Optimization of injected dose based on noise equivalent count rates for 2- and 3-dimensional whole-body PET.

Carole Lartizien1, Claude Comtat, Paul E Kinahan, Nuno Ferreira, Bernard Bendriem, Régine Trébossen.   

Abstract

UNLABELLED: The noise equivalent count (NEC) rate index is used to derive guidelines on the optimal injected dose to the patient for 2-dimensional (2D) and 3-dimensional (3D) whole-body PET acquisitions.
METHODS: We performed 2D and 3D whole-body acquisitions of an anthropomorphic phantom modeling the conditions for (18)F-FDG PET of the torso and measured the NEC rates for different activity levels for several organs of interest. The correlations between count rates measured from the phantom and those from a series of whole-body patient scans were then analyzed. This analysis allowed validation of our approach and estimation of the injected dose that maximizes NEC rate as a function of patient morphology for both acquisition modes.
RESULTS: Variations of the phantom and patient prompt and random coincidence rates as a function of single-photon rates correlated well. On the basis of these correlations, we demonstrated that the patient NEC rate can be predicted for a given single-photon rate. Finally, we determined that patient single-photon rates correlated with the mean dose per weight at acquisition start when normalized by the body mass index. This correlation allows modifying the injected dose as a function of patient body mass index to reach the peak NEC rate in 3D mode. Conversely, we found that the peak NEC rates were never reached in 2D mode within an acceptable range of injected dose.
CONCLUSION: The injected dose was adapted to patient morphology for 2D and 3D whole-body acquisitions using the NEC rate as a figure of merit of the statistical quality of the sinogram data. This study is a first step toward a more comprehensive comparison of the image quality obtained using both acquisition modes.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12215569

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  26 in total

1.  Application of a generalized scan statistic model to evaluate TOF PET images.

Authors:  Suleman Surti; Joel S Karp
Journal:  IEEE Trans Nucl Sci       Date:  2011       Impact factor: 1.679

2.  FDG-PET: procedure guidelines for tumour imaging.

Authors:  Emilio Bombardieri; Cumali Aktolun; Richard P Baum; Angelika Bishof-Delaloye; John Buscombe; Jean François Chatal; Lorenzo Maffioli; Roy Moncayo; Luc Mortelmans; Sven N Reske
Journal:  Eur J Nucl Med Mol Imaging       Date:  2003-12       Impact factor: 9.236

3.  Simulations of the Effect of Collimation on Count Rates of an LSO PET System.

Authors:  Ruth E Schmitz; Paul E Kinahan; Robert L Harrison; Thomas K Lewellen
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2006

4.  Treatment response evaluation using 18F-FDOPA PET in patients with recurrent malignant glioma on bevacizumab therapy.

Authors:  Johannes Schwarzenberg; Johannes Czernin; Timothy F Cloughesy; Benjamin M Ellingson; Whitney B Pope; Tristan Grogan; David Elashoff; Cheri Geist; Daniel H S Silverman; Michael E Phelps; Wei Chen
Journal:  Clin Cancer Res       Date:  2014-03-31       Impact factor: 12.531

5.  Optimization of noise equivalent count rate performance for a partially collimated PET scanner by varying the number of septa.

Authors:  Ruth E Schmitz; Robert L Harrison; Charles W Stearns; Thomas K Lewellen; Paul E Kinahan
Journal:  IEEE Trans Med Imaging       Date:  2007-07       Impact factor: 10.048

6.  Measured count-rate performance of the Discovery STE PET/CT scanner in 2D, 3D and partial collimation acquisition modes.

Authors:  L R Macdonald; R E Schmitz; A M Alessio; S D Wollenweber; C W Stearns; A Ganin; R L Harrison; T K Lewellen; P E Kinahan
Journal:  Phys Med Biol       Date:  2008-06-23       Impact factor: 3.609

7.  A simulation study for estimating scatter fraction in whole-body 18F-FDG PET/CT.

Authors:  Shota Hosokawa; Kazumasa Inoue; Daisuke Kano; Fuminori Shimizu; Kazuya Koyama; Yoshihiro Nakagami; Yoshihisa Muramatsu; Masahiro Fukushi
Journal:  Radiol Phys Technol       Date:  2016-12-28

8.  Reliability of predicting image signal-to-noise ratio using noise equivalent count rate in PET imaging.

Authors:  Tingting Chang; Guoping Chang; John W Clark; Rami H Diab; Eric Rohren; Osama R Mawlawi
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

9.  Direct comparison between 2-dimensional and 3-dimensional PET acquisition modes for myocardial blood flow absolute quantification with O-15 water and N-13 ammonia.

Authors:  Véronique Roelants; Anne Bol; Xavier Bernard; Ann Coppens; Jacques Melin; Bernhard Gerber; Jean-Louis Vanoverschelde
Journal:  J Nucl Cardiol       Date:  2006 Mar-Apr       Impact factor: 5.952

Review 10.  Advantages and limitations of FDG PET in the follow-up of breast cancer.

Authors:  Peter Lind; Isabel Igerc; Thomas Beyer; Peter Reinprecht; Klaus Hausegger
Journal:  Eur J Nucl Med Mol Imaging       Date:  2004-04-15       Impact factor: 9.236

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.