Reliability and failure modes of narrow implant systems.

Ronaldo Hirata1, Estevam A Bonfante2, Rodolfo B Anchieta1,3,4, Lucas S Machado1,4, Gileade Freitas1,5, Vinicius P Fardin6, Nick Tovar1, Paulo G Coelho1,7,8.   

Abstract

OBJECTIVES: Narrow implants are indicated in areas of limited bone width or when grafting is nonviable. However, the reduction of implant diameter may compromise their performance. This study evaluated the reliability of several narrow implant systems under fatigue, after restored with single-unit crowns.
MATERIALS AND METHODS: Narrow implant systems were divided (n = 18 each), as follows: Astra (ASC); BioHorizons (BSC); Straumann Roxolid (SNC), Intra-Lock (IMC), and Intra-Lock one-piece abutment (ILO). Maxillary central incisor crowns were cemented and subjected to step-stress accelerated life testing in water. Use level probability Weibull curves and reliability for a mission of 100,000 cycles at 130- and 180-N loads (90 % two-sided confidence intervals) were calculated. Scanning electron microscopy was used for fractography.
RESULTS: Reliability for 100,000 cycles at 130 N was ∼99 % in group ASC, ∼99 % in BSC, ∼96 % in SNC, ∼99 % in IMC, and ∼100 % in ILO. At 180 N, reliability of ∼34 % resulted for the ASC group, ∼91 % for BSC, ∼53 % for SNC, ∼70 % for IMC, and ∼99 % for ILO. Abutment screw fracture was the main failure mode for all groups.
CONCLUSIONS: Reliability was not different between systems for 100,000 cycles at the 130-N load. A significant decrease was observed at the 180-N load for ASC, SNC, and IMC, whereas it was maintained for BSC and ILO. CLINICAL RELEVANCE: The investigated narrow implants presented mechanical performance under fatigue that suggests their safe use as single crowns in the anterior region.

Entities:  

Keywords:  Biomechanics; Dental implants; Fatigue; Reliability; Weibull

Mesh:

Substances:

Year:  2015        PMID: 26526325     DOI: 10.1007/s00784-015-1636-8

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.573


  29 in total

1.  Effect of implant-abutment connection design on reliability of crowns: regular vs. horizontal mismatched platform.

Authors:  Amilcar C Freitas Júnior; Estevam A Bonfante; Nelson R F A Silva; Leonard Marotta; Paulo G Coelho
Journal:  Clin Oral Implants Res       Date:  2011-08-03       Impact factor: 5.977

2.  Effect of implant diameter on reliability and failure modes of molar crowns.

Authors:  Amilcar C Freitas-Junior; Estevam A Bonfante; Leandro M Martins; Nelson R F A Silva; Leonard Marotta; Paulo G Coelho
Journal:  Int J Prosthodont       Date:  2011 Nov-Dec       Impact factor: 1.681

3.  Reliability testing of indirect composites as single implant restorations.

Authors:  Marcelo Suzuki; Estevam Bonfante; Nelson Rfa Silva; Paulo G Coelho
Journal:  J Prosthodont       Date:  2011-08-31       Impact factor: 2.752

4.  Mechanical Evaluation of Two Grades of Titanium Used in Implant Dentistry.

Authors:  Ronaldo Hirata; Estevam A Bonfante; Lucas S Machado; Nick Tovar; Paulo G Coelho
Journal:  Int J Oral Maxillofac Implants       Date:  2015 Jul-Aug       Impact factor: 2.804

5.  Influence of implant/abutment joint designs on abutment screw loosening in a dental implant system.

Authors:  Tsuyoshi Kitagawa; Yasuhiro Tanimoto; Misako Odaki; Kimiya Nemoto; Masahiro Aida
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2005-11       Impact factor: 3.368

6.  Laboratory simulation of Y-TZP all-ceramic crown clinical failures.

Authors:  P G Coelho; E A Bonfante; N R F Silva; E D Rekow; V P Thompson
Journal:  J Dent Res       Date:  2009-04       Impact factor: 6.116

7.  The impact of loads on standard diameter, small diameter and mini implants: a comparative laboratory study.

Authors:  Simon Rupert Allum; Rachel Anne Tomlinson; Rajendra Joshi
Journal:  Clin Oral Implants Res       Date:  2008-06       Impact factor: 5.977

8.  Effect of implant connection and restoration design (screwed vs. cemented) in reliability and failure modes of anterior crowns.

Authors:  Amilcar C Freitas; Estevam A Bonfante; Eduardo P Rocha; Nelson R F A Silva; Leonard Marotta; Paulo G Coelho
Journal:  Eur J Oral Sci       Date:  2011-08       Impact factor: 2.612

Review 9.  Mini implants for definitive prosthodontic treatment: a systematic review.

Authors:  Avinash S Bidra; Khalid Almas
Journal:  J Prosthet Dent       Date:  2013-03       Impact factor: 3.426

10.  Reliability of one-piece ceramic implant.

Authors:  Nelson R F A Silva; Paulo G Coelho; Carlos A O Fernandes; Jose M Navarro; Renata Alves Dias; Van P Thompson
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2009-02       Impact factor: 3.368

View more
  1 in total

1.  A Comparative Analysis of Implants Presenting Different Diameters: Extra-Narrow, Narrow and Conventional.

Authors:  Henrique Tuzzolo Neto; Alessandra Sayuri Tuzita; Sérgio Alexandre Gehrke; Renata de Vasconcellos Moura; Márcio Zaffalon Casati; Alfredo Mikail Melo Mesquita
Journal:  Materials (Basel)       Date:  2020-04-17       Impact factor: 3.623

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.