Direct detector radiography versus dual reading computed radiography: feasibility of dose reduction in chest radiography.

Michael Gruber1, Martin Uffmann, Michael Weber, Mathias Prokop, Csilla Balassy, Cornelia Schaefer-Prokop.   

Abstract

The image quality of dual-reading computed radiography and dose-reduced direct radiography of the chest was compared in a clinical setting. The study group consisted of 50 patients that underwent three posteroanterior chest radiographs within minutes, one image obtained with a dual read-out computed radiography system (CR; Fuji 5501) at regular dose and two images with a flat panel direct detector unit (DR; Diagnost, Philips). The DR images were obtained with the same and with 50% of the dose used for the CR images. Images were evaluated in a blinded side-by-side comparison. Eight radiologists ranked the visually perceivable difference in image quality using a three-point scale. Then, three radiologists scored the visibility of anatomic landmarks in low and high attenuation areas and image noise. Statistical analysis was based on Friedman tests and Wilcoxon rank sum tests at a significance level of P<0.05. DR was judged superior to CR for the delineation of structures in high attenuation areas of the mediastinum even when obtained with 50% less dose (P<0.001). The visibility of most pulmonary structures was judged equivalent with both techniques, regardless of acquisition dose and speed level. Scores for image noise were lower for DR compared with CR, with the exception of DR obtained at a reduced dose. Thus, in this clinical preference study, DR was equivalent or even superior to the most modern dual read-out CR, even when obtained with 50% dose. A further dose reduction does not appear to be feasible for DR without significant loss of image quality.

Entities:  

Mesh:

Year:  2006        PMID: 16404566     DOI: 10.1007/s00330-005-0077-1

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  13 in total

1.  Digital chest radiography with a large-area flat-panel silicon X-ray detector: clinical comparison with conventional radiography.

Authors:  S P Hennigs; M Garmer; H J Jaeger; R Classen; A Jacobs; H M Gissler; A Christmann; K Mathias
Journal:  Eur Radiol       Date:  2001       Impact factor: 5.315

2.  Comparison of low-contrast detail perception on storage phosphor radiographs and digital flat panel detector images.

Authors:  S Peer; U Neitzel; S M Giacomuzzi; R Peer; E Gassner; I Steingruber; W Jaschke
Journal:  IEEE Trans Med Imaging       Date:  2001-03       Impact factor: 10.048

3.  Radiation dose reduction in chest radiography using a flat-panel amorphous silicon detector.

Authors:  W P Hosch; C Fink; B Radeleff; A kampschulte a; G W Kauffmann; J Hansmann
Journal:  Clin Radiol       Date:  2002-10       Impact factor: 2.350

4.  Comparison of visual grading analysis and determination of detective quantum efficiency for evaluating system performance in digital chest radiography.

Authors:  Patrik Sund; Magnus Båth; Susanne Kheddache; Lars Gunnar Månsson
Journal:  Eur Radiol       Date:  2003-10-16       Impact factor: 5.315

5.  Computed radiography and direct radiography: influence of acquisition dose on the detection of simulated lung lesions.

Authors:  Martin Uffmann; Mathias Prokop; Edith Eisenhuber; Michael Fuchsjäger; Michael Weber; Cornelia Schaefer-Prokop
Journal:  Invest Radiol       Date:  2005-05       Impact factor: 6.016

6.  Clinical comparative study with a large-area amorphous silicon flat-panel detector: image quality and visibility of anatomic structures on chest radiography.

Authors:  Christian Fink; Peter J Hallscheidt; Gerd Noeldge; Annette Kampschulte; Boris Radeleff; Waldemar P Hosch; Günter W Kauffmann; Jochen Hansmann
Journal:  AJR Am J Roentgenol       Date:  2002-02       Impact factor: 3.959

7.  Routine chest radiography using a flat-panel detector: image quality at standard detector dose and 33% dose reduction.

Authors:  Michael Strotzer; Markus Völk; Rüdiger Fründ; Okka Hamer; Niels Zorger; Stefan Feuerbach
Journal:  AJR Am J Roentgenol       Date:  2002-01       Impact factor: 3.959

8.  Improvement of signal-to-noise and contrast-to-noise ratios in dual-screen computed radiography.

Authors:  C C Shaw; T P Wang; D S Breitenstein; D Gur
Journal:  Med Phys       Date:  1997-08       Impact factor: 4.071

9.  Chest imaging with flat-panel detector at low and standard doses: comparison with storage phosphor technology in normal patients.

Authors:  A Herrmann; H Bonél; A Stäbler; C Kulinna; C Glaser; N Holzknecht; B Geiger; M Schätzl; F Reiser
Journal:  Eur Radiol       Date:  2001-11-13       Impact factor: 5.315

10.  Comparison of an amorphous silicon/cesium iodide flat-panel digital chest radiography system with screen/film and computed radiography systems--a contrast-detail phantom study.

Authors:  X J Rong; C C Shaw; X Liu; M R Lemacks; S K Thompson
Journal:  Med Phys       Date:  2001-11       Impact factor: 4.071

View more
  8 in total

1.  Single-exposure dual-energy subtraction chest radiography: detection of pulmonary nodules and masses in clinical practice.

Authors:  Zsolt Szucs-Farkas; Michael A Patak; Seyran Yuksel-Hatz; Thomas Ruder; Peter Vock
Journal:  Eur Radiol       Date:  2007-09-27       Impact factor: 5.315

2.  Image quality of supine chest radiographs: intra-individual comparison of computed radiography and low-dose linear-slit digital radiography.

Authors:  Zsolt Szucs-Farkas; Peter Vock
Journal:  Eur Radiol       Date:  2009-05-05       Impact factor: 5.315

3.  Assessment of patient doses in CR examinations throughout a large health region.

Authors:  Yogesh Thakur; Thorarin A Bjarnason; Kevin Hammerstrom; Lorie Marchinkow; Tim Koch; John E Aldrich
Journal:  J Digit Imaging       Date:  2012-02       Impact factor: 4.056

4.  [Dose reduction and adequate image quality in digital radiography: a contradiction?].

Authors:  S Pötter-Lang; M Dünkelmeyer; M Uffmann
Journal:  Radiologe       Date:  2012-10       Impact factor: 0.635

5.  [Detection of lung nodules. New opportunities in chest radiography].

Authors:  S Pötter-Lang; S Schalekamp; C Schaefer-Prokop; M Uffmann
Journal:  Radiologe       Date:  2014-05       Impact factor: 0.635

6.  Detection of urinary stones at reduced radiation exposure: a phantom study comparing computed radiography and a low-dose digital radiography linear slit scanning system.

Authors:  Zsolt Szucs-Farkas; Dev P Chakraborty; Harriet C Thoeny; Christos Loupatatzis; Peter Vock; Harald M Bonel
Journal:  AJR Am J Roentgenol       Date:  2009-06       Impact factor: 3.959

Review 7.  Digital chest radiography: an update on modern technology, dose containment and control of image quality.

Authors:  Cornelia Schaefer-Prokop; Ulrich Neitzel; Henk W Venema; Martin Uffmann; Mathias Prokop
Journal:  Eur Radiol       Date:  2008-04-23       Impact factor: 5.315

8.  An Experimental Study for Minimum Level of Decalcification to Detect the Osteolytic Bone Metastasis of Long Bone on Plain Radiography.

Authors:  Jun Ho Baek; Il Hyung Park; Sung Hwa Seo
Journal:  J Bone Metab       Date:  2016-08-31
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.