An elasto-plastic finite element model for polyethylene wear in total hip arthroplasty.

S H Teoh1, W H Chan, R Thampuran.   

Abstract

A new finite element model (FEM) based on an elasto-plastic behavior of ultra high molecular weight polyethylene (UHMWPE) was used to study the wear behavior of UHMWPE acetabular cup, which has a 32 mm diameter femoral head. The model imposed a plastic yield stress of 8 MPa on the UHMWPE so that any stresses beyond this would automatically be redistributed to its neighbor. The FEM model adopted a unique mesh design based on an open cube concept which eliminated the problems of singularities. Wear prediction combined the influences of contact stress, sliding distance and a surface wear coefficient. The new model predicted significantly higher volumetric wear rate (57 mm(3)/yr) well within the average reported clinical values. The model was also used to study the effect of friction and clearance between the acetabular cup and the femoral head. Increase in friction increased the volumetric wear rate but did not appear to affect the linear wear rate, which remained at 0.12 +/- 0.02 mm/yr. The predicted wear was sensitive to clearance. It was found that when the clearance was close to 0 and >0.5mm, severe wear occurred. The best clearance range was between 0.1 and 0.15 mm where the average linear wear rate was 0.1mm/yr and the volumetric wear was 55 mm(3)/yr. The present work indicates the importance of avoiding too tight or too loose a diametrical clearance.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11858807     DOI: 10.1016/s0021-9290(01)00215-9

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  14 in total

1.  Wear analysis of chamfered elongated acetabular cup liners.

Authors:  Hsiao-Che Lin; Tzuo-Liang Luo; Jian-Horng Chen
Journal:  Med Biol Eng Comput       Date:  2011-12-22       Impact factor: 2.602

2.  Computational analysis of polyethylene wear in anatomical and reverse shoulder prostheses.

Authors:  C Quental; J Folgado; P R Fernandes; J Monteiro
Journal:  Med Biol Eng Comput       Date:  2014-11-02       Impact factor: 2.602

3.  Wear testing and particle characterisation of sequentially crosslinked polyethylene acetabular liners using different femoral head sizes.

Authors:  Carmen Zietz; Christian Fabry; Lars Middelborg; Gerhard Fulda; Wolfram Mittelmeier; Rainer Bader
Journal:  J Mater Sci Mater Med       Date:  2013-04-25       Impact factor: 3.896

4.  Polyethylene wear is related to patient-specific contact stress in THA.

Authors:  Robert Košak; Veronika Kralj-Iglič; Aleš Iglič; Matej Daniel
Journal:  Clin Orthop Relat Res       Date:  2011-09-24       Impact factor: 4.176

5.  Quantification of clearance and creep in acetabular wear measurements.

Authors:  Mo Saffarini; Thomas Gregory; Eric Vandenbussche
Journal:  Ann Transl Med       Date:  2016-04

6.  Cross-shear implementation in sliding-distance-coupled finite element analysis of wear in metal-on-polyethylene total joint arthroplasty: intervertebral total disc replacement as an illustrative application.

Authors:  Curtis M Goreham-Voss; Philip J Hyde; Richard M Hall; John Fisher; Thomas D Brown
Journal:  J Biomech       Date:  2010-06-18       Impact factor: 2.712

7.  Dual mobility cups provide biomechanical advantages in situations at risk for dislocation: a finite element analysis.

Authors:  Alexandre Terrier; Adeliya Latypova; Maika Guillemin; Valérie Parvex; Olivier Guyen
Journal:  Int Orthop       Date:  2017-01-09       Impact factor: 3.075

8.  In vitro evaluation of stiffness graded artificial hip joint femur head in terms of joint stresses distributions and dimensions: finite element study.

Authors:  H Fouad
Journal:  J Mater Sci Mater Med       Date:  2011-04-20       Impact factor: 3.896

9.  Case reports: fractures of threaded cups: rare complications of a well-established implant.

Authors:  Arnd T Hoburg; Manav Mehta; Stephan Tohtz; Carsten Perka
Journal:  Clin Orthop Relat Res       Date:  2008-09-27       Impact factor: 4.176

10.  A novel formulation for scratch-based wear modelling in total hip arthroplasty.

Authors:  Karen M Kruger; Nishant M Tikekar; Anneliese D Heiner; Thomas E Baer; John J Lannutti; John J Callaghan; Thomas D Brown
Journal:  Comput Methods Biomech Biomed Engin       Date:  2013-01-10       Impact factor: 1.763

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.