flap: what should be done?].

Authors:  E Bertelmann; N Minko; N Torun
Journal:  Ophthalmologe       Date:  2012-09       Impact factor: 1.059

Review 6.  An overview of home tonometry and telemetry for intraocular pressure monitoring in humans.

Authors:  Edward Yung; Valerie Trubnik; L Jay Katz
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-05-29       Impact factor: 3.117

Review 7.  Ageing and ocular surface immunity.

Authors:  Alireza Mashaghi; Jiaxu Hong; Sunil K Chauhan; Reza Dana
Journal:  Br J Ophthalmol       Date:  2016-07-04       Impact factor: 4.638

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.