Evaluation of applanation resonator sensors for intra-ocular pressure measurement: results from clinical and in vitro studies.

A Eklund1, C Lindén, T Bäcklund, B M Andersson, O A Lindahl.   

Abstract

Glaucoma is an eye disease that, in its most common form, is characterised by high intra-ocular pressure (IOP), reduced visual field and optic nerve damage. For diagnostic purposes and for follow-up after treatment, it is important to have simple and reliable methods for measuring IOP. Recently, an applanation resonator sensor (ARS) for measuring IOP was introduced and evaluated using an in vitro pig-eye model. In the present study, the first clinical evaluation of the same probe has been carried out, with experiments in vivo on human eyes. There was a low but significant correlation between IOP(ARS) and the IOP measured with a Goldmann applanation tonometer (r = 0.40, p = 0.001, n = 72). However, off-centre positioning of the sensor against the cornea caused a non-negligible source of error. The sensor probe was redesigned to have a spherical, instead of flat, contact surface against the eye and was evaluated in the in vitro model. The new probe showed reduced sensitivity to off-centre positioning, with a decrease in relative deviation from 89% to 11% (1 mm radius). For normalised data, linear regression between IOP(ARS) and direct IOP measurement in the vitreous chamber showed a correlation of r = 0.97 (p < 0.001, n = 108) and a standard deviation for the residuals of SD < or = 2.18 mm Hg (n = 108). It was concluded that a spherical contact surface should be preferred and that further development towards a clinical instrument should focus on probe design and signal analysis.

Entities:  

Mesh:

Year:  2003        PMID: 12691439     DOI: 10.1007/BF02344887

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   3.079


  9 in total

1.  A catheter tactile sensor for measuring hardness of soft tissue: measurement in a silicone model and in an in vitro human prostate model.

Authors:  A Eklund; A Bergh; O A Lindahl
Journal:  Med Biol Eng Comput       Date:  1999-09       Impact factor: 2.602

2.  Applanation tonometry and central corneal thickness.

Authors:  N Ehlers; T Bramsen; S Sperling
Journal:  Acta Ophthalmol (Copenh)       Date:  1975-03

3.  [Intraocular pressure in normal human eyes].

Authors:  W LEYDHECKER; K AKIYAMA; H G NEUMANN
Journal:  Klin Monbl Augenheilkd Augenarztl Fortbild       Date:  1958

4.  The Goldmann applanation tonometer.

Authors:  R A MOSES
Journal:  Am J Ophthalmol       Date:  1958-12       Impact factor: 5.258

5.  The effect of corneal thickness on applanation tonometry.

Authors:  M M Whitacre; R A Stein; K Hassanein
Journal:  Am J Ophthalmol       Date:  1993-05-15       Impact factor: 5.258

6.  Clinical comparison of the XPERT non-contact tonometer and the conventional Goldmann applanation tonometer.

Authors:  M K Hansen
Journal:  Acta Ophthalmol Scand       Date:  1995-04

Review 7.  Sources of error with use of Goldmann-type tonometers.

Authors:  M M Whitacre; R Stein
Journal:  Surv Ophthalmol       Date:  1993 Jul-Aug       Impact factor: 6.048

8.  A resonator sensor for measurement of intraocular pressure--evaluation in an in vitro pig-eye model.

Authors:  A Eklund; T Bäcklund; O A Lindahl
Journal:  Physiol Meas       Date:  2000-08       Impact factor: 2.833

9.  Tonometry in adults and children. A manometric evaluation of pneumatonometry, applanation, and TonoPen in vitro and in vivo.

Authors:  D L Eisenberg; B G Sherman; C A McKeown; J S Schuman
Journal:  Ophthalmology       Date:  1998-07       Impact factor: 12.079

  9 in total
  5 in total

1.  Underestimation of intraocular pressure after photorefractive keratectomy: a biomechanical analysis.

Authors:  Per Hallberg; Anders Eklund; Kenneth Santala; Timo Koskela; Olof Lindahl; Christina Lindén
Journal:  Med Biol Eng Comput       Date:  2006-07-26       Impact factor: 2.602

2.  Symmetric sensor for applanation resonance tomometry of the eye.

Authors:  Per Hallberg; Christina Lindén; Tomas Bäcklund; Anders Eklund
Journal:  Med Biol Eng Comput       Date:  2006-03       Impact factor: 2.602

3.  Prostate tissue stiffness as measured with a resonance sensor system: a study on silicone and human prostate tissue in vitro.

Authors:  Ville Jalkanen; Britt M Andersson; Anders Bergh; Börje Ljungberg; Olof A Lindahl
Journal:  Med Biol Eng Comput       Date:  2006-06-10       Impact factor: 2.602

4.  A generalised porous medium approach to study thermo-fluid dynamics in human eyes.

Authors:  Alessandro Mauro; Nicola Massarotti; Mohamed Salahudeen; Mario R Romano; Vito Romano; Perumal Nithiarasu
Journal:  Med Biol Eng Comput       Date:  2018-03-22       Impact factor: 2.602

5.  Repeatability and reproducibility of applanation resonance tonometry: a cross-sectional study.

Authors:  Laura Ottobelli; Paolo Fogagnolo; Paolo Frezzotti; Stefano De Cillà; Elena Vallenzasca; Maurizio Digiuni; Ruggiero Paderni; Ilaria Motolese; Simone Alex Bagaglia; Eduardo Motolese; Luca Rossetti
Journal:  BMC Ophthalmol       Date:  2015-04-10       Impact factor: 2.209

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.