Determination of saturation, heart rate, and respiratory rate at forearm using a Nellcor™ forehead SpO2-saturation sensor.

Jarkko Harju1, Antti Vehkaoja2, Ville Lindroos3, Pekka Kumpulainen2, Sasu Liuhanen4, Arvi Yli-Hankala5,3, Niku Oksala3,6.   

Abstract

Alterations in arterial blood oxygen saturation, heart rate (HR), and respiratory rate (RR) are strongly associated with intra-hospital cardiac arrests and resuscitations. A wireless, easy-to-use, and comfortable method for monitoring these important clinical signs would be highly useful. We investigated whether the Nellcor™ OxiMask MAX-FAST forehead sensor could provide data for vital sign measurements when located at the distal forearm instead of its intended location at the forehead to provide improved comfortability and easy placement. In a prospective setting, we recruited 30 patients undergoing surgery requiring postoperative care. At the postoperative care unit, patients were monitored for two hours using a standard patient monitor and with a study device equipped with a Nellcor™ Forehead SpO2 sensor. The readings were electronically recorded and compared in post hoc analysis using Bland-Altman plots, Spearman's correlation, and root-mean-square error (RMSE). Bland-Altman plot showed that saturation (SpO2) differed by a mean of -0.2 % points (SD, 4.6), with a patient-weighted Spearman's correlation (r) of 0.142, and an RMSE of 4.2 points. For HR measurements, the mean difference was 0.6 bpm (SD, 2.5), r = 0.997, and RMSE = 1.8. For RR, the mean difference was -0.5 1/min (4.1), r = 0.586, and RMSE = 4.0. The SpO2 readings showed a low mean difference, but also a low correlation and high RMSE, indicating that the Nellcor™ saturation sensor cannot reliably assess oxygen saturation at the forearm when compared to finger PPG measurements.

Entities:  

Keywords:  Heart rate; Intraoperative monitoring; Plethysmography; Pulse oximetry; Respiratory rate

Mesh:

Substances:

Year:  2016        PMID: 27752932     DOI: 10.1007/s10877-016-9940-7

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  32 in total

Review 1.  Maximizing the laboratory setting for testing devices and understanding statistical output in pulse oximetry.

Authors:  Paul B Batchelder; Dena M Raley
Journal:  Anesth Analg       Date:  2007-12       Impact factor: 5.108

2.  Combined photoplethysmographic monitoring of respiration rate and pulse: a comparison between different measurement sites in spontaneously breathing subjects.

Authors:  L Nilsson; T Goscinski; S Kalman; L-G Lindberg; A Johansson
Journal:  Acta Anaesthesiol Scand       Date:  2007-08-15       Impact factor: 2.105

3.  Validation of a wrist monitor for accurate estimation of RR intervals during sleep.

Authors:  Ph Renevey; J Sola; P Theurillat; M Bertschi; J Krauss; D Andries; C Sartori
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2013

4.  Confidence interval estimation for the Bland-Altman limits of agreement with multiple observations per individual.

Authors:  G Y Zou
Journal:  Stat Methods Med Res       Date:  2011-06-24       Impact factor: 3.021

5.  ESC Guidelines on the diagnosis and treatment of peripheral artery diseases: Document covering atherosclerotic disease of extracranial carotid and vertebral, mesenteric, renal, upper and lower extremity arteries: the Task Force on the Diagnosis and Treatment of Peripheral Artery Diseases of the European Society of Cardiology (ESC).

Authors:  Michal Tendera; Victor Aboyans; Marie-Louise Bartelink; Iris Baumgartner; Denis Clément; Jean-Philippe Collet; Alberto Cremonesi; Marco De Carlo; Raimund Erbel; F Gerry R Fowkes; Magda Heras; Serge Kownator; Erich Minar; Jan Ostergren; Don Poldermans; Vincent Riambau; Marco Roffi; Joachim Röther; Horst Sievert; Marc van Sambeek; Thomas Zeller
Journal:  Eur Heart J       Date:  2011-08-26       Impact factor: 29.983

6.  Accuracy of pulse oximeter readings from probe placement on newborn wrist and ankle.

Authors:  N Phattraprayoon; S Sardesai; M Durand; R Ramanathan
Journal:  J Perinatol       Date:  2011-07-07       Impact factor: 2.521

7.  Respiration can be monitored by photoplethysmography with high sensitivity and specificity regardless of anaesthesia and ventilatory mode.

Authors:  L Nilsson; A Johansson; S Kalman
Journal:  Acta Anaesthesiol Scand       Date:  2005-09       Impact factor: 2.105

8.  Estimation of respiratory rate from ECG, photoplethysmogram, and piezoelectric pulse transducer signals: a comparative study of time-frequency methods.

Authors:  Shishir Dash; Kirk H Shelley; David G Silverman; Ki H Chon
Journal:  IEEE Trans Biomed Eng       Date:  2010-05       Impact factor: 4.538

9.  Postanesthesia monitoring revisited: frequency of true and false alarms from different monitoring devices.

Authors:  L Wiklund; B Hök; K Ståhl; A Jordeby-Jönsson
Journal:  J Clin Anesth       Date:  1994 May-Jun       Impact factor: 9.452

10.  Estimating respiratory and heart rates from the correntropy spectral density of the photoplethysmogram.

Authors:  Ainara Garde; Walter Karlen; J Mark Ansermino; Guy A Dumont
Journal:  PLoS One       Date:  2014-01-22       Impact factor: 3.240

View more
  3 in total

1.  Comparison of non-invasive blood pressure monitoring using modified arterial applanation tonometry with intra-arterial measurement.

Authors:  Jarkko Harju; Antti Vehkaoja; Pekka Kumpulainen; Stefano Campadello; Ville Lindroos; Arvi Yli-Hankala; Niku Oksala
Journal:  J Clin Monit Comput       Date:  2017-01-19       Impact factor: 2.502

2.  Photoplethysmography for blood volumes and oxygenation changes during intermittent vascular occlusions.

Authors:  T Y Abay; P A Kyriacou
Journal:  J Clin Monit Comput       Date:  2017-05-25       Impact factor: 2.502

Review 3.  Sensors and Functionalities of Non-Invasive Wrist-Wearable Devices: A Review.

Authors:  Aida Kamišalić; Iztok Fister; Muhamed Turkanović; Sašo Karakatič
Journal:  Sensors (Basel)       Date:  2018-05-25       Impact factor: 3.576

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.