Abstract

OBJECTIVE: The purpose of this study was to measure the in vitro range of motion of a mobile bearing inlay knee prosthesis under dynamic isokinetic loading conditions. Additionally, the effect on the range of motion of rotational malalignment of the tibia baseplate was determined.
DESIGN: Specimens with implanted knee prostheses were mounted onto a custom built knee simulator. 3-D inlay movement was measured by an ultrasonic tracking system.
BACKGROUND: More recent knee prostheses include mobile bearing inlays type designs. These systems are intended to allow higher conformity of the tibiofemoral joint and thereby decrease contact stress without decreasing the knee's range of motion.
METHODS: Dynamic testing in the knee simulator mimicked both the speed and resulting moment of a knee isokinetic extension test. The tibia baseplate was first implanted with no rotational malalignment, followed by sequential internal and external rotation of upto 15 degrees.
RESULTS: Correctly aligned, the inlay center moved 3.5 mm (SD, 1.5 mm) posterior during extension. With the tibia baseplate externally rotated more than 10 degrees the movement pattern changed.
CONCLUSION: At up to 10 degrees of rotational malalignment the primary motion pattern of the mobile bearing is maintained. However, beyond 10 degrees unintended motion may occur. RELEVANCE: These test results correlate to radiographic measurements of in vivo movements of mobile bearing inlays showing "paradoxical" movement of the mobile inlay compared to physiologic meniscal movement.

Entities:  

Mesh:

Year:  2002        PMID: 11779646     DOI: 10.1016/s0268-0033(01)00103-6

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  12 in total

1.  Influence of bi- and tri-compartmental knee arthroplasty on the kinematics of the knee joint.

Authors:  Markus Wünschel; Jiahsuan Lo; Torsten Dilger; Nikolaus Wülker; Otto Müller
Journal:  BMC Musculoskelet Disord       Date:  2011-01-27       Impact factor: 2.362

2.  In vitro investigation of the influence of tibial slope on quadriceps extension force after total knee arthroplasty.

Authors:  Sven Ostermeier; Christof Hurschler; Henning Windhagen; Christina Stukenborg-Colsman
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2006-04-21       Impact factor: 4.342

3.  Dynamic in-vivo tibio-femoral and bearing motions in mobile bearing knee arthroplasty.

Authors:  S Fantozzi; A Leardini; S A Banks; M Marcacci; S Giannini; F Catani
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2003-07-09       Impact factor: 4.342

4.  Femoro-tibial kinematics after TKA in fixed- and mobile-bearing knees in the sagittal plane.

Authors:  Kiriakos Daniilidis; Steffen Höll; Georg Gosheger; Ralf Dieckmann; Nicolo Martinelli; Sven Ostermeier; Carsten O Tibesku
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-04-04       Impact factor: 4.342

5.  Infrapatellar fat pad pressure and volume changes of the anterior compartment during knee motion: possible clinical consequences to the anterior knee pain syndrome.

Authors:  Michael Bohnsack; Christof Hurschler; Taskin Demirtas; Oliver Rühmann; Christina Stukenborg-Colsman; Carl-Joachim Wirth
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2004-10-29       Impact factor: 4.342

6.  [A novel knee endoprosthesis with a physiological joint shape. Part 1: Biomechanical basics and tribological studies].

Authors:  K-H Frosch; T Floerkemeier; C Abicht; P Adam; H Dathe; J Fanghänel; K M Stürmer; D Kubein-Meesenburg; H Nägerl
Journal:  Unfallchirurg       Date:  2009-02       Impact factor: 1.000

7.  Rotating platform versus fixed-bearing total knees: an in vitro study of wear.

Authors:  Hani Haider; Kevin Garvin
Journal:  Clin Orthop Relat Res       Date:  2008-08-29       Impact factor: 4.176

Review 8.  Clinical and radiological outcomes of fixed- versus mobile-bearing total knee replacement: a meta-analysis.

Authors:  Toby O Smith; Farshid Ejtehadi; Rachel Nichols; Leigh Davies; Simon T Donell; Caroline B Hing
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2009-09-25       Impact factor: 4.342

9.  [A new total knee arthroplasty with physiologically shaped surfaces. Part 2: First clinical results].

Authors:  K-H Frosch; H Nägerl; D Kubein-Meesenburg; J Buchholz; J Dörner; H Dathe; O Hellerer; C Dumont; K M Stürmer
Journal:  Unfallchirurg       Date:  2009-02       Impact factor: 1.000

10.  Prosthesis alignment affects axial rotation motion after total knee replacement: a prospective in vivo study combining computed tomography and fluoroscopic evaluations.

Authors:  Melinda K Harman; Scott A Banks; Stephan Kirschner; Jörg Lützner
Journal:  BMC Musculoskelet Disord       Date:  2012-10-23       Impact factor: 2.362

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.